CTRL Option (4)
Pinning (3)
Input Type (2)
Model Numbering
Refer to Applications Notes
DC/DC Converter
Output Power
Input Voltage (1)
Output Voltage
Single or Dual
MED-1
REV.: 05/2016www.recom-power.com
Notes:
Note1: for 4:1 Input Voltage Type add “W”, see Note 2.
2:1 4:1 “W”
xx= 4.5-9 Vin = “05” xx= 9-36Vin = “24”
xx= 9-18Vin = “12” xx= 18-75Vin = “48”
xx= 18-36Vin = “24”
xx= 36-75Vin = “48”
Note2: Blank for Standard 2:1 Input Voltage Range; „W“ suffix for 4:1 Input Voltage Range
Note3: „A“ suffix for A pinning; „C“ suffix for C pinning, for more details refer to Package Style and Pinning
Note4: „CTRL“ suffix for control pin option, for A pinning only, for C pinning not aviable
Examples:
REM6-0512D/A = 2:1 Input, 4.5-9Vin, ±12Vout, pinout „A“, without control pin
REM6-1215S/C = 2:1 Input, 9-18Vin, 15Vout, pinout „C“, without control pin
REM6-4815SW/A/CTRL = 4:1 Input, 36-75Vin, 15Vout, pinout „A“ with control pin
REM6 -243.3SW/C = 4:1 Input, 9-36Vin, 3.3Vout, pinout „C“, without control pin
6 Watt
2:1 & 4:1
DIP24
Single and Dual
Output
5
Y
E
A
R
w
a
r
r
a
n
t
y
5
REM 6- __ __ __ __ / __ __
Features
Regulated
Converters
• Reinforced Insulation for 250VAC Working Voltage
• Clearance and Creepage Distance: 8mm
• 5kVAC I/P to O/P 2MOPP Isolation
• 2µA Patient Leakage Current
• Industry Standard Pinout
• 2:1 and 4:1 Wide Input Range
Description
The REM6 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute
isolation with low 2µA leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The
compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions.
The outputs are short circuit and overload protected. The converters are available in two different pin-
ning options and optionally with an external control pin for standby consumption as low as 12.5mW. The
converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the CE and UL marks.
Selection Guide
Part Input Output Output Efficiency Max. Capacitive
Number Voltage Range Voltage Current typ. Load
(VDC) (VDC) (mA) (%) (µF)
REM6-xx3.3S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 3.3 1800 81.5 / 83.5 / 83 / 82.5 2100
REM6-xx05S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 5 1200 86 / 86 / 86 / 86.5 1500
REM6-xx12S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 12 500 86 / 89 / 89 / 88 260
REM6-xx15S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 15 400 87.5 / 88.5 / 88.5 / 88.5 210
REM6-xx24S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 24 250 87 / 88.5 / 88.5 / 88 75
REM6-xxx05D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 ±5 ±600 84 / 85 / 85 / 85 ±860
REM6-xx12D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 ±12 ±250 86.5 / 89 / 88.5 / 88 ±150
REM6-xx15D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75 ±15 ±200 87.5 / 88 / 88.5 / 87 ±110
REM6-xx3.3SW/ (3,4) 9-36 / 18-75 3.3 1800 83 / 82.5 2100
REM6-xx05SW/ (3,4) 9-36 / 18-75 5 1200 86 / 86.5 1500
REM6-xx12SW/ (3,4) 9-36 / 18-75 12 500 89 / 88 260
REM6-xx15SW/ (3,4) 9-36 / 18-75 15 400 89 / 88.5 210
REM6-xx24SW/ (3,4) 9-36 / 18-75 24 250 88.5 / 88 75
REM6-xx05DW/ (3,4) 9-36 / 18-75 ±5 ±600 85 / 85 ±860
REM6-xx12DW/ (3,4) 9-36 / 18-75 ±12 ±250 88.5 / 88 ±150
REM6-xx15DW/ (3,4) 9-36 / 18-75 ±15 ±200 88.5 / 87 ±110
REM6
IEC-60601-1 Certified
ES-60601-1 Certifed
EN-55011 Certified
EN-55022 Certified
2MOPP
250VAC
E314885