MTB080 -- ICCNexergy Page 1 of 3 MTB080 Series 65 - 100 W Medical Open Frame Power Supply * High Efficiency: Level V * High Power Density 12.5W/in3 * Output Floating * Lifetime Expectation >5 years * Hold-up Time 28ms at full load * Medical Approval - EN60601-1 Class I Elpac Part Number Output Voltage Output Current Peak Current Total Regulation Typical Efficiency MTB080009A 9.0V 7.2A (65W) 10.0A (90W) 5% 88% MTB080012A 12.0V 6.7A (80W) 8.3A (100W) 5% 90% MTB080015A 15.0V 5.4A (80W) 6.7A (100W) 5% 90% MTB080024A 24.0V 3.4A (80W) 4.2A (100W) 5% 91% Input Input Voltage 85 - 264VAC 100 - 240VAC Nominal Input Frequency 47 - 63Hz Input Current <2A rms Inrush Current <37A at 230VAC cold start Zero Load Power Consumption <0.3W Earth Leakage Current <75A @ 132VAC @ 60Hz <120A @ 264VAC @ 60Hz Patient Leakage Current <40A @ 132VAC @ 60Hz <60A @ 264VAC @ 60Hz Output Output Voltage See Table http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mtb080/ 5/27/2011 MTB080 -- ICCNexergy Page 2 of 3 Total Regulation +/-5% Minimum Load No minimum load required Start-Up Delay <0.2s Hold-Up Time >28ms at full load Ripple & Noise <1% pk-pk ** Over Voltage Protection 120-150% Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable Over Current Protetion 110 - 190% Short Circuit Protection shutdown, auto-restart (hiccup mode) General Efficiency Avg Efficiency 89.6% @ 115VAC; 90.3% @ 230VAC MTBF min. 200,000 hours demonstrated Size 4.00 Weight 0.37 lbs (.166 kg) Power Density 12.5W/in3 Environmental Operating Temperature 0 - 50C (Full load to 50C, derate linearly to 50% load at 70C) Storage Temperature -40C to +85C Relative Humidity 5-95%, non-condensing Cooling Natural Convection Vibration All units production tested to 19.6m/s2 EMC & Safety Emissions FCC class B, CISPR11 class B EN61000-3-2, -3 Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11 Certified by TUV to the following: cTUVus UL 60601-1 CAN/CSA-22.2 No.601.1-M90 http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mtb080/ 5/27/2011 MTB080 -- ICCNexergy Page 3 of 3 CB per IEC60601-1 CE marked to LVD MTB080 | ICCNEXERGY | (c) 2011 | http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mtb080/ 5/27/2011