MTB080 Series
65 - 100 W Medical Open Frame
Power Supply
High Efficiency: Level VHigh Power Density 12.5W/in3Output FloatingLifetime Expectation >5 yearsHold-up Time 28ms at full loadMedical Approval - EN60601-1 Cla ss I
Elpac Part
Number Output
Voltage Output
Current Peak
Current Total
Regulation Typical
Efficiency
MTB080009A 9.0V 7.2A (65W) 10.0A (90W) ±5% 88%
MTB080012A 12.0V 6.7A (80W) 8.3A (100W) ±5% 90%
MTB080015A 15.0V 5.4A (80W) 6.7A (100W) ±5% 90%
MTB080024A 24.0V 3.4A (80W) 4.2A (100W) ±5% 91%
Input
Input Voltage 85 - 264VAC 100 - 240VAC Nominal
Input Frequency 47 - 63Hz
Input Current <2A rms
Inrush Current <37A at 230VAC cold start
Zero Load Power Consumption <0.3W
Earth Leakage Current <75µA @ 132VAC @ 60Hz
<120µA @ 264VAC @ 60Hz
Patient Leakage Current <40µA @ 132VAC @ 60Hz
<60µA @ 264VAC @ 60Hz
Output
Output Voltage See Table
Pa
g
e 1 of 3MTB080
ICCNexer
gy
5/27/2011htt
p
://iccnexer
gy
.com/el
p
ac/medical-
p
owe
r
-su
pp
lies/o
p
en-frame/mtb080
/
Total Regulation +/-5%
Minimum Load No minimum load required
Start-Up Delay <0.2s
Hold-Up Time >28ms at full load
Ripple & Noise <1% pk-pk **
Over Voltage Protection 120-150%
Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable
Over Current Protetio n 110 - 190%
Short Circuit Protection shutdown, auto- restart (hiccup mode)
General
Efficiency Avg Efficiency 89.6% @ 115VAC; 90.3% @ 230VAC
MTBF min. 200,000 hours demons tr ated
Size 4.00
Weight 0.37 lbs (.166 kg)
Power Density 12.5W/in3
Environmental
Operating Temperature 0 – 50°C (Full load to 50°C , derate linearly to 50% load at 70°C)
Storage Temperature -40°C to +85°C
Relative Humidity 5-95%, non-condensing
Cooling Natural Convection
Vibration All units production tested to 19.6m/s2
EMC & Safety
Emissions FCC class B, CISPR11 class B EN61000-3-2, -3
Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11
Certified by TUV to the following: cTUVus
UL 60601-1
CAN/CSA-22.2 No.601.1-M90
Pa
g
e 2 of 3MTB080
ICCNexer
gy
5/27/2011htt
p
://iccnexer
gy
.com/el
p
ac/medical-
p
owe
r
-su
pp
lies/o
p
en-frame/mtb080
/
CB per IEC60601-1
CE marked to LVD