Amendment 1 to Operating Instructions A5E01132948-04
A5E31311334-03, 08/2013 3
In Section 4.3.3, Table 4.4 'Electric connection for PROFIBUS connection cable (accessory)' of items 1 and 3 the supply
voltage rating in the function column has changed to read now 19 ... 30.2 V DC.
In Section 5.1.4, Table 5.5 'Electric connection for PROFIBUS connection cable' of items 1 and 3 the supply voltage rating in
the function column has changed to read now 19 ... 30.2 V DC.
Section 5.2 applies to all FM and CSA applications and therefore its header is renamed to read 'FM/CSA'. Throughout this
section any 'FM' statement is modified to read 'FM/CSA' instead.
The following paragraph is added to this Section:
Special Conditions for Safe Use
● Always consult the manufacturer for dimensional information on the flameproof joints in case of repair.
● The painted surface of the SITRANS SL may be subject to electrostatic charge and thus become a source of ignition in
applications with a low relative humidity of less than 30 % when the painted surface is free of surface contamination such
as dirt, dust or oil. If necessary clean the painted surface only using a damp cloth.
● The SITRANS SL analyzers used for process measurement and control shall not measure oxygen concentrations
greater than 21%.
In Section 11.1, Table 3 'Electric characteristics' the rating of item 'Power supply' has changed. The correct rating is 24 V DC
nominal (19 ... 30.2 V DC).
The following two paragraphs are added to Annex A:
Classification according to Pressure Equipment Directive (97/23/EC)
For gases of fluid group 1 and liquids of fluid group 1 the SITRANS SL gas analyzer complies with the requirements of the
Pressure Equipment Directive 97/23/EC, article 3, paragraph 3 (sound engineering practice).
A marking in accordance with KC (Korea)
KC marking for use in Korea only
A급 기기
이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을 주의하시기 바라며
가정 외의 지역에서 사용하는 것을 목적으로 합니다.
Class A device This device is classified as having electromagnetic resistance suitable for industrial use (Class
A) and may be used universally, apart from in domestic use.
Siemens AG
Industry Sector
Amendment 1 to Operating Instructions A5E01132948
-04