Detailed specifications
Ti400 Ti300 Ti200
Temperature
Temperature measurement range
(not calibrated below -10 °C)
-20 °C to +1200 °C
(-4 °F to +2192 °F)
-20 °C to +650 °C
(-4 °F to +1202 °F)
Temperature measurement
accuracy ± 2 °C or 2 % (at 25 °C nominal, whichever is greater)
On-screen emissivity correction Yes (by number and table)
On-screen reflected background
temperature compensation Yes
On-screen transmission correction Yes
Imaging performance
Image capture frequency 9 Hz refresh rate or 60 Hz refresh rate depending upon model variation
Detector type Focal Plane Array, uncooled
microbolometer, 320 x 240 pixels
Focal Plane Array, uncooled
microbolometer, 240 X 180 pixels
Focal Plane Array, uncooled
microbolometer, 200 X 150 pixels
Thermal sensitivity (NETD) ≤ 0.05 °C at 30 °C target temp (50 mK) ≤ 0.075 °C at 30 °C target temp (75 mK)
Total pixels 76,800 43,200 30,000
Infrared spectral band 7.5 μm to 14 μm (long wave)
Visual (visible light) camera Industrial performance 5.0 megapixel
Standard infrared lens type
Field of view 24 ° x 17 °
Spatial resolution (IFOV) 1.31 mRad 1.75 mRad 2.09 mRad
Minimum focus distance 15 cm (approx. 6 in)
Optional telephoto infrared lens type, available soon
Field of view 12 ° x 9 °
Spatial resolution (IFOV) 0.65 mRad 0.87 mRad 1.05 mRad
Minimum focus distance 45 cm (approx. 18 in)
Optional wide-angle infrared lens type, available soon
Field of view 46 ° x 34 °
Spatial resolution (IFOV) 2.62 mRad 3.49 mRad 4.19 mRad
Minimum focus distance 15 cm (approx. 6 in)
Focus mechanism
LaserSharp® Auto Focus System Yes
Advanced Manual Focus Yes
Image presentation
Palettes
Standard Ironbow, Blue-Red, High Contrast, Amber, Amber Inverted, Hot Metal, Grayscale, Grayscale Inverted
Ultra Contrast™ Ironbow Ultra, Blue-Red Ultra, High Contrast Ultra, Amber Ultra, Amber Inverted Ultra,
Hot Metal Ultra, Grayscale Ultra, Grayscale Inverted Ultra
Level and span Smooth auto-scaling and manual scaling of level and span
Fast auto toggle between manual
and auto modes Yes
Fast auto-rescale in manual mode Yes
Minimum span (in manual mode) 2.0 °C (3.6 °F)
Minimum span (in auto mode) 3.0 °C (5.4 °F)
IR-Fusion® information
Picture-In-Picture (PIP) Yes
Full screen infrared Yes
AutoBlend™ mode Yes
Color alarms (temperature alarms) High-temperature , low-temperature, and isotherm (user-selectable)
Image capture and data storage
Image capture, review,
save mechanism One-handed image capture, review, and save capability
Storage medium Micro SD Memory Card, on-board flash memory, save-to-USB capability, direct download via USB-to-PC connection
File formats Non-radiometric (.bmp) or (.jpeg) or fully-radiometric (.is2) Video*: non-radiometric (MPEG - encoded .AVI) and fully-radiometric (.IS3)
No analysis software required for non-radiometric (.bmp, .jpg and .avi*) files
Export file formats w/SmartView®
software BMP, DIB, GIF, JPE, JFIF, JPEG, JPG, PNG, TIF, and TIFF
Memory review Thumbnail view navigation and review selection
Other time-saving and productivity features
Voice annotation 60 seconds maximum recording time per image; reviewable playback on imager
IR-PhotoNotes™ Yes
Wi-Fi connectivity Yes, to PC, iPhone®, iPad® and WiFi to LAN*
Text annotation* Yes
Video recording* Standard and Radiometric
Streaming Video Via USB to PC and HDMI to HDMI compatible screen
CNX™ Wireless System* Yes*
Fluke Connect compatible Yes
Auto capture (temperature and interval)* Yes*
Remote control and operation (for
special and advanced applications)
Yes No No
* Firmware updates for these features are not available yet in all countries.. Users notified via SmartView software when available.
2 Fluke Corporation Ti400, Ti300, and Ti200 Industrial Thermal Imagers