BCD.00287_AA_July-12-2013 1 www.power-one.com
DATA SHEET
Dialysis
Monitoring
Pumps
Surgical Devices
Home Health
Ultrasound
MBC60 SERIES MEDICAL 60W AC/DC
60 W convection cooled
-20 to 50 deg C full load operation
90-264 VAC input
2 x MOPP
2” x 4” x 1.2” (101.6 x 50.8 x 30.5 mm)
No minimum load required - single output models
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
No load power < 0.3 W
FEATURES
APPLICATIONS
MBC60 Series 2 www.power-one.com
DATA SHEET
TECHNICAL DATA:
Input
DESCRIPTION/CONDITION
Universal Input
90 - 264 Vac
47 to 400 Hz (3)
264 Vac (cold start)
40 A max.
264 Vac 50 / 60 Hz
250 μA max
120 Vac @ Full load
230 Vac @ Full load
< 140 uA
< 250 uA
Output
PARAMETER
DESCRIPTION/CONDITION
Voltage Adjustment
V1
± 10%
Transient Response
Main output 50 to 100% load change, 50/60
Hz, 50% duty cycle, 0.1 A / uSec
< 10%, recovery time < 5 mSec
Over Voltage Protection
130% Typical
Short Circuit Protection
130% Typical
Remote Sense
V1
0.5 Vdc compensation
Efficiency
>85%
Set Point Tolerance
V1 = ± 3%, V2 & V3 = ± 5%
Rise Time
<100 mSec
Ordering Information
PRODUCT
FAMILY
VOLTS
(VDC)
OUTPUT CURRENT
MAX (A)
MINIMUM
LOAD (A) (4)
RIPPLE &
NOISE
TOTAL
REGULATION
OVP THRESHOLD
MBC60-1005G
5.1
10.0
0
1%
± 0.8%
130% Typical
MBC60-1012G
12
5.0
0
1%
± 0.8%
130% Typical
MBC60-1015G
15
4.0
0
1%
± 0.8%
130% Typical
MBC60-1024G
24
2.5
0
1%
± 0.8%
130% Typical
MBC60-1048G
48
1.25
0
1%
± 0.8%
130% Typical
MBC60-3000G
5.2
8.0
0.5
1%
± 0.8%
130% Typical
12.5
3.0
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
MBC60-3001G
5.2
8.0
0.5
1%
± 0.8%
130% Typical
24.0
1.5
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
MBC60-3002G
5.2
8.0
0.5
1%
± 0.8%
130% Typical
14.6
2.5
0.1
1%
± 5.3%
-14.8
0.5
0.0
1%
± 5.3%
MBC60-3003G
3.3
8.0
1.0
1.5%
± 0.8%
130% Typical
5.2
3.0
0.1
1%
± 5.3%
-12.8
0.5
0.0
1%
± 5.3%
BCD.00287_AA_July-12-2013 3 www.power-one.com
DATA SHEET
Notes:
1. Maximum outputs for each output. Max power rating should not be exceeded.
2. Output noise measurement is made with a 20 MHz bandwidth using a 6” twisted pair, terminated with a 10 uF tantalum
capacitor in parallel with a 0.1 uF ceramic capacitor.
3. Safety Approved: 47 to 63 Hz.
4. Minimum load specified to meet cross regulation.
5. Specifications are for nominal input voltage, 25° C and max load unless otherwise stated.
6. Class 1 models have Earthing tab J4. Class 2 models (-2 suffix) have no Earthing tab.
7. Specifications subject to change without notice.
8. Warranty 2 years.
General Specifications
PARAMETER
DESCRIPTION/CONDITION
Hold Up Time
6 mSec
120 Vac input
MTBF
>100 khrs
MIL-HDBK-217F
Switching Frequency
67 kHz typical
Isolation Voltage
Input to Output: Min 4242 Vdc
Input to Ground 2120 Vdc
Weight
150 g (0.33 lbs)
Environmental
PARAMETER
DESCRIPTION/CONDITION
Operating Temperature
-20 to 50° C Full Load
50 to 70° C
Derate linearly to 50% at 70° C
Altitude
Operating 10,000 ft.
Non-operation 40,000 ft.
Relative Humidity
95%
Non-condensing
Conducted Emissions
EN55022, FCC part 15 Level B
Radiated Emissions
EN55022, FCC part 15 Level B
To be controlled in end system
Electromagnetic Susceptibility
EN61000-4
2, 3, 4, 5 Level 3
Harmonic Current
EN61000-3-2, Class A
Safety
PARAMETER
DESCRIPTION/CONDITION
EN / UL / CSA
60601-1 3rd Edition
MBC60 Series 4 www.power-one.com
DATA SHEET
Figure 1 Output Power Vs. Temperature
Figure 2 Dimension Drawing (Top and Side View)
Mechanical
INPUT = J1
EARTHING TAB = J4
DC OUTPUT = J2
REMOTE SENSE = J3
Pin 1: AC Neutral
Pin 2: AC Line
Molex: 19705-4301
Pin 1 = V1
Pin 2 = V1
Pin 3 = RTN
Pin 4 = RTN
Pin 5 = V3
Pin 6 = V2
Pin 1 = +V1 Sense
Pin 2 = -V1 Sense
Mating Connector:
Molex: 09-50-3031
Pins: 08-50-0106
Mating Connector:
Molex: 190030001
Mating Connector:
Tyco: 647402-6; Pins: 3-647409-1
Mating Connector:
Molex: 22-01-2021
Copyright © 2010 Power-One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherwise, the
trademarks and service marks of Power-One Inc. in the U.S. and other countries. All other product or service names are the property of their respective holders.
Power-One products are protected under numerous U.S. and foreign patents and pending applications, maskwork rights, and copyrights. Power-One reserves the
right to make changes to any products and services at any time without notice. Power-One assumes no responsibility or liability arising out of the application or use
of any information, product, or service described herein except as expressly agreed to in writing by Power-One Inc.
NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not designed, intended for use in, or authorized for use as critical components in life sup-
port systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president of
Power-One, Inc.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufac-
tured. Specifications are subject to change without notice.