Specifications
Measurements
Available measurements Vrms - Volts RMS VDF - Volts Distortion Factor
Arms - Amps RMS VTIF - Volts Telephone Influence Factor
WATT - Watts ATHD - Amps Total Harmonic Distortion
VA - Volt-Amps ADF - Amps Distortion Factor
VAR - Volt-Amps Reactive ATIF - Amps Telephone Influence Factor
FRQ - Frequency VF - Fundamental Volts rms
PF - Power factor AF - Fundamental Amps rms
VPK+ - Volts peak (positive) IMP - Impedance
VPK- - Volts peak (negative) RES - Resistance
APK+ - Amps peak (positive) REA - Reactance
APK- - Amps peak (negative) HR - Integrator time
VDC - Volts DC WHR - Watt Hours
ADC - Amps DC VAH - VA Hours
VRMN - Volts rectified mean VRH - VAr Hours
ARMN - Amps rectified mean AHR - Amp Hours
VCF - Voltage crest factor WAV - Average Watts
ACF - Current crest factor PFAV - Average Power Factor
VTHD - Volts Total Harmonic Distortion CVAR - Correction VArs
Voltage and current ranges
Voltage Ranges 2000 Vpeak, 1000 Vpeak, 500 Vpeak, 200 Vpeak, 100 Vpeak, 50 Vpeak, 20 Vpeak, 5 Vpeak, 2 Vpeak
Current Ranges (30 A Shunt) 200 Apeak, 100 Apeak, 50 Apeak, 20 Apeak, 10 Apeak, 5 Apeak, 2 Apeak, 1 Apeak, 0.5 Apeak, 0.2 Apeak, 0.1 Apeak
Current Ranges (1 A Shunt) 5 Apeak, 2 Apeak, 1 Apeak, 0.5 Apeak, 0.2 Apeak, 0.1 Apeak, 0.25 Apeak, 0.0125 Apeak, 0.005 Apeak, 0.0025 Apeak
Current Ranges (External
Shunt)
3 Vpeak, 1.5 Vpeak, 0.75 Vpeak, 0.3 Vpeak, 0.15 Vpeak, 0.075 Vpeak, 0.03 Vpeak, 0.015 Vpeak, 0.0075 Vpeak
Measurement accuracy - voltage
and current
Voltage Accuracy, Vrms ( 45 Hz
- 850 Hz )
± 0.04% of Reading ± 0.04% of Range ± 0.02 V
Voltage Accuracy, Vrms ( 10 Hz
- 45 Hz, 850 Hz - 1 MHz )
± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± 0.02 V (typical)
Voltage Accuracy, DC ± 0.05% of Reading ± 0.1% of Range ± 0.05 V
Voltage Accuracy, Vrmn ± 0.2% of Reading ± 0.1% of Range ± 0.1 V
Datasheet