SPECIFICATION RW-3023 ISSUE 7 Page 3
© 2004 Tyco Electronics Corporation. All rights reserved
If this document is printed it becomes uncontrolled. Check for the latest revision
4.3.1 Dimensions and Longitudinal Change
Three 150mm (6 inch) specimens of tubing, as supplied, shall be measured for length, to an accuracy of
± 1 mm (±1/32 inch) and inside diameter in accordance with ASTM D 2671. The specimens then shall be
conditioned for 3 minutes in a 200 ± 3°C (392 ± 5°F) oven, removed from the oven, cooled to 23 ± 3°C
(73 ± 5°F), re-measured for length, inside diameter and wall thickness in accordance with ASTM D 2671.
The longitudin al change shall be calculated as follows:
C = L1 - L0
L0 x 100
Where: C = Longitudinal Change (percent)
L
0 = Length Before Conditioning [mm (inches)]
L
1 = Length After Conditioning [mm (inches)]
4.3.2 Tensile Strength and Ultimate Elongation
The tensile strength and ultimate elongation of the tubing shall be determined in accordance with
ASTM D 2671 using 25mm (1 inch) bench marks and a 25 mm (1 inch) initial jaw separation. The speed of
jaw separation shall be 500 ± 50 mm (20 ± 2 inches) per minute.
4.3.3 Secant Modulus
The secant modulus of the tubing shall be tested using tubing as supplied in accordance with ASTM D 2671.
4.3.4 Copper Stability
Three 150mm (6 inch) specimens of tubing shall be slipped over a snug filling, straight, clean, bare copper
conductor. For tubing sizes 1/4 and smaller a solid conductor shall be used; for tubing sizes 3/8 and larger a
solid or tubular conductor shall be used. The specimens on the conductors shall be conditioned for 24 hours
in a desiccator or similar humidity chamber at 90 to 95 percent relative humidity and 23 ± 3°C (73 ± 5°F)
(three specimens shall be conditioned for 7 days in 158.0 ± 1.0°C (316.4± 1.8°F) oven. After conditioning,
the specimens shall be removed from the oven and cooled to 23 ± 3°C (73 ± 5°F). The copper conductor then
shall be removed from the tubing, the tubing and conductor shall then be examined. Darkening of the copper
due to normal air oxidation shall not be cause for rejection. The tubing then shall be conditioned at room
temperature for 16 to 96 hours and tested for ultimate elongation in accordance with 4.3.2.
4.3.5 Dielectric Withstand, Breakdown and Strength
The dielectric strength of the tubing shall be measured under oil in accordance with ASTM D 2671. Five
150mm (6 inch) specimens of tubing shall be recovered over a metal mandrel by conditioning for 3 minutes
in a 200 ± 3°C (392 ± 5°F) oven. The mandrel diameter shall be slightly larger than the fully recovered
inside diameter of the tubing being tested. The metal mandrel shall serve as one electrode and a 25mm (1
inch) wide strip of lead foil wrapped around the outside of the tubing as the other electrode. The test voltage
shall be applied at a rate of rise of 500 volts per second. Thickness measurements for calculating dielectric
strength shall be made adjacent to the point of breakdown. Specimens for dielectric withstand shall be held
for 60 seconds at 2500 volts.
4.3.6 Corrosive Effect
Three specimens of tubing shall be tested for copper contact corrosion in accordance with ASTM D 2671,
Procedure B. Three specimens shall be conditioned for 7 days in a 158.0 ± 1.0°C (316.4 ± 1.8°F) oven.
After conditioning, the specimens shall be visually examined for evidence of corrosion.
4.4 REJECTION AND RETEST
Failure of any samples of tubing to conform to any one of the requirements of this specification shall be
cause for rejection of the lot represented. Tubing which has been rejected may be replaced or reworked to
correct the defect and then resubmitted for acceptance. Before resubmitting, full particulars concerning the
rejection and the action taken to correct the defect shall be furnished to th e inspector.