Signal Analyzers
24.24
HP 11970, 11974 Series Mixers
HP 11974 Series Preselected Millimeter Mixers
Eliminate the need for signal identification at millimeter frequencies.
The HP 11974 series mixers are preselected from 26.5 to 75 GHz for
faster, easier testing of millimeter devices and systems. Preselection
reduces mixer overload from broadband signals and reduces radiation
of local oscillator harmonics back to the device-under-test. Equipment
operators can quickly locate true signals, and software development for
automated measurements is greatly simplified.
These mixers feature advanced barium-ferrite technology and
come with a standalone power supply. They are particularly useful
for broadband millimeter signal analysis, millimeter electromagnetic-
interference (EMI) measurements, and unattended monitoring of
millimeter signals.
HP 11974 series preselected mixers are available in four bands
Sensitivity11 dB
Frequency (displayed avg. Calibration Image Gain
HP range noise level/10 Hz) accuracy1rejection1compression
model (GHz) (dBm) (dB) (dB) (dBm)
11974A 26.5 to 40 –111 < ±2.3 –54 +6
11974Q 33 to 50 –106 < ±2.3 –50 +0
11974U 40 to 60 –109 < ±2.6 –50 +0
11974V 50 to 75 –100 < ±4.5 –40 +3
1 Specifications apply when connected to the HP 8566B or 70000 series spectrum
analyzers.
Compatibility
Upgrade kits are available to assure the compatibility of HP 8566A/B
spectrum analyzers and the HP 70907A external mixer interface mod-
ule. Consult your HP sales representative to determine requirements.
All HP 8560 E-series spectrum analyzers and the HP 70907B external
mixer interface modules are fully compatible with the HP 11974 series.
50 to 75 GHz Sweep Without 50 to 75 GHz Sweep Using
Preselection HP 11974 Series Mixer
HP 11970 Series Harmonic Mixers
The HP 11970 series waveguide mixers are general-purpose harmonic
mixers. They employ a dual-diode design to achieve flat-frequency re-
sponse and low conversion loss. These are achieved without external dc
bias or tuning stubs. Manual operation and computer-controlled hard-
ware operation are simplified because mixer bias and tuning adjust-
ments are not required.
HP 11970 series harmonic mixers are available in six bands
Noise Gain
Frequency LO level (dB) Freq.1compres-
HP range harm Conversion 1 kHz response sion
model (GHz) number loss (dB) RBW (dB) (dBm)
11970K 18 to 26.5 6+ 24 –105 ± 1.9 –3
11970A 26.5 to 40 8+ 26 –102 ± 1.9 –5
11970Q 33 to 50 10+ 28 –101 ± 1.9 –7
11970U 40 to 60 10+ 28 –101 ± 1.9 –7
11970V 50 to 75 14+ 40 –92 ± 2.1 –3
11970W 75 to 110 18+ 47 –85 ± 3.0 –1
1Frequency response of the mixers is reduced by 1 dB for LO range of 14 to 18 dBm.
Compatibility
The HP 11970 series mixers extend the frequency range of the HP 8560
E-series portable spectrum analyzers, of the HP 8566B spectrum analyzer
(used with the HP 11975A amplifier) and of the HP 70000 modular mea-
surement system (used with the HP 70907A/B external mixer inter-
face modules).
HP 11970 and 11974 Series Specifications
IF Range: dc to 1.3 GHz
LO Amplitude Range: +14 dB to +16 dB; +16 dB optimum
Calibration Accuracy: +2 dB for HP 11970 series with
optimum LO amplitude
Typical RF Input SWR: < 2.2:1, < 3.0:1 for HP 11974 series
Bias Requirements: None
Typical Odd-Order Harmonic Suppression: > 20 dB
(does not apply to HP 11974 series)
Maximum CW RF Input Level: +20 dBm (100 mW), +25 dBm for
HP 11974 series
Maximum Peak Pulse Power: 24 dBm (250 mW) with < 1 µs pulse
(avg. power = +20 dBm)
Bandwidth: 100 MHz minimum (HP 11974 series only)
Environmental: Meets MIL-T-28800, Type III, Class 3, Style C
IF/LO Connectors: SMA (female)
TUNE IN Connector: BNC
LO Range: 3 to 6.1 GHz
Key Literature
HP 11970 Series Technical Data, p/n 5954-2714
HP 11974 Series Technical Data, p/n 5952-2748
Ordering Information
HP 11974A 26.5 to 40 GHz Preselected Mixer
HP 11974Q 33 to 50 GHz Preselected Mixer
HP 11974U 40 to 60 GHz Preselected Mixer
HP 11974V 50 to 75 GHz Preselected Mixer
Opt 003 Delete Power Supply (HP 11974 series only)
HP 11970K 18 to 26.5 GHz Mixer
HP 11970A 26.5 to 40 GHz Mixer
HP 11970Q 33 to 50 GHz Mixer
HP 11970U 40 to 60 GHz Mixer
HP 11970V 50 to 75 GHz Mixer
HP 11970W 75 to 110 GHz Mixer
HP 11970
Opt 009 Mixer Connection Set adds three 1-m
low-loss SMA cables, wrench, Allen driver for
any HP 11970 series mixer
HP 11975A 2 to 8 GHz Amplifier
HP 281A/B Coaxial to Waveguide Adapters
R281A 26.5 to 40 GHz, 2.4 mm (f)
R281B 26.5 to 40 GHz, 2.4 mm (m)
Q281A 33 to 50 GHz, 2.4 mm (f)
Q281B 33 to 50 GHz, 2.4 mm (m)
• Preselected mixers to eliminate signal identification
• State-of-the-art technology
• Easier automated measurements
• Low conversion loss
• Individually amplitude calibrated
• No bias or tuning adjustments
• High 100 mW safe input level