749
Head-separated Dual Display Digital Pressure Sensor DPS-400 SERIES DPH-100 SERIES
FIBER
SENSORS
LASER
SENSORS
PHOTO-
ELECTRIC
SENSORS
MICRO
PHOTO-
ELECTRIC
SENSORS
AREA
SENSORS
SAFETY LIGHT
CURTAINS /
SAFETY
COMPONENTS
PRESSURE /
FLOW
SENSORS
INDUCTIVE
PROXIMITY
SENSORS
PARTICULAR
USE
SENSORS
SENSOR
OPTIONS
SIMPLE
WIRE-SAVING
UNITS
WIRE-SAVING
SYSTEMS
MEASURE-
MENT
SENSORS
STATIC
CONTROL
DEVICES
LASER
MARKERS
PLC
HUMAN
MACHINE
INTERFACES
ENERGY
MANAGEMENT
SOLUTIONS
FA
COMPONENTS
MACHINE
VISION
SYSTEMS
UV
CURING
SYSTEMS
Selection
Guide
Pressure/
Digital Display
Pressure/
Head-separated
Flow
DPC-L100/
DPH-L100
DPS-400/
DPH-100
DPC-100/
DPH-100
SPECIFICATIONS
Item
Type Compound pressure Positive pressure Vacuum pressure
±100 kPa type 1 MPa type –101 kPa type
Model No. (Note 3)
DPH-101(-R)
DPH-101-M3(-R)DPH-101-M5(-R)
DPH-102
DPH-102-M5(-R)
DPH-103(-R)
DPH-103-M3(-R)DPH-103-M5(-R)
CE marking directive compliance
EMC Directive, RoHS Directive
Type of pressure Gauge pressure
Rated pressure range –100.0 to +100.0 kPa 0 to +1.000 MPa 0 to –101.0 kPa
Pressure withstandability 500 kPa 1.5 MPa 500 kPa
Applicable uid Air, non-corrosive gas
Supply voltage 12 to 24 V DC ±10 % Ripple P-P 10 % or less
Current consumption 15 mA or less
Ambient temperature 0 to +50 °C +32 to +122 °F (No dew condensation allowed), Storage: –10 to +60 °C +14 to +140 °F
Ambient humidity 35 to 85 % RH, Storage: 35 to 85 % RH
Temperature characteristics Over ambient temperature range 0 to +50 °C +32 to +122 °F: within ±2 % F.S. of detected pressure at +25 °C +77 °F
Material Front case: PBT, Rear case: PBT (glass ber reinforced), Pressure port: Stainless steel (SUS303), O-ring: NBR
Pressure element: Silicon diaphragm, PPS
Cable
0.2 mm
2
4-core oil resistant cabtyre cable (Models with “-R” afxed to the Model No. have bending-resistant, oil-resistant cabtyre cable)
Cable extension Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable.
Weight Net weight DPH-10□(-R): Head 10 g approx. / Cable 40 g approx., DPH-10□-M3/M5(-R): Head 6 g approx. / Cable 40 g approx.
Gross weight DPH-10□(-R): 80 g approx., DPH-10□-M3/M5(-R): 70 g approx.
Accessory Connector (e-CON): 1 pc. (Note 4)
Notes: 1) Where measurement conditions have not been specied precisely, the conditions used were an ambient temperature of +25 °C +77 °F.
2) The sensor head can be used independently.
3) Model No. having the sufx “-R” is bending-resistant cable type. The bending-resistant cable type of DPH-102 is not available.
4) Connectors (e-CON) are available as spare parts. CN-EP2: 5 pcs.
Sensor heads
Controllers
Type For compound pressure / vacuum pressure For positive pressure
Item Model No. DPS-401 DPS-402
CE marking directive compliance
EMC Directive, RoHS Directive
Applicable sensor head Compound pressure : DPH-101□
Vacuum pressure : DPH-103□Positive pressure : DPH-102□
Rated pressure range Compound pressure: –100.0 to +100.0 kPa
Vacuum pressure: 0 to –101.0 kPa Positive pressure: 0 to +1.000 MPa
Display / set pressure range Compound pressure: –199.9 to +199.9 kPa
Vacuum pressure: +101.3 to –101.3 kPa Positive pressure: –1.050 to +1.050 MPa
Supply voltage 12 to 24 V DC ±10 % Ripple P-P 10 % or less
Power consumption
(Note 2)
Normal operation: 650 mW or less (Current consumption 25 mA or less at 24 V supply voltage)
ECO mode: 500 mW or less (Current consumption 20 mA or less at 24 V supply voltage)
Sensor head supply voltage Same as supply voltage
Sensor head input Input voltage range: 1 to 5 V DC (over rated pressure range)
Comparative outputs
(Comparative output 1, 2)
NP N open-collector transistor
• Maximum sink current: 50 mA (Note 3)
• Applied voltage: 30 V DC or less
(between comparative output and 0 V)
• Residual voltage: 1.5 V or less (Note 4) [at 50 mA (Note 3) sink current]
Output operation NO/NC, selectable by jog switch
Hysteresis Minimum 0 digit, variable by hysteresis setting
Repeatability
With compound pressure type connected: within ±0.2 % F.S. (±4 digits), With vacuum/positive pressure type connected: within ±0.2 % F.S. (±2 digits)
Response time 150 μs, 500 μs, 1 ms, 5 ms, 10 ms, 50 ms, 100 ms, 500 ms, selectable by jog switch
Ambient temperature
–10 to +50 °C +14 to +122 °F (cascading 8 to 16 controllers: –10 to +45 °C +14 to +133 °F) (No dew condensation or icing allowed), Storage: –20 to +70 °C – 4 to +158 °F
Ambient humidity 35 to 85 % RH, Storage: 35 to 85 % RH
Temperature characteristics Over ambient temperature range –10 to +50 °C +14 to +122 °F: within ±0.5 % F.S. of pressure at +25 °C +77 °F
Material Enclosure: Heat-resistant ABS, Protective cover: Polycarbonate, Jog switch: ABS, MODE key: Acrylic
Weight Net weight: 20 g approx.
Notes: 1) Where measurement conditions have not been specied precisely, the conditions used were an ambient temperature of +20 °C +68 °F.
2) Excluding current consumption of the applicable pressure sensor head.
3) 25 mA max. if 5 or more controllers are connected together.
4) In case of using the quick-connection cable (cable length: 5 m 16.404 ft)(optional)
5) The cable for controller connection is not supplied as an accessory. Make sure to use the quick-connection cables (optional) given below.
Main cable (4-core): CN-74-C1 (cable length: 1m 3.281 ft), CN-74-C2 (cable length: 2 m 6.562 ft), CN-74-C5 (cable length: 5 m 16.404 ft)
Sub cable (2-core): CN-72-C1 (cable length: 1m 3.281 ft), CN-72-C2 (cable length: 2 m 6.562 ft), CN-72-C5 (cable length: 5 m 16.404 ft)
6) The values specied above are applied only to the controller.