SPECIFICATION
25W AC-DC Single Output Medical Adaptor GSM25E serie s
File Name:GSM25E-SPEC 2014-01-02
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE Note.2
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
RIPPLE & NOISE (max.) Note.3
OUTPUT
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 7)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION
OTHERS
NOTE
PACKING
PROTECTION
5V
GSM25E05
GSM25E05-P1J
GSM25E07-P1J
GSM25E09-P1J
GSM25E12-P1J
GSM25E15-P1J
GSM25E18-P1J
GSM25E24-P1J
GSM25E48-P1J
7.5V
GSM25E07
9V
GSM25E09
12V
GSM25E12
15V
GSM25E15
18V
GSM25E18
24V
GSM25E24
48V
GSM25E48
60mVp-p 80mVp-p 80mVp-p 120mVp-p 120mVp-p 150mVp-p 180mVp-p 240mVp-p
VOLTAGE TOLERANCE Note.4
LINE REGULATION Note.5
LOAD REGULATION
±6.0% ±5.0% ±5.0% ±3.0% ±3.0% ±3.0% ±2.0% ±2.0%
±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0%
±6.0% ±5.0% ±5.0% ±3.0% ±3.0% ±3.0% ±2.0% ±2.0%
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
80% 83%
7.88 ~ 10.5V
84%
9.45 ~ 13V
86%
12.6 ~ 17.2V
86%
15.75 ~ 20.25V
86%
18.9 ~ 25.2V
87%
25.2 ~ 32.4V
88%
50.4 ~ 64.8V
0.7A / 115VAC 0.35A / 230VAC
55A / 230VAC 30A / 115VAC
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
796.7K hrs min. MIL-HDBK-217F(25℃)
-25 ~ +60℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85℃, 10 ~ 95% RH
±0.03% / ℃ (0~40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
79*54*33mm (L*W*H)
230g ; 60pcs / 14.8Kg / CARTON
CONNECTOR
PLUG
CABLE
Standard type ψ ψ for stock ther type available by customer requestedP1J: 2.1 * 5.5 * 11mm, tuning fork type, center positive ; O
See page 2
1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives.
For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
OVERLOAD
OVER VOLTAGE
105 ~ 170% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.25 ~ 7.5V
4A 2.93A 2.77A 2.08A 1.66A 1.38A 1.04A 0.52A
0 ~ 4A 0 ~ 2.93A 0 ~ 2.77A 0 ~ 1.66A 0 ~ 1.38A 0 ~ 1.04A
0 ~ 0.52A
20W 22W 25W 25W
0 ~ 2.08A
25W 25W 25W 25W
Protection type : Shut down o/p voltage, re-power on to recover
500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load
16ms / 230VAC 16ms / 115VAC at full load
SETUP, RISE TIME Note.6
HOLD UP TIME (Typ.)
I/P-O/P:4KVAC
I/P-O/P / 25℃/ 70% RH:100M Ohms / 500VDC
EMC IMMUNITY
EMC EMISSION
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A
TUV EN60601-1 / 60601-1-11 approved
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3
Touch current < 50 A/264VAC
μ