12
Relative Humidity : Maximum relative humidity 80% for temperature up to 31oC
decreasing linearly to 50% relative humidity at 40oC
Altitude : Operating below 2000m
Storage Temperature : -20oC to 60oC, < 80% R.H. (with battery removed)
Temperature Coefficient : nominal 0.15 x (specified accuracy)/ oC @(0oC -18oC or 28oC
-40oC), or otherwise specified
Sensing : Average sensing for model 131; True RMS sensing for model 132 & 135
Safety : Meets IEC61010-2-032(1994), EN61010-2-032(1995), UL3111-2-032(1999).
Measurement Category III 600 Volts ac & dc
Transient protection : 6.5kV (1.2/50µs surge) for all models
Pollution degree : 2
E.M.C. : Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 45 digits
Performance above 3V/m is not specified
Overload Protections :
ACA Clamp-on jaws : AC 1000A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Power Supply : standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Power Consumption :
Voltage & ACA functions: 3.5mA typical
Ohm & Temperature functions: 4mA typical
APO Timing : Idle for 16 minutes
APO Consumption : 10µA typical
Dimension : L224mm X W78mm X H40mm
Weight : 224 gm approx
Jaw opening & Conductor diameter : 45mm max
Special features : THD%-R Total harmonic distortion-RMS (ACD-40PQ only); Display
Backlight ACD-16 TRMS & ACD-40PQ only); Auto-Hold; Display Hold; On screen
stand-alone Hi-Lo logging (5400 minutes) at sampling speed of faster than:
20 per second for Voltage & ACA functions
4 per second for Ohm & Temperature functions
2 per second for Hz function
Accessories : Test leads (MTL-90B), batteries installed, user's manual, TPK-59 banana
plug type-K thermocouple & soft carrying pouch
O
ptional accessories : RS-232 KIT2 interface kit.
Electrical Specifications
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23 oC ±5 oC &
less than 75% R.H.