Extended Specications
1586A Super-DAQ
Precision Temperature Scanner
• Measurethermocouples,PRTs,
thermistors,dcvoltage,dccurrent,
andresistance
• Best-in-classtemperaturemeasurement
accuracy:
-PRTs:±0.005°C(usingexternal
DAQ-STAQMultiplexer)
-Thermocouples:±0.5°C
(usingHigh-CapacityModuleand
internalCJC)
-Thermistors:±0.002°C
• InputChannels:Upto40isolated
universalinputs
• Flexibleconfiguration:InternalHigh-
CapacityModuleand/orDAQ-STAQ
Multiplexer
• Selectablescanspeed:Upto10
channelspersecond
• Fourmodesofoperation:Scan,
Monitor,Measure,DigitalMultimeter
(DMM)
• Real-timecolortrending:Chartuptofour
channelssimultaneously
• Automatedsensorcalibration:Control
FlukeCalibrationtemperaturesources
suchasdry-wellsormicro-bathsfor
automatedcalibrationroutines
• Datastorage:Recordsupto20MBof
dataandsetupfilestointernalnon-volatile
memoryortoanexternalUSBdrive.
TransferdatatoaPCusingtheUSBdrive
orLANconnectionandviewdatain
Microsoft®Excel
• Datasecurity:Administratoranduser
profilesforprotectingsettingsand
ensuringtesttraceability
• Mx+Bscalingandchanneloffsetzero
function
• Alarms:Twoindependent,user-defined
alarmsforeachchannelindicatewhenan
upperorlowerrangehasbeenexceeded
The1586ASuper-DAQ
isthemostaccurate
andflexibletemperature
dataacquisitionsystemon
themarket.Itscansandrecordstemperature,dcvoltage,dccurrent,
andresistanceforupto40inputchannelsandscanspeedsasfast
as10channelspersecond.TheSuper-DAQcanbeconfiguredforuse
asamulti-channeldataloggerinthefactoryorasaprecisionreference
thermometerforbenchtopsensorcalibrationinthelab.
2 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner
1586A Super-DAQ at a glance
Colordisplaywithchannel
statusindicators.Chartup
to4channelssimultaneously.
Front panel
Back panel
Backlitmainfunctionkeys
alwaysknowthemodeof
operationandrecording
status.
Memorymanagement—
fordownloadingandmoving
dataandsetuples.
Measure/DMMfunction
forquickmeasurements
andrecordingwithminimal
setup.
Tworearpanelslotscanbe
conguredfortheinternal
High-CapacityModuleor
theexternalDAQ-STAQ
Multiplexeroroneofeach.
Functionsoftkeyssupport
fastsetupandoperation.
TotalizerandDigitalI/O
inputports.
Digitalexternalalarm
triggeroutputs.
Standbykeyturnsthe
displayandkeypadoff—
eliminateswarm-uptime.
USBandEthernet
portsusedforremote
operation.
Five-waybindingpostfor
referencethermometers.
Alsomeasuresdcvoltage,
dccurrent,andresistance.
Linevoltageand
fuseselector.
Navigationkeysforfast
selectionofinformation.
Powerswitchandline
cordconnector.
USBsupportforeasy
transferofdataand
setuples.
Dedicatedconnector
forcontrolofFluke
Calibrationdry-wellsor
temperaturebathsfor
automatedtests.
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 3
Twenty2-wireorten3-,
or4-wireconnections.
Built-incablestrainreliefto
helpsecureconnections.
1586A modules at a glance
1586-2588 DAQ-STAQ Multiplexer
1586-2586 High-Capacity Module
Twodedicateddc
currentchannels.
ElectronicIDwithserialnumber
andcalibrationdatereportsto
the1586ASuper-DAQ.
Twentymini-jackthermocouple
inputsanduptotenfour-wire,
gold-platedPRT/thermistor
connectors.
Cleartopcasewithembedded
channelnumbersforeasy
wiringidentication.
Universal input channels designed to measure temperature,
resistance, dc voltage, and dc current for data acquisition
applications on the factory floor where channel count and
scan speed are important.
Designed for high-accuracy measurements in secondary
temperature calibration labs. Easily connect/disconnect
thermocouples, PRTs, and thermistors.
4 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner
1586A Super-DAQ general specications
Mains Voltage 100 V Setting 90 V to 110 V
120 V Setting 108 V to 132 V
220 V Setting 198 V to 242 V
240 V Setting 216 V to 264 V
Frequency 47 Hz to 440 Hz
Power Consumption 36 VA peak (24 W average)
Environment Temperature Operating 0 °C to 50 °C
Full accuracy 18 °C to 28 °C
Storage −20 °C to 70 °C
Warm-up 1 hour to full accuracy specifications
Relative Humidity
(non-condensing)
Operating 0 °C to 30 °C <80 %
30 °C to 50 °C <50 %
Storage −20 °C to 70 °C <95 %
Altitude Operating 2,000 m
Storage 12,000 m
Vibration and Shock Complies with MIL-PRF-28800F Class 3
Channel Capacity Total analog channels 45
Voltage/resistance channels 41
Current channels 5
Digital I/O 8 bits
Totalizer 1
Alarm outputs 6
Trigger input 1
Input Protection 50 V all functions, terminals and ranges
Math Channels Number of channels 20
Operations sum, difference, multiply, divide, polynomial,
power, square root, reciprocal, exponential,
logarithm, absolute value, average, maxi-
mum, minimum
Triggers Interval, external (trigger input), alarm, remote (bus), manual, automated test
Memory Scan data RAM 75,000 readings with timestamp
Data/Setup flash memory 20 MB
USB Host Port Connector type Type A
Function Memory
File system FAT32
Memory capacity 32 GB
USB Device Port Connector type Type B
Class Instrument
Function Control and data transfer
Command protocol SCPI
General
Maximum input 50 V
Offset voltage <2 μV
3-wire internal resistance mismatch <50 mΩ
Basic CJC accuracy 0.25 °C
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 5
1586-2586 High-Capacity Module specications
1586-2588 DAQ-STAQ Multiplexer specications
General
Maximum input 50 V
Offset voltage <2 μV
3-wire internal resistance mismatch <50 mΩ
Basic CJC accuracy 0.25 °C
General
Maximum input 50 V
Offset voltage <2 μV
3-wire internal resistance mismatch <50 mΩ
Basic CJC accuracy 0.6 °C
LAN Function Control and data transfer
Network protocols Ethernet 10/100, TCP/IP
Command protocol SCPI
RS-232 Connector D-sub 9 pin (DE-9)
Baud rates 1200, 2400, 4800, 9600, 19200, 38400
Function Temperature source control output
Dimensions Height 150 mm
Width 245 mm
Depth 385 mm
Weight 6 kg (typical configuration)
Shipping weight 9.5 kg (typical configuration)
Conformity CE, CSA, IEC 61010 3rd edition
Accuracy specifications generally apply with medium and slow sample rates (unless otherwise noted), after a warm-up
time of 1 hour, and within an environment temperature range of 18 °C to 28 °C, and may depend on the channel.
The confidence level for accuracy specifications is 95 % within one year of calibration.
Measurement specications
Scan rate Fast 10 channels per second max (0.1 seconds per channel)
Medium 1 channel per second (1 second per channel)
Slow 4 seconds per channel
Display resolution 4 ½ to 6 ½ digits, depending on function and Sample Rate (see Measurement
Characteristics tables below to find the display resolution of temperature readings)
PRT/RTD
Temperature range 200 °C to 1200 °C (depending on the sensor)
Resistance range 0 Ω to 4 kΩ
Offset compensation 0 Ω to 400 Ω, 4-wire automatic current reversal
400 Ω to 4000 Ω or 3-wire none
Source current reversal interval
(0 Ω to 400 Ω range)
Fast sample rate 2 ms
Medium sample rate 250 ms
Slow sample rate 250 ms
Maximum lead resistance (4-wire Ω) 2.5 % of range per lead for 400 Ω and 4 kΩ ranges.
6 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner
PRT/RTD resistance accuracy
Accuracy is given as % of measurement or ohms, whichever is greater. Basic accuracy is for 4-wire PRT/RTD. When using
3-wire PRT/RTD add 0.013 Ω to the accuracy specification for internal resistance mismatch and voltage offset if using
Channel 1, or add 0.05 Ω if using channels x01 through x20. If the environment temperature is outside the specified range,
multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification.
Range Sample Rate DAQ-STAQ Multiplexer
and Channel 1
High-Capacity Module T.C./ °C Outside
18 °C to 28 °C
0 Ω to 400 Ω
Slow 0.002 % or 0.0008 Ω 0.003 % or 0.003 Ω 0.0001 % or 0.0008 Ω
Medium 0.002 % or 0.002 Ω 0.003 % or 0.003 Ω 0.0001 % or 0.0008 Ω
Fast 0.002 % or 0.005 Ω 0.003 % or 0.006 Ω 0.0001 % or 0.0008 Ω
400 Ω to 4 kΩ
Slow 0.004 % or 0.06 Ω 0.006 % or 0.06 Ω 0.0001 % or 0.008 Ω
Medium 0.004 % or 0.1 Ω 0.006 % or 0.1 Ω 0.0001 % or 0.008 Ω
Fast 0.004 % or 0.18 Ω 0.006 % or 0.18 Ω 0.0001 % or 0.008 Ω
PRT/RTD temperature accuracy
Accuracy is for 4-wire 100 Ω nominal PRT/RTD. When using 3-wire PRT/RTD add 0.039 °C to the accuracy specification
for internal resistance mismatch and voltage offset if using Channel 1, or add 0.15 °C if using channels x01 through x20.
If the environment temperature is outside the specified range, multiply the temperature coefficient number by the tem-
perature deviation and add to the accuracy specification. Linear interpolation may be used between points in the table.
Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor
and characterization.
Sample Rate Temperature DAQ-STAQ Multiplexer and
Channel 1
High-Capacity Module T.C./ °C Outside
18 °C to 28 °C
Slow
−200 °C
0 °C
300 °C
600 °C
0.002 °C
0.005 °C
0.012 °C
0.02 °C
0.008 °C
0.008 °C
0 .018 ° C
0.03 °C
0.002 °C
0.003 °C
0.006 °C
0.01 °C
Medium
−200 °C
0 °C
300 °C
600 °C
0.005 °C
0.005 °C
0.012 °C
0.02 °C
0.008 °C
0.008 °C
0 .018 ° C
0.03 °C
0.002 °C
0.003 °C
0.006 °C
0.01 °C
Fast
−200 °C
0 °C
300 °C
600 °C
0.013 °C
0.013 °C
0.014 °C
0.02 °C
0.015 ° C
0.015 ° C
0.018 °C
0.0 3 °C
0.002 °C
0.003 °C
0.006 °C
0.01 °C
PRT/RTD measurement characteristics
Range Temperature Display Resolution Source Current
Slow / Medium
Sample Rate
Fast
Sample Rate
0 Ω to 400 Ω 0.001 °C 0.01 °C ±1 mA
400 Ω to 4 kΩ 0.001 °C 0.01 °C 0.1 mA
Thermistor
Temperature range 200 °C to 400 °C (depending on the sensor)
Resistance range 0 Ω to 1 MΩ
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 7
Thermistor resistance accuracy
Accuracy is given as ± (% of measurement + Ω). The basic accuracy specification is for 4-wire thermistor, slow sample
rate. When using medium or fast sample rate, add the number given in the table to the accuracy specification. If the
environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature
deviation and add to the accuracy specification. For 2-wire thermistor add 0.02 Ω internal resistance if using Channel 1 or
1.5 Ω if using channels x01 through x20, and add external lead wire resistance.
Range Slow
Sample Rate
Medium
Sample Rate
Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
0 Ω to 2.2 kΩ 0.004 % + 0.2 Ω add 0.3 Ω add 1 Ω 0.0005 % + 0.05 Ω
2.1 kΩ to 98 kΩ 0.004 % + 0.5 Ω add 0.5 Ω add 1.3 Ω 0.0005 % + 0.1 Ω
95 kΩ to 1 MΩ 0.015 % + 5 Ω add 5 Ω add 13 Ω 0.001 % + 2 Ω
Thermistor temperature accuracy
Accuracy specifications are for 4-wire thermistor. When using 2-wire thermistor, add the number given in the table to the
specification for internal resistance. If the environment temperature is outside the specified range, increase the accuracy
specification by 25 % for every 1 °C outside the specified environment temperature range. Specications do not include
sensor accuracy. The practical range of temperature measurement depends on the sensor.
Range Accuracy 2.2 kΩ Thermistor
Slow
Sample Rate
Medium
Sample Rate
Fast
Sample Rate
2-wire
−40 °C 0.001 °C 0.001 °C 0.01 °C add 0.001 °C
0 °C 0.003 °C 0.004 °C 0.01 °C add 0.004 °C
25 °C 0.006 °C 0 . 0 11 ° C 0.02 °C add 0.016 °C
50 °C 0.008 °C 0.018 °C 0.04 °C add 0.05 °C
100 °C 0.047 °C 0 .114 °C 0.28 °C add 0.34 °C
150 °C 0.23 °C 0.56 °C 1.34 °C add 1.7 °C
Range Accuracy 5 kΩ Thermistor
Slow
Sample Rate
Medium
Sample Rate
Fast
Sample Rate
2-wire
−40 °C 0.003 °C 0.004 °C 0.01 °C add 0.001 °C
0 °C 0.002 °C 0.002 °C 0.01 °C add 0.002 °C
25 °C 0.004 °C 0.006 °C 0.01 °C add 0.007 °C
50 °C 0.005 °C 0.009 °C 0.02 °C add 0.022 °C
100 °C 0.022 °C 0.052 °C 0.13 °C add 0.16 °C
150 °C 0.096 °C 0.24 °C 0.57 °C add 0.7 °C
Range Accuracy 10 kΩ Thermistor
Slow
Sample Rate
Medium
Sample Rate
Fast
Sample Rate
2-wire
−40 °C 0.003 °C 0.004 °C 0.01 °C add 0.001 °C
0 °C 0.002 °C 0.002 °C 0.01 °C add 0.002 °C
25 °C 0.003 °C 0.004 °C 0.01 °C add 0.004 °C
50 °C 0.005 °C 0.009 °C 0.02 °C add 0.011 °C
100 °C 0.011 °C 0.024 °C 0.06 °C add 0.067 °C
150 °C 0.04 °C 0.098 °C 0.24 °C add 0.29 °C
8 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner
Thermistor measurement characteristics
Range Temperature Display Resolution Source Current
Slow / Medium
Sample Rate
Fast
Sample Rate
0 Ω to 2.2 kΩ 0.0001 °C 0.001 °C 10 μA
2.1 kΩ to 98 kΩ 0.0001 °C 0.001 °C 10 μA
95 kΩ to 1 MΩ 0.0001 °C 0.001 °C 1 μA
Thermocouple voltage accuracy
Accuracy is given as ± (|% of measurement| + μV). Basic accuracy specification is for medium or slow sample rate. When
using a fast sample rate add the number given in the table to the accuracy specification. If the environment temperature
is outside the specified range, multiply the temperature coefcient numbers by the temperature deviation and add to the
accuracy specification.
Thermocouple reference junction accuracy
Range Accuracy
Channel 1
Ch. x01 – x20 Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
−15 mV to 100 mV 0.004 % + 4 μV add 2 μV add 1 μV 0.0005 % + 0.0005 mV
Module CJC Accuracy T.C./ °C Outside
18 °C to 28 °C
DAQ-STAQ Multiplexer 0.25 °C 0.02 °C
High-Capacity Module 0.6 °C 0.05 °C
Thermocouple
Temperature range 200 °C to 2315 °C (depending on the sensor)
Voltage range 15 mV to 100 mV
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 9
Thermocouple temperature accuracy
Accuracy specifications apply using medium or slow sample rate. When using fast sample rate, increase the accuracy
specification by 25 %. If the environment temperature is outside the specified range, increase the accuracy specification
by 12 % for every 1 °C outside the specified environment temperature range. Accuracy with fixed/external CJC does not
include the accuracy of the reference junction temperature. Linear interpolation may be used between points in the table.
Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor.
Type (Range) Temperature Accuracy
Fixed / External CJC Internal CJC
Channel 1 Ch. x01 – x20 DAQ-STAQ Multiplexer High-Capacity Module
K
270 °C to
1372 °C
−200 °C
0 °C
1000 °C
0.28 °C
0 .10 ° C
0 .14 ° C
0 .41 ° C
0.15 °C
0.20 °C
0.76 ° C
0.29 °C
0.32 °C
1.60 °C
0.62 °C
0.64 °C
T
270 °C to
400 °C
−200 °C
0 °C
200 °C
400 °C
0.27 °C
0 .10 ° C
0.08 °C
0.08 °C
0.40 °C
0 .15 ° C
0 .12 ° C
0 .11 °C
0.76 °C
0.30 °C
0.23 °C
0.20 °C
1.60 °C
0.65 °C
0.47 °C
0.41 °C
R
−50 °C to
1768 ° C
0 °C
300 °C
1200 °C
1600 °C
0.76 °C
0.42 °C
0.33 °C
0.34 °C
1.13 ° C
0.63 °C
0.47 ° C
0.49 °C
1.16 °C
0.64 °C
0.48 °C
0.50 °C
1.28 °C
0 .71 ° C
0.52 °C
0.54 °C
S
−50 °C to
1768 ° C
0 °C
300 °C
1200 °C
1600 °C
0.74 °C
0.45 °C
0.37 °C
0.39 °C
1.11 ° C
0.67 °C
0.54 °C
0.56 °C
1.14 ° C
0.68 °C
0.55 °C
0.57 °C
1.26 °C
0.76 °C
0.60 °C
0.63 °C
J
210 °C to
1200 °C
−200 °C
0 °C
1000 °C
0.20 °C
0.08 °C
0 .11 °C
0.29 °C
0 .12 ° C
0 .14 ° C
0.65 °C
0.28 °C
0.25 °C
1. 41 ° C
0.61 °C
0.53 °C
N
270 °C to
1300 °C
−200 °C
0 °C
500 °C
1000 °C
0.42 °C
0 .15 ° C
0 .12 ° C
0 .14 ° C
0.62 °C
0.23 °C
0 .17 ° C
0 .19 ° C
0.90 °C
0.34 °C
0.24 °C
0.26 °C
1.69 °C
0.64 °C
0.44 °C
0.45 °C
E
270 °C to
1000 °C
−200 °C
0 °C
300 °C
700 °C
0 .17 ° C
0.07 °C
0.06 °C
0.08 °C
0.25 °C
0 .10 ° C
0.09 °C
0 .10 ° C
0.64 °C
0.27 °C
0.21 °C
0.21 °C
1.42 °C
0.61 °C
0.46 °C
0.45 °C
B
100 °C to
1820 °C
300 °C
600 °C
120 0 ° C
16 0 0 ° C
1.32 °C
0.68 °C
0 .41 ° C
0.38 °C
1.97 °C
1.02 °C
0.60 °C
0.55 °C
1.97 °C
1.02 °C
0.60 °C
0.55 °C
1.97 °C
1.02 °C
0.60 °C
0.55 °C
C
0 °C to
2315 °C
600 °C
120 0 ° C
2000 °C
0.23 °C
0.28 °C
0.44 °C
0.33 °C
0.40 °C
0.60 °C
0.37 °C
0.45 °C
0.66 °C
0.54 °C
0.63 °C
0 .91 °C
D
0 °C to
2315 °C
600 °C
120 0 ° C
2000 °C
0.22 °C
0.26 °C
0.39 °C
0.32 °C
0.36 °C
0.53 °C
0.34 °C
0.39 °C
0.56 °C
0.44 °C
0.49 °C
0.69 °C
G
0 °C to
2315 °C
600 °C
120 0 ° C
2000 °C
0.24 °C
0.22 °C
0.33 °C
0.36 °C
0.32 °C
0.46 °C
0.36 °C
0.32 °C
0.46 °C
0.36 °C
0.33 °C
0.46 °C
L
200 °C to
900 °C
−200 °C
0 °C
800 °C
0 .13 ° C
0.08 °C
0.09 °C
0.19 °C
0.12 °C
0.12 °C
0.45 °C
0.28 °C
0.23 °C
0.99 °C
0.62 °C
0.48 °C
M
−50 °C to
1410 °C
0 °C
500 °C
1000 °C
0 . 11 ° C
0.10 °C
0.10 °C
0.16 °C
0.15 °C
0.14 °C
0.30 °C
0.25 °C
0.21 °C
0.64 °C
0.51 °C
0.41 °C
U
200 °C to
600 °C
−200 °C
0 °C
400 °C
0.25 °C
0 .10 ° C
0.08 °C
0.37 °C
0 .15 ° C
0. 11 ° C
0.71 ° C
0.30 °C
0.20 °C
1.48 °C
0.63 °C
0.40 °C
W
0 °C to
2315 °C
600 °C
120 0 ° C
2000 °C
0.24 °C
0.22 °C
0.33 °C
0.36 °C
0.32 °C
0.46 °C
0.36 °C
0.32 °C
0.46 °C
0.36 °C
0.33 °C
0.46 °C
10 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner
Thermocouple measurement characteristics
Range Temperature Display Resolution
Slow / Medium Sample Rate Fast Sample Rate
−270 °C to 2315 °C 0.01 °C 0.1 °C
DC voltage
Maximum input 50 V on any range
Common mode rejection 140 dB at 50 Hz or 60 Hz (1 kΩ unbalance in LOW lead) ± 50 V peak maximum
Normal mode rejection 55 dB for power line frequency ± 0.1 %, ± 120 % of range peak maximum
A/D linearity 2 ppm of measurement + 1 ppm of range
Input bias current 30 pA at 25 °C
DC voltage accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for Channel 1, medium or slow sample
rate. For channels x01 through x20 or when using Fast sample rate, add the numbers given in the table to the accuracy
specification. If the environment temperature is outside the specied range, multiply the temperature coefficient numbers
by the temperature deviation and add to the accuracy specification.
Range Accuracy
Channel 1
Ch. x01 – x20 Fast
Sample Rate
T.C./ °C Outside
18 °C to 28 °C
±100 mV 0.0037 % + 0.0035 % add 2 μV add 0.0008 % of range 0.0005 % + 0.0005 %
±1 V 0.0025 % + 0.0007 % add 2 μV add 0.0008 % of range 0.0005 % + 0.0001 %
±10 V 0.0024 % + 0.0005 % add 0.0008 % of range 0.0005 % + 0.0001 %
±50 V 0.0038 % + 0.0012 % add 0.0008 % of range 0.0005 % + 0.0001 %
DC voltage input characteristics
Range Resolution Input Impedance
Slow / Medium Fast
±100 mV 0.1 μV 1 μV 10 GΩ 1
±1 V 1 μV 10 μV 10 GΩ 1
±10 V 10 μV 100 μV 10 GΩ 1
±50 V 100 μV 1 mV 10 MΩ ±1 %
1 Input beyond ±12 V is clamped. The clamp current is up to 3 mA.
DC current accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for medium or slow sample rate.
When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment temper-
ature is outside the specified range, multiply the temperature coefcient numbers by the temperature deviation and add to
the accuracy specification.
DC current
Input protection 0.15 A resettable PTC
Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 11
Range Accuracy Fast Sample Rate T.C./ °C Outside
18 °C to 28 °C
±100 μA 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 %
±1 mA 0.015 % + 0.0011 % add 0.0008 % of range 0.002 % + 0.001 %
±10 mA 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 %
±100 mA 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 %
DC current input characteristics
Range Resolution Burden Voltage
Slow / Medium Fast
±100 μA 0.1 nA 1 nA <1 mV
±1 mA 1 nA 10 nA <1 mV
±10 mA 10 nA 100 nA <1 mV
±100 mA 100 nA 1 μA <1 mV
Resistance accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for 4-wire resistance, medium or
slow sample rate. For 2-wire resistance add 0.02 Ω internal resistance if using Channel 1, or 1.5 Ω if using channels x01
through x20, and add external lead wire resistance. When using fast sample rate, add the numbers given in the table to
the accuracy specication. If the environment temperature is outside the specified range, multiply the temperature coef-
ficient numbers by the temperature deviation and add to the accuracy specification.
Range Accuracy Fast Sample Rate T.C./ °C Outside
18 °C to 28 °C
100 Ω 0.004 % + 0.0035 % add 0.001 % of range 0.0001 % + 0.0005 %
1 kΩ 0.003 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 %
10 kΩ 0.004 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 %
100 kΩ 0.004 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 %
1 MΩ 0.006 % + 0.001 % add 0.002 % of reading plus
0.0008 % of range 0.0005 % + 0.0002 %
10 MΩ 0.015 % + 0.001 % add 0.002 % of reading plus
0.0008 % of range 0.001 % + 0.0004 %
100 MΩ 0.8 % + 0.01 % add 0.001 % of range 0.05 % + 0.002 %
Resistance input characteristics
Range Resolution Source Current
(open-circuit voltage)
Slow / Medium Fast
100 Ω 0.1 mΩ 1 mΩ 1 mA (4 V)
1 kΩ 1 mΩ 10 mΩ 1 mA (4 V)
10 kΩ 10 mΩ 100 mΩ 100 μA (6 V)
100 kΩ 100 mΩ 1 Ω 100 μA (12 V)
1 MΩ 1 Ω 10 Ω 10 μA (12 V)
10 MΩ 10 Ω 100 Ω 1 μA (12 V)
100 MΩ 100 Ω 1 kΩ 0.1 μA (12 V)
Resistance
Max. lead resistance (4-wire ohms) 10 Ω per lead for 100 Ω and 1 kΩ ranges. 1 kΩ per lead on all other ranges
Fluke Calibration
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke Calibration. Precision, performance, confidence.
For more information call:
In the U.S.A. (877) 355-3225 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40
2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2013 Fluke Calibration.
Specifications subject to change without notice.
Printed in U.S.A. 9/2013 6000790B_EN
Pub-ID 12118-eng
Modification of this document is not permitted
without written permission from Fluke Calibration.
Ordering information
1586A Super-DAQ Precision Temperature Scanner
Super-DAQ and DAQ-STAQ Multiplexer
1586A/1DS Super-DAQ, 1 DAQ-STAQ Multiplexer
1586A/1DS/C Super-DAQ, 1 DAQ-STAQ Multiplexer,
Accredited Calibration
1586A/2DS Super-DAQ, 2 DAQ-STAQ Multiplexers
1586A/2DS/C Super-DAQ, 2 DAQ-STAQ Multiplexers,
Accredited Calibration
Super-DAQ and High-Capacity Module
1586A/1HC Super-DAQ, 1 High-Capacity Module
1586A/1HC/C Super-DAQ, 1 High-Capacity Module,
Accredited Calibration
1586A/2HC Super-DAQ, 2 High-Capacity Modules
1586A/2HC/C Super-DAQ, 2 High-Capacity Modules,
Accredited Calibration
Super-DAQ, High-Capacity Module, and DAQ-STAQ Multiplexer
1586A/DS-HC Super-DAQ, 1 High-Capacity Module, 1 DAQ-STAQ Multiplexer
1586A/DSHC/C Super-DAQ, 1 High-Capacity Module, 1 DAQ-STAQ Multiplexer,
Accredited Calibration
Accessories
1586-2586 High-Capacity Module without Relay Card
1586-2586-KIT High-Capacity Module with Relay Card
1586-2588 DAQ-STAQ Multiplexer without Adapter Card
1586-2588-KIT DAQ-STAQ Multiplexer, Adapter Card, Interface Cable
1586-2588-CBL DAQ-STAQ Multiplexer Interface Cable
Y1586S Rack Mount Kit, Single (half rack)
Y1586D Rack Mount Kit, Dual (full rack)
1586-CASE Super-DAQ Carrying Case (mainframe and internal modules)
1586/DS-CASE Super-DAQ/DAQ-STAQ Carrying Case (mainframe and external module)