USPDCR-4000 - PDS-A109 - 4/99 3
cd
Pressure Measurement
Specification
Operating Pressure Ranges
PDCR 4000
1, 2 psi gauge
5, 10, 15, 20, 30, 50, 100, 200, 300, 500, and 900
psi gauge or absolute
1000, 2000, 3000, 5000, 7500, and 10,000 psi
sealed gauge or absolute
PDCR 4100
1, 2, 5, 10, 15, 20, 25, 30, 50, 75, 100, 200, 300,
and 500 psi differential
Other pressure units can be specified e.g. millibar,
bar, kg/cm2, KPa, etc.
Overpressure
The operating pressure range may be exceeded
by the following multiples with negligible effect
on calibration.
PDCR 4000
Gauge and absolute:
10 x for ranges 1 to 2 psi
6 x for range 5 psi
4 x for ranges 10 psi to 900 psi up to maximum
of 2000 psi
2 x for range 1000 through 10,000 psi to a
maximum of 15,000 psi
PDCR 4100
Differential (positive side):
10 x for ranges 1 and 2 psi
6 x for range 5 psi
4 x for ranges 10 to 300 psi
3 x for range 500 psi
Differential (negative side):
6 x for ranges 1, 2 psi
4 x for range 5 psi
2 x for ranges 10 to 75 psi
150 psi for ranges 100 to 500 psi
This overpressure capability can be further
improved by selecting a range higher than
required and operating with a lower output.
Pressure Containment
Gauge and differential (positive side):
12 x for ranges up to 5 psi
6 x for ranges 10 to 900 psi (3000 psi maximum)
Differential (negative side):
8 x for ranges up to 5 psi
4 x for ranges 10 to 500 psi
200 psi maximum for ranges 100 to 500 psi
Sealed gauge and absolute:
3000 psi for ranges up to 900 psi
20,000 psi for ranges 1000 psi and above
Pressure Media
Fluids compatible with Hastelloy C276 and
stainless steel 316L.
Line Pressure
1000 psig maximum
Excitation Voltage
10 Volts at 5 mA nominal.
For pulse powered operation, the recommended
power-on time is 10ms before sample.
Output is ratiometric to supply within the
following limits:
1 V to 12 V for ranges up to 900 psi
5 V to 12 V for ranges 1000 psi and above
Output Voltage
(based on 10Vdc excitation)
50 mV for ranges 1, 2 and 5 psi
100 mV for 10 psi and above
Transducers with ranges up to 900 psi can be
over ranged 2 x Full Scale to provide up to
200mV output. Linearity is slightly degraded
but stability is improved.
For higher outputs up to 10 V refer to PMP 4000
series datasheet.
Common Mode Voltage
Typically +3.5 V to +9 V with respect to the –ve
supply at 10 V excitation.
Output Impedance
2 kê nominal.
Load Impedance
Greater than 100 k ê for quoted performance.
Performance Specification
Accuracy
Combined effects of Non-linearity, Hysteresis and
Repeatability.
Standard : ± 0.08% FS BSL maximum
Option (A) : ± 0.04% FS BSL maximum
Higher accuracies can be selected from
the core stock database.
Zero Offset & Span Setting
Zero: ±3 mV maximum
Span :±10 mV maximum. Units of the
same range are matched to closer
than ±3 mV.
Improved settings are available where
interchangeability is critical.
Option (D): Consult factory
Stability
± 0.1% FS typically per annum
Long term stability is improved by using a
lower pressure range in the overrange
condition at a reduced excitation voltage.
Operating Temperature Range
-5° to +175°F standard
-65 to +250°F available
Temperature Effects
Standard: ± 0.3% FS TEB over 32 to 122°F
± 1.0% FS TEB over -5 to 175°F
For ranges up to 5 psi multiply x3
Improved Temperature Error Band (TEB) can
be selected from the core stock database.
Acceleration Sensitivity
Typically 0.02% FS/g for 5 psi
decreasing to 0.0003% FS/g for ranges
above 900 psi, along the sensitive axis.
Mechanical Shock
1000g, 1ms half sine pulse in each of 3
mutually perpendicular axes will not affect
performance.
Vibration
Response less than 0.05% FS/g at 30g
peak 10Hz to 2kHz, limited by 12 mm double
amplitude, (MIL-STD 810C Proc 514.2-2
Curve L)
Physical Specification
Pressure Connection
1 psi to 900 psi ranges
1/4” NPT Female
1/8” NPT Male w/Bulkhead Mount - standard
1/4” NPT Male
1/4” Tube Swagelok w/Bulkhead Mount
MS 33656-4 (1/4 AN)
Others available on request
1000 to 10,000 psi ranges: 1/4” NPT Female -
standard
Adaptors available on request
Weight
4.5 oz nominal (1 to 900 psi)
6 oz nominal (1000 to 10,000 psi)
7 oz nominal for differential types
Electrical Connection
A wide range of cable and connector versions are
available. Refer to ordering information and
installation drawings overleaf.
Options
(A) Improved accuracy ± 0.04% FS BSL.
(B) Internal Shunt Calibration (ranges up to 900
psi). Connecting an external link results in a
positive span shift of 80% ±5% FS.
(Not available for PDCR 4X2X, PDCR 4X7X).
(C) Mating electrical connector (PDCR 4X6X).
(D) Improved zero and span settings
(consult factory).
(E) Negative calibration
Accessories
A traceable calibration certificate with
installation notes is supplied as standard.
Calibration Standards
Pressure transducers manufactured by Druck are
calibrated against precision pressure calibration
equipment which is traceable to the National
Institute of Standards and Technology (NIST).
Druck is an ISO 9001
registered company.
Continuing development sometimes necessitates
specification changes without notice.