SPECIFICATION
MPS-45-3.3 MPS-45-5 MPS-45-7.5 MPS-45-12 MPS-45-13.5 MPS-45-15 MPS-45-24 MPS-45-27 MPS-45-48
MODEL
DCVOLTAGE
RATEDCURRENT
CURRENTRANGE
RATEDPOWER
OUTPUT
LINEREGULATION
LOADREGULATION
SETUP,RISETIME
HOLDUP TIME (Typ.)
VOLTAGERANGE
FREQUENCY RANGE
EFFICIENCY(Typ.)
INPUT
INRUSHCURRENT (Typ.)
LEAKAGECURRENT
WORKINGHUMIDITY
WORKINGTEMP.
STORAGETEMP.,HUMIDITY
TEMP.COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
OVERVOLTAGE
ACCURRENT (Typ.)
3.3V 5V 7.5V 12V 13.5V 15V 24V 27V 48V
8A 8A 5.4A 3.7A 3.3A 3A 1.9A 1.7A 1A
0~10.7A 0~10.5A 0~7A 0~4.4A 0~3.9A 0~3.5A 0~2.2A 0~1.95A 0~1.1A
26.4W 40W 40.5W 44.4W 44.6W 45W 45.6W 45.9W 48W
52W(+3.3V:35W)with18CFMmin.Forcedairconvection
80mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p
3.0% 3.0% 3.0% 2.0%
2.0%
2.0% 2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0%
3.0% 3.0% 3.0% 2.0%
90~264VAC127~370VDC
47~440Hz
65% 72% 75% 76% 76% 77% 78% 78% 78%
1.2A/115VAC0.7A/230VAC
COLDSTART 17A/115VAC35A/230VAC
<0.3mA /264VAC
53~75W(3.3V:36~55W) ratedoutputpower
Protectiontype: Hiccupmode,recoversautomaticallyafterfaultconditionisremoved
Protectiontype: Hiccupmode,recoversautomaticallyafterfaultconditionisremoved
-20~+85 ,10~95%RH
0.04%/ (0~50
10~500Hz,2G10min./1cycle,periodfor60min. eachalongX, Y,Zaxes
366.1Khrsmin.MIL-HDBK-217F(25 )
127*76*28mm(L*W*H)
0.18Kg;72pcs/15.1Kg/1.35CUFT
FileName:MPS-45-SPEC2010-10-18
800ms,30ms/230VAC1200ms,30ms/115VACatfullload
50ms/230VAC16ms/115VACatfullload
-10~+60 (Refertooutputloadderatingcurve)
20~90%RHnon-condensing
SAFETY STANDARDS
HARMONICCURRENT
EMSIMMUNITY
UL2601-1, TUVEN60601-1,IEC60601-1approved
CompliancetoEN55011(CISPR11)ClassB
Complianceto EN61000-3-2,-3
Complianceto EN61000-4-2,3,4,5,6,8,11;ENV50204,EN60601-1-2,medicallevel,criteria A
45WSingleOutputMedical Type MPS-45 serie s
Universal AC input / Full range
Low leakage current 0.3mA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
Features :
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
ENVIRONMENT
SAFETY &
EMC
(Note4)
PROTECTION
EMICONDUCTION&RADIATION
WITHSTANDVOLTAGE
ISOLATIONRESISTANCE
I/P-O/P:4KVACI/P-FG:1.5KVACO/P-FG:0.5KVAC
I/P-O/P,I/P-FG:100MOhms/500VDC /25 /70%RH
RIPPLE&NOISE(max.) Note.2
VOLTAGETOLERANCE Note.3
OUTPUTPOWER(max.)
3.8~4.46V 5.75~6.75V 8.63~10.1V 13.8~16.2V 15.5~18.2V
17.25~20.25V
27.6~32.4V 31~36.45V 55.2~64.8V
VOLTAGE ADJ.RANGE 2.97~3.63V 4.5~5.5V 6.75~8.25V 10.8~13.2V 12.2~14.85V 13.5~16.5V 21.6~26.4V 24.3~29.7V 43.2~52.8V
EN60601-1
IEC60601-1
http://store.iiic.cc/