(R) FOM/FOS Fiber Optic Meter Fiber Optic Source Service Information Warning To avoid injury: do not service the FOM or FOS unless you are qualified to do so. The service information provided in this document is for the use of qualified personnel only. Caution The FOM and FOS contain parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. Introduction Refer to the Instruction Sheet for the complete warranty statement. This Service Information provides the following service information for the Fiber Optic Meter (hereafter referred to as the FOM) and the Fiber Optic Source (hereafter referred to as the FOS). To order parts, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call: * * * * * * * * * Parts and warranty service information Specifications Schematic Cleaning procedures Required equipment Performance tests Procedures for disassembling and reassembling the FOM and FOS Calibration adjustments Parts and accessories lists For operating instructions, refer to the Fiber Optic Meter/Fiber Optic Source Instruction Sheet. Parts and Warranty Service The FOM and FOS are warranted to be free from defects in material and workmanship for one year, while under normal use. Parts and repairs are warranted for 90 days. 800-44FLUKE (800-443 5853) in the U.S.A. and Canada. (31 40) 267 8200 in Europe + 1 206-356-5500 from other countries Address correspondence to: Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. Fluke Europe B.V. P.O. Box 1186 5602 B.D Eindhoven The Netherlands Specifications Accuracy is specified for a period of one year after calibration, at 18C to 28C (64F to 82F) with relative humidity to 75%. Specifications are shown in Tables 1 and 2. PN 617891 January 1997 1996 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies. FOM/FOS Service Information Table 1. FOM Specifications Output Input Connector Type Photodetector Type Application range Calibrated Wavelengths Acceptable Fiber Types Operating Range Maximum Power Level Absolute Accuracy Relative Accuracy Repeatability Battery Life 1 mV per 1 dB Fixed ST Germanium 800 to 1600 nm 850, 1300, and 1550 nm +3 to -50 dBm +5 dBm 0.25 dB (specified at 25C -10 dBm per NIST standard) 0.15 dB, (specified over any 10 dB within measurement range) 0.04 dB Low Battery Indication Operating Temperature Storage Temperature -20 to +70C Humidity 0 to 40C up to 75% RH Table 2. FOS Specifications Type Wavelength Output power Output connector type Beam divergence Pulse duration Maximum output Stability Temperature coefficient Battery Life 2 Infrared LED 850 30 nm 1300 -40/+50 nm 850/1300 30 nm, -40/+50 nm -20 dBm, nominal into 62.5/125 micron multimode fiber Fixed ST 0.3 radians Continuous wave 200 W (radiated into free space) 0.2 dB per 8 hours at 20C after 20 minute warmup -0.08 dB per C, < 18C or > 20C Low Battery Indication Operating Temperature 9V alkaline, NEDA 1604A or IEC 6LR61 16 hours minimum, 24 hours typical, 9V alkaline Blinking LED indicator 0 to +40C Storage Temperature -20 to +70C Humidity 0 to 40C up to 75% RH Battery Type CLASS I LED PRODUCT The FOS is rated as a class 1 LED source per IEC 825. 9/125 to 100/140 m 9V alkaline, NEDA 1604A or IEC 6LR61 16 hours minimum, 100 hours typical, 9V alkaline Blinking LED indicator 0 to +40C Battery Type Safety Information General Description The FOM measures optical power on fiber optic cables. It indicates power loss on tested cables using any digital multimeter (DMM) or graphical multimeter (GMM) that has a 10 Me input impedance, standard diameter banana jacks, and mVdc capability. The FOS is used as a light source with the FOM or other fiber optic meters. Low Battery Detection For both the FOM and FOS, a steady, green LED indicates a good battery. A blinking LED indicates a low battery that needs to be replaced. Cleaning the Optical Connector Most problems with optical power meters and light sources result from contaminated connectors. Therefore, always clean the connector before troubleshooting or calibration. To clean the FOM or FOS ST connector, wipe the internal portion gently with an optical-grade swab dampened with optical-grade alcohol. To remove loose dirt and dust from the connector, use filtered, compressed air. Always cover the connector with a dust cap when the unit is not in use. Equipment Required The following equipment is required for servicing the FOM: * * * * * * * 0-10V power supply Digital multimeter Optical power meter calibrated at 850, 1300, and 1550 nm traceable to NIST Standards 850 nm laser source; Po = -10 dBm 1300 nm laser source; Po = -10 dBm 1550 nm laser source: Po = -10 dBm Variable optical attenuator Fiber Optic Meter/Source Performance Tests * 2 ST/ST multimode fiber optic patch cables The following equipment is required for servicing the FOS: * Calibrated FOM * Digital multimeter * One ST/ST multimode fiber optic patch cable Performance Tests Use the following performance tests to confirm that the FOM or FOS is working properly. If the FOM or FOS fails any of these tests, it needs calibration adjustments or repair. FOM -10 dBm Checkout 1. Turn on the FOM and the three light sources. Allow the FOM to warm up for 10 minutes. Allow the sources to warm up according to the manufacturer's specifications. Perform the remaining steps using the light sources in the following order: (1) 850 nm, (2) 1550 nm, and (3) 1300 nm. 2. Connect the light source to the reference power meter using a fiber optic patch cable. 3. Adjust the output of the source for a power meter reading of -10.00 0.05 dBm. Record this reading as Po (power out) to use in step 5. 4. Set the FOM switch to the wavelength position being tested. 5. Connect the FOM to the light source using the patch cable from step 2. Allow a few seconds for the reading to stabilize; then verify that the reading is Po 0.25 dBm. 4. Allow a few seconds for the reading to stabilize; then verify that the reading is Po 0.4 dBm. 5. Turn off the DMM, FOM, and the light sources. Disconnect the FOM from the DMM. Disconnect the fiber optic patch cables from the attenuator, source, and FOM. FOS Checkout 1. Connect the FOS to the calibrated FOM with a patch cable. Turn the FOS on to the wavelength being tested and let it stabilize for 20 minutes. 2. Connect the FOM to a DMM with at least 0.1 mV resolution. Turn the FOM on and set the switch to the wavelength being tested. 3. Verify that the FOS output is -20 0.5 dBm (-19.5 to -20.5 mV on the DMM). 4. For a dual source, repeat steps 1 through 3 for each wavelength. Disassembling and Reassembling the FOM and FOS To perform calibration adjustments for the FOM and the FOS, you must disassemble them. When you finish the adjustments, reassemble them. A Phillips-head screwdriver is required for disassembling and reassembling the FOM and FOS. To disassemble the FOM or FOS, proceed as follows: 1. Set the FOM or FOS switch to OFF. 6. Disconnect the light source from the FOM. 2. Remove the battery door; then disconnect the battery. FOM -50 dBm Checkout 3. Remove the two Phillips screws found under the battery door; then separate the top and bottom cases. 1. Connect the 1300 nm light source to the optical attenuator. Connect the attenuator to the reference power meter using a second fiber optic patch cable. 2. Set the optical attenuator for a power meter reading of -50.00 0.05 dBm. Record this reading as Po to use in step 4. 3. Connect the FOM to the attenuator using the patch cable connected to the reference power meter. 4. To remove the pca, remove the Phillips screw near SW1; then lift out the pca. To reassemble the FOM or FOS, proceed as follows: 1. Reinsert the pca; then replace the screw that holds the pca. 2. Tuck the battery wires into the notch at the side of the battery compartment. 3 FOM/FOS Service Information 3. Tuck the battery wires down against the pca so they will not interfere with SW1. 4. Set SW1 to the off position. 5. Set the slide switch actuator on the top case to the OFF position. For the FOS, place the lanyard into the notch in the bottom case. 6. Put the top and bottom cases together; then replace the two screws under the battery door. Calibration Adjustments for the FOM To ensure that the FOM performs to specifications, calibrate it annually using the procedure that follows. Always perform the complete calibration adjustment procedure. Preparing for Calibration To prepare for calibration, proceed as follows: 1. Remove the dust cap from the FOM. 2. Connect a voltage supply of 8V 1.0V dc to the FOM battery strap. 3. Set the FOM switch to the 1300 nm position. Allow the FOM to warm up for 10 minutes. Note For the remaining calibration adjustments, use the cable that was used in step 1 for connecting the attenuator to the reference power meter or the FOM. To reduce errors, limit the movement of the cable when making connections between the attenuator and the reference power meter or the FOM. To limit cable movement, coil and tape the excess cable to the work surface. 2. Connect the attenuator output to the FOM. 3. Measure Vdet between U1 pin 7 and the analog ground reference at U1 pin 5 with the DMM. (See Figure 1.) 4. Adjust R2 for an output on the DMM of 0.23 0.01V dc. 1300 nm Fine Calibration Adjustment 1. Connect the 1300 nm light source to the attenuator input. Connect the attenuator output to the reference power meter. Set the attenuator for 0 dB attenuation. Set the 1300 nm light source for a power meter reading of -10 0.05 dBm. Record this reading as Po to use in step 3. 2. Connect the attenuator output to the FOM. Connect the FOM to the DMM. 4. Allow all sources to warm up according to the manufacturer's specifications. 3. Adjust R10 for an output on the DMM of Po 0.05 dBm. Coarse Calibration Adjustment 4. Connect the attenuator output to the reference power meter. Set the attenuator for a power meter reading of -45 0.05 dBm. Refer to the FOM drawing (Figure 1) for the locations of measurements and components. 1. Connect the 1300 nm light source to the attenuator input. Connect the attenuator output to the reference power meter. Set the attenuator for 0 dB attenuation. Set the 1300 nm light source for a power meter reading of -10 0.05 dBm. Record this reading as Po to use in step 6. 5. Connect the attenuator output to the FOM. 6. Adjust R2 for an output on the DMM of Po 0.05 dBm. 7. Repeat steps 1 through 6 until the FOM's -10 dBm and -45 dBm readings are within specifications. 8. Apply insulating varnish, such as red GLPT, to the R10 and R2 potentiometers. 4 Fiber Optic Meter/Source Calibration Adjustments for the FOS 850 nm Calibration Adjustment 1. Set the switch on the FOM to the 850 nm position. 2. Connect the light source to the reference power meter. Set the 850 nm light source for a power meter reading of -10 0.05 dBm. Record this level as Po to use in step 4. 3. Connect the 850 nm light source to the FOM. Connect the FOM to the DMM. 4. Adjust R1 for an output on the DMM of Po 0.05 dBm. 5. Apply insulating varnish, such as red GLPT, to the R1 potentiometer. 3. For 850 nm calibration, adjust R21 until the output on the DMM reads -20 0.1 mV. For 1300 nm calibration, adjust R20 until the output on the DMM reads -20 0.1 mV. 4. Apply insulating varnish, such as red GLPT, to each potentiometer adjusted. Parts and Accessories Tables 3 and 4 contain replacement parts for the FOM and FOS that are available from Fluke. Table 5 lists available accessories. Table 3. Fluke Replacement Parts for the FOM Description 1550 nm Calibration Adjustment 1. Set the switch on the FOM to the 1550 nm position. 2. Connect the light source to the reference power meter. Set the 1550 nm light source to -10 0.05 dBm. Record this reading as Po to use in step 4. FOM Plastic Case Assembly 617339 Battery Door 200474 Slide Switch Actuator 200482 PD1: Germanium Photodiode, 1 mm 602788 Table 4. Fluke Replacement Parts for the FOS 3. Connect the 1550 nm light source to the FOM. Connect the FOM to the DMM. 4. Adjust R3 for an output on the DMM of Po 0.05 dBm. Fluke Part Number Description Fluke Part Number FOS-850 Plastic Case Assembly 617354 FOS-1300 Plastic Case Assembly 617388 FOS-850/1300 Plastic Case Assembly 617412 Battery Door 200474 Slide Switch Actuator 200482 Verifying Calibration 850 LED 602755 Follow the procedure under "Performance Tests" earlier in this document. 1300 LED 602760 5. Apply insulating varnish, such as red GLPT, to the R3 potentiometer. At this point, the calibration adjustment is complete. Table 5. Accessories for the FOM and FOS Calibration Adjustments for the FOS Refer to the FOS drawing (Figure 2) for the locations of components. 1. Connect the FOS to the calibrated FOM with a patch cable. Turn the FOS on and let it stabilize for 20 minutes. 2. Connect the FOM to a DMM with at least 0.1 mV resolution. Turn the FOM on and set it to the desired wavelength. Description Fluke Part Number ST to ST Multimode Patch Cable, 1m 617453 ST to ST Mating Sleeve 602810 ST to FC Patch Cable 617487 ST to SC Patch Cable 617511 ST to SMA Patch Cable 616545 5 FOM/FOS Service Information J1 (Red) J2 (Black) U1 Pin 7, (Vdet) U1 Pin 5, (Analog GND) FPM1-00-0100 eq1f.eps Figure 1. FOM Calibration Measurement and Adjustment Points 6 Fiber Optic Meter/Source Calibration Measurement and Adjustment Points J1 (Red) J2 (Black) 850 Source Components 1300 Source Components FLS1-01-0100 eq2f.eps Figure 2. FOS Calibration Adjustment Points 7 FOM/FOS Service Information FPM1-00-0300 eq4f.eps Figure 3. Fiber Optic Meter Schematic 8 cut cut Fiber Optic Meter/Source Fiber Optic Meter Schematic FLS1-00-0300 eq3f.eps Figure 4. Fiber Optic Source Schematic 9 FOM/FOS Service Information 10