GENH Series
Programmable DC Power Supplies
750W in a 1U half-rack size
Built in RS-232 & RS-485 Interface
Advanced Parallel Standard
Optional Interfaces:
IEEE488.2 SCPI (GPIB)
Isolated Analog Programming
Compliant LAN
The GenesysTM family of programmable power supplies sets a new standard for flexible, reliable,
AC/DC power systems in OEM, Industrial and Laboratory applications.
Front Panel Description
1. AC ON/OFF Switch
2. Air Intake allows zero stacking for maximum system flexibility and power density.
3. Reliable encoder controls Output Voltage and sets Address.
4. Volt Display shows Output Voltage and directly displays OVP, UVL and Address settings.
5. Reliable encoder controls Output Current, sets baudrate and Advanced Parallel mode.
6. Current Display shows Output Current and displays baudrate.
7. Function/Status LEDs:
8. Pushbuttons allow flexible user configuration
Coarse and Fine adjustment of Output Voltage/Current and Advanced Parallel Master or Slave
Preview settings and set Voltage/Current with Output OFF, Front Panel Lockout
Set OVP and UVL Limits
Set Current Foldback
Local/Remote Mode and select Address and Baudrate
Output ON/OFF and Auto-Start/Safe-Start Mode
Alarm
Fine Control
Preview Settings
Foldback Mode
Remote Mode
Output On
Features include:
Worldwide Safety Agency Approvals; CE Mark for LVD and EMC Regulation
Genesys™ GENH750W-1U
1
High Power Density available:
750W in 1U half-rack size.
Wide Range Input (85 - 265Vac Continuous)
Active Power Factor Correction (0.99 typical)
Output Voltage up to 600V, Current up to 100A
Built-in RS-232/RS-485 Interface
Front Panel Lockout
Last-Setting Memory
High Resolution 16 bit ADCs & DACs
Reliable Encoders for Voltage and Current Adjustment
Constant Voltage/Constant Current auto-crossover
Adv anced Parallel reports total current up to f our identical units
Global Commands for Serial RS-232/RS-485 Interface
Independent Remote ON/OFF and Remote
Enable/Disable
External Analog Programming and Monitoring
Reliable Modular and SMT Design
19" Rack Mounted ATE benchtop and OEM applications
Side-by-side mounting of two units in a 19” rack
Optional Interfaces
Isolated Analog Programming and Monitoring Interface
(0-5V/0-10V & 4-20mA)
IEEE 488.2 SCPI (GPIB) Multi-Drop
Compliant LAN
LabView® and LabWindows® drivers
Five Year W arranty
1. Remote/Local Output Voltage Sense Connections.
2. DIP Switches select 0-5V or 0-10V Programming and other functions.
3. DB25 (Female) connector allows (Non-isolated) Analog Program and Monitor and other functions.
4. RS-485 OUT to other Genesys™ Power Supplies.
5. RS-232/RS-485 IN Remote Serial Programming.
6. Wide-Range Input 85-265VAC continuous, 47/63Hz with Active Power Factor Correction (0.99 typical)
AC Input Connector: IEC320.
7. Exit air assures reliable operation when zero stacked.
8. Output Connections: Rugged busbars for 6V up to 60V Output; Connector for Outputs >60V.
9. Optional Interface Position for IEEE 488.2 SCPI (shown) or Isolated Analog Interface or LAN Interface .
Rear Panel Description
Test and Measurement
Last-Setting memory simplifies test design and requires
no battery backup.
Built-in RS-232/RS-485 gives maximum system flexibility
along with 0-5V and 0-10V, selectable analog
programming.
Wide range of available inputs allows testing of many
different devices.
Semiconductor Burn-in
Safe-Start may be ENABLED to re-start at Output OFF
to protect load.
Wide range input (85-265Vac) with Active Power Factor
correction rides through input transients easily.
Component Test
High power density, zero stacking and single wire parallel
operation give maximum system flexibility.
Laser Diode
OVP is directly set on Voltage Display, assuring accurate
protection settings.
Current Limit Fold Back assures load is protected from
current surges.
Heater Supplies
Smooth, reliable encoders enhance front panel control.
Remote analog programming is user selectable 0-5V or
0-10V.
RF Amplifiers and Magnets
Robust design assures stable operation under a wide
variety of loads.
High linearity in voltage and current mode.
Applications
Genesys™ power supplies have been designed to meet the demands of a wide variety of applications.
2
3
*1: Minimum voltage is guaranteed to maximum 0.2% of Vo Rated.
*2: Minimum current is guaranteed to maximum 0.4% of Io Rated
*3: At maximum output power.
*4: 85~132Vac or 170~265Vac, constant load.
*5: From No-load to Full-load, constant input voltage.
*6: For load voltage change, equal to the unit voltage rating, constant input voltage.
*7: For 6V models the ripple is measured at 2~6V output voltage and full output current. For other models, the ripple is measured at 10~100% output voltage and full output current.
*8: Time for the output voltage to recover within 0.5% of its rated for a load change 10~90% of rated output current, Output set-point:10~100%.
*9: For 6V~300V models: measured with JEITA RC-9131A 1:1 probe. For 600V model: measured with 10:1 probe
Accuracy -Values have been calculated at Vo Rated & Io Rated
Genesys ™ GENH750W Specifications
1.0 MODEL
1.Rated output voltage (*1)
2.Rated Output Current (*2)
3.Rated Output Power
4.Efficiency at 100/200Vac (*3)
1.1 CONSTANT VOLTAGE MODE
1.Max.line regulation ( 0.01% of Vo+ 2mV )(*4)
2.Max load regulation ( 0.01% of Vo+2mV )(*5)
3.Ripple and noise p-p 20MHz (*9)
4.Ripple r.m.s 5Hz~1MHz (*9)
5.Remote sense compensation/line
6.Temp. coefficient
7.Up-prog. response time, 0~Vo Rated
8.Down-prog response time full-load
9.Down-prog response time no-load
10.Transient response time (*8)
1.2 CONSTANT CURRENT MODE
1.Max.line regulation (0.01% of Io+ 2mA)(*4)
2.Max.load regulation (0.02% of Io+5mA)(*6)
3.Ripple r.m.s 5Hz~1MHz . (*7)
4.Temp. coefficient
1.3 PROTECTIVE FUNCTIONS
1. OCP
2. OCP Foldback
3. OVP type
4. OVP trip point
5. Over Temp. Protection
2.Display
3.Indications
1.5 FRONT PANEL
1.Control functions
1.4 ANALOG PROGRAMMING AND MONITORING
1.Vout Voltage Programming
2.Iout Voltage Programming
3.Vout Resistor Programming
4.Iout Resistor Programming
5.On/Off control (rear panel)
6.Output Current monitor
7.Output Voltage monitor
8.Power Supply OK signal
9. CV/CC indicator
10. Enable/Disable
11. Local/Remote analog control
2. Remote Current Programming (16 bit)
Resolution (0.012% of Io Rated)
Accuracy (0.1% of Io Rated+0.1% of Io Actual Output)
1.6 Interface RS-232&RS-485 or Optional GPIB / LAN Interface
Model
1. Remote Voltage Programming (16 bit)
Resolution (0.012% of Vo Rated)
Accuracy (0.05%Vo Rated+0.05% of Vo Actual Output)
4. Readback Current
Resolution (0.012% of Io Rated )
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output)
3. Readback Voltage
Resolution (0.012% of Vo Rated)
Accuracy (0.1%Vo Rated+0.1% of Vo Actual Output)
5. OVP/UVL Programming
Resolution (0.1% of Vo Rated)
Accuracy (1% of Vo Rated)
600-1.3
600
1.3
780
83/87
62
62
300
60
5
4000
250
250
2.13
5.26
8
5~660V
0.16
2.6
600
72
600
72
5.2
1200
0.16
600
6000
300-2.5
300
2.5
750
83/87
32
32
150
25
5
2500
2.25
5.5
13
5~330V
0.30
5.0
300
36
300
36
10
600
0.30
300
3000
5~165V
0.60
10
150
18
150
18
20
300
0.60
150
1500
150-5
150
5
750
83/87
17
17
100
10
5
150
2000
2.5
6.0
18
5~110V
0.90
15
100
12
100
12
30
200
0.90
100
1000
100-7.5
100
7.5
750
83/87
12
12
80
8
5
1500
2.75
6.5
23
5~88V
1.14
19
80
9.6
80
9.6
38
160
1.14
80
800
80-9.5
80
9.5
760
83/87
10
10
80
8
4
150mS , N.L/F.L , resistive load
1200
2.95
6.9
29
5~66V
1.50
25
60
7.2
60
7.2
50
120
1.50
60
600
60-12.5
60
12.5
750
83/87
8
8
60
8
3
1100
3.25
7.5
38
2~44V
2.28
38
40
4.80
40
4.80
76
80
2.28
40
400
40-19
40
19
760
83/87
6
6
60
8
2
80
1000
3.9
8.8
48
2~36V
3.0
50
30
3.60
30
3.60
100
60
3.0
30
300
30-25
30
25
750
82/85
5
5
60
8
1.5
900
4.5
10
63
1~24V
4.56
76
20
2.40
20
2.40
152
40
4.56
20
200
20-38
20
38
760
82/85
4
4
60
8
1
800
5.8
12.6
76
1~15V
7.2
120
12.5
1.50
12.5
1.50
240
25
7.2
12
125
12.5-60
12.5
60
750
81/84
3.3
3.3
60
8
1
50
700
8.0
17
120
0.5~10V
10.8
180
8
0.96
8.0
0.96
360
16
10.8
8
80
8-90
8
90
720
78/81
2.8
2.8
60
8
1
600
11
23
180
0~105% Constant Current
Output shut down when power supply change from CV to CC. User selectable.
Inverter shut-down, manual reset by AC input recycle or by OUT button or by communication port
0.5~7.5V
User selectable , latched or non latched
Voltage
Current
Voltage, Current, Alarm, Fine, Preview, Foldback, Local, Output On, Front Panel Lock
4 digits , accuracy: 0.5%+/-1 count
4 digits, accuracy: 0.5%+/-1 count
Vout/ Iout manual adjust by separate encoders (coarse and fine adjustment selectable)
OVP/UVL manual adjust by Volt. Adjust encoder
Address selection by Voltage (or current) adjust encoder. Number of addresses:31
AC on/off, Output on/off, Re-start modes (auto, safe), Foldback control (CV to CC), Go to local control
RS232/485 and IEEE488.2 selection by IEEE enable switch and DIP switch
Baudrate selection: 1200,2400,4800,9600 and 19,200
Front Panel Lock
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:+/-0.5% of rated Vout.
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:+/-1% of rated Iout.
0~100%, 0~5/10Kohm full scale,user select.,Accuracy and linearity:+/-1% of rated Vout.
0~100%, 0~5/10Kohm full scale,user select. Accuracy and linearity:+/-1.5% of rated Iout.
By electrical. Voltage: 0~0.6V/2~15V,or dry contact ,user selectable logic
0~5V or 0~10V , accuracy:1% , user selectable
0~5V or 0~10V ,accuracy:1% ,user selectable
TTL High=OK, 0V-Fail 500ohm impedance
CV: TTL high (4~5V) source: 10mA, CC: TTL low (0~0.6V) sink current:10mA
Dry contact. Open:off , Short: on. Max. voltage at Enable/Disable in: 6V
By electrical signal or Open/Short: 0~0.6V or short: Remote analog, 4~5V or open: Local.
12
200
6
0.72
6.0
0.72
400
12
12
6
60
6-100
6
100
600
76/78
2.6
2.6
60
8
1
100PPM/°C of rated output voltage,following 30 minutes warm up
80mS , N.L/F.L , resistive load
10
500
Less than 1mSec for models up to and including 100V. 2msec for models above 100V
12
25
200
100
PPM/°C
from rated output current, following 30 minutes warm up
mA
mA
V
mV
mV
mA
mA
mV
mV
mV
mV
GENH
V
A
W
%
mV
mV
mV
mV
V
PPM/°C
mS
mS
mS
mA
mA
mA
PPM/°C
12. Local/Remote analog control indicator Open collector, Local: Open, Remote: On. Maximum voltage: 30V, maximum sink current: 5mA.
General Specifications Genesys™ GENH750W
*1: For cases where conformance to various safety standards (UL, IEC etc.) is required, to be described as 100-240Vac (50/60Hz).
All specifications subject to change without notice.
2.1 INPUT CHARACTERISTICS
1. Input voltage/freq. (*1)
2. Power Factor 0.99 @100/200Vac, rated output power.
3. EN61000-3-2,3 compliance Complies with EN61000-3-2 class A and EN61000-3-3 at 20~100% output power.
4. Input current 100/200Vac
5. Inrush current 100/200Vac Less than 25A,
6. Hold-up time More than 20mS , 100Vac , at 100% load.
10.5A / 5A,
85~265Vac continuous, 47~63Hz, single phase
2.7 RELIABILITY SPECS
1. Warranty 5 years.
1. Parallel Operation
2. Series Operation Up to 4 units in master/slave mode with single wire current balance connection
Up to 2 units. with external diodes. 600V Max to Chassis ground
2.2 POWER SUPPLY CONFIGURATION
2.3 ENVIRONMENTAL CONDITIONS
1. Operating temp
2. Storage temp -20~70°C
3. Operating humidity 30~90% RH (non-condensing).
4. Storage humidity 10~95% RH (non-condensing).
5. Vibration MIL-810E, method 514.4 , test cond. I-3.3.1. The EUT is fixed to the vibrating surface.
6. Shock Less than 20G , half sine , 11mSec. Unit is unpacked.
7. Altitude Operating: 10000ft (3000m), Derat output current by 2%/100m above 2000m, Non operating: 40000ft (12000m).
0~50°C, 100% load.
More than 100Mohm at 25°C , 70% RH, 500Vdc
2.5 SAFETY
1.Applicable standards:
400<Vout<600V:Output is hazardous, IEEE/Isolated analog are not SELV.
2.Withstand voltage 60<Vout<600V models: Input-Haz. Output: 2.5KVrms 1min, Input-SELV: 3KVrms 1min.
Hazardous Output.-SELV: 1.9KVrms 1min, Hazardous Output-Ground:1.9KVrms 1min.
3.Insulation resistance
CE Mark, UL60950,EN60950 listed. Vout<60V:Output is SELV , IEEE/Isolated analog are SELV.
60<Vout<400V: Output is hazardous, IEEE/Isolated analog are SELV.
Vout<60V models :Input-Outputs (SELV): 3.0KVrms 1min, Input-Ground: 2.0KVrms 1min.
Input-Ground: 2KVrms 1min.
2.6 MECHANICAL CONSTRUCTION
1. Cooling
2. Dimensions (WxHxD) W: 214.0mm, H: 43.6mm, (57.0mm Benchtop version), D: 437.5mm (excluding connectors, encoders, handles, etc.)
3. Weight 4.5Kg (9.9 Lbs)
4. AC Input connector IEC320 AC Inlet.
5.Output connectors 6V to 60V models: Bus-bars (hole Ø 6.5mm). 80V to 600V models: Meating plug, Phoenix P/N: GIC 2.5/4-ST-7.62.
Forced air flow: from front to rear. No ventilation holes at the top or bottom of the chassis; Variable fan speed.
EN55022A, FCC part 15-A, VCCI-A.
2.4 EMC
1.Applicable Standards:
2.ESD IEC1000-4-2. Air-disch.-8KV, contact disch.-4KV
3.Fast transients IEC1000-4-4. 2KV
4.Surge immunity IEC1000-4-5. 1KV line to line, 2KV line to ground
5.Conducted immunity IEC1000-4-6, 3V
6.Radiated immunity IEC1000-4-3, 3V/m
7.Conducted emission EN55022B,FCC part 15J-B,VCCI-B
8.Radiated emission EN55022A,FCC part 15-A,VCCI-A
9.Voltage dips EN61000-4-11
10. Conducted emission
11. Radiated emission EN55022B, FCC part 15-B, VCCI-B.
4
Also available GenesysTM
1U full Rack 750W/1500W & 2U 3300W
Genesys™ Power Benchtop Parallel and Series Configurations
Remote Programming via RS-232 & RS-485 Interface
Standard Serial Interface allows chain control of up to 31 power
supplies on the same bus with built-in RS-232 & RS-485 Interface.
Benchtop Power Supply
5
Parallel operation - Master/Slave:
Active current sharing allows up to four identical units to be connected
in an auto-parallel configuration for four times the output power.
In Adv anced Parallel Master/Sla ve Mode, total current is pr ogrammed
and reported by the Master, Up to four supplies act as one.
Series operation
Up to two units may be connected in series to increase the output
voltage or to provide bipolar output. (Max 600V to Chassis Ground).
Programming Options (Factory installed)
Digital Programming via IEEE Interface P/N: IEEE
IEEE 488.2 SCPI Compliant
Program Voltage Program Current
Measure Voltage Measure Current
Over Voltage setting and shutdown Current Foldback shutdown
Error and Status Messages
New! Multi-Drop
Allows IEEE Master to control up to 31 slaves over RS-485 daisy-chain
Only the Master needs be equipped with IEEE Interface
Isolated Analog Programming
Four Channels to Program and Monitor Voltage and Current.
Isolation allows operation with floating references in harsh electrical environments.
Choose between programming with Voltage or Current.
Connection via removable terminal block: Phoenix MC1,5/8-ST-3.81.
Voltage Programming, user-selectable 0-5V or 0-10V signal. P/N: IS510
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
Current Programming with 4-20mA signal. P/N: IS420
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
LAN Interface Compliant to Class C P/N: LAN
• Meets all LXI-C Requirements
• Address Viewable on Front Panel
• Fixed and Dynamic Addressing
• Compatible with most standard Networks
VISA & SCPI Compatible
• LAN Fault Indicators
• Auto-detects LAN Cross-over Cable
• Fast Startup
Benchtop applications P/N:GENH/MO
The benchtop stacking kit allows the units to be Zero stacked for maximum
system flexibility and power density without increasing the 1U height of the units.
To install a GENH750W two units or three units one on top of the other use
option kit P/N:GENH/MO
Rack Mounting applications P/N:GENH/RM
The Rack Mounted kit allows the units to be zero stacking for maximum system flexibility and
power density without increasing the 1U height of the units
To install one GENH750W unit or two units side-by-side in a standard 19" rack in 1U(1.75")
height, use option kit P/N:GENH/RM
Single unit installation
Single GENH750W power supply in a standard 19" rack
in 1U(1.75") height,
Dual unit installation
Two GENH750W power supplies side-by-side in a standard 19" rack
in 1U(1.75") height,
Accessories
Communication cable
RS-232/RS-485 Cable is used to connect the power supply to the PC Controller.
Serial link cable*
Daisy-chain up to 31 Genesys™ power supplies.
Mode
RS-485 Power Supply Connector Communication Cable
EIA/TIA-568A (RJ-45) Shield Ground L=50cm GEN/RJ45
P/N
* Included with power supply
Mode
P/N
RS-485 RS-232 RS-232
GEN/485-9 GEN/232-9 GEN/232-25
PC Connector
Communication Cable
Power Supply Connector
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-25F
FShield Ground L=2m
EIA/TIA-568A (RJ-45)
6
12
NOTES:
1. Bus-bars 6V to 60V models
Connector 80V to 600V model
Header Phoenix P/N: GIC 2.5/4-G-7.62
Mating plug Phoenix P/N: GIC 2.5/4-ST-7.62
2. Mating plug Phoenix P/N: MC1.5/5-ST-3.81
3. Mating plug AMP P/N: 745211-2
Mating plugs supplied with power supply.
4. Benchtop assembly x 2 (removable)
Screws: 4 x M3x8 marked “A”.
Supplied with the power supply.
NOTE 1
NOTE 1 NOTE 3
NOTE 2
AA
A
A
NOTE 4
6.5mm
M6x16
GENH Models 80V to 600V.
Outline Drawings Genesys™ GENH 750W
7
Power Supply Identification / Accessories
AC Cords sets
How to order
Europe
750W
10A/250Vac L=2m
INT'L 7/VII
IEC320-C13
P/N: GEN/E
Japan
750W
13A/125Vac L=2m
IEC320-C13
P/N: GEN/J
Middle East
750W
10A/250Vac L=2m
SI-32
IEC320-C13
P/N: GEN/I
North America
750W
13A/125Vac L=2m
NEMA 5-15P
IEC320-C13
P/N : GEN/U
Region
Output Power
AC Cords
Wall Plug
Power Supply
Connector
Part Number
United Kingdom
750W
10A/250Vac L=2m
BS1363
IEC320-C13
P/N: GEN/GB
GENH
Series
Name
60
Output
Voltage
(0~60V)
12.5
Output
Current
(0~12.5A)
-
IEEE
IS510
IS420
LAN
Option:
-
E - Europe
GB - United Kingdom
J - Japan
I - Middle East
U- North America
Region:
Factory Options AC Cable option
-
Models GENH750W
Output
Power
( W )
600
750
760
750
760
720
750
760
750
750
750
780
Output
Voltage
VDC
0~6V
0~12.5V
0~20V
0~30V
0~40V
0~8V
0~60V
0~80V
0~100V
0~150V
0~300V
0~600V
Output
Current
( A )
0~100
0~60
0~38
0~25
0~19
0~90
0~12.5
0~9.5
0~7.5
0~5
0~2.5
0~1.3
Model
GENH6-100
GENH12.5-60
GENH20-38
GENH30-25
GENH40-19
GENH8-90
GENH60-12.5
GENH80-9.5
GENH100-7.5
GENH150-5
GENH300-2.5
GENH600-1.3
8
Factory option P/N
RS-232/RS-485 Interface built-in Standard -
GPIB Interface IEEE
Voltage Programming Isolated Analog Interface IS510
Current Programming Isolated Analog Interface IS420
LAN Interface (Complies with Class C) LAN
Programmable DC Power Supplies
750W/1500W in 1U
Built in RS-232 & RS-485 Interface
Advanced Parallel Standard
Optional Interfaces:
IEEE488.2 SCPI (GPIB)
Isolated Analog Programming
Compliant LAN
Test and Measurement
Last-Setting memory simplifies test design and requires no battery backup.
Built-in RS-232/RS-485 gives maximum system flexibility along with 0-5V and 0-10V, selectable analog programming.
Wide range of available inputs allows testing of many different devices.
Semiconductor Burn-in
Safe-Start may be ENABLED to re-start at Output OFF to protect load.
Wide range input (85-265Vac) with Active Power Factor correction rides through input transients easily.
Component Test
High power density, zero stacking and single wire parallel operation give maximum system flexibility.
Laser Diode
OVP is directly set on Voltage Display, assuring accurate protection settings.
Current Limit Fold Back assures load is protected from current surges.
Heater Supplies
Smooth, reliable encoders enhance front panel control.
Remote analog programming is user selectable 0-5V or 0-10V.
RF Amplifiers and Magnets
Robust design assures stable operation under a wide variety of loads.
High linearity in voltage and current mode.
Applications
Genesys™
power supplies have been designed to meet the demands of a wide variety of applications.
Genesys™
750W/1500W-1U
1|
The GenesysTM
family of programmable power supplies sets a new standard for flexible, reliable,
AC/DC power systems in OEM, Industrial and Laboratory applications.
Features include:
High Power Density: 1500W in 1U
Wide Range Input (85 - 265Vac Continuous, single phase, 47/63Hz)
Active Power Factor Correction (0.99 typical)
Output Voltage up to 600V, Current up to 200A
Built-in RS-232/RS-485 Interface Standard
Last-Setting Memory
Front Panel Lock selectable from Front Panel or Software
High Resolution 16 bit ADCs & DACs
Reliable Encoders for Voltage and Current Adjustment
Constant Voltage/Constant Current auto-crossover
Advanced Parallel reports total current up to four identical units
Global Commands for Serial RS-232/RS-485 Interface
Independent Remote ON/OFF and Remote Enable/Disable
External Analog Programming and Monitoring (user selectable 0-5V & 0-10V)
Reliable Modular and SMT Design
19" Rack Mounted ATE and OEM applications
Optional Interfaces
Isolated Analog Programming and Monitoring Interface (0-5V/0-10V & 4-20mA)
IEEE 488.2 SCPI (GPIB) Multi-Drop
Compliant LAN
LabView® and LabWindows® drivers
Five Year W arranty
Worldwide Safety Agency Approvals; CE Mark for LVD and EMC Regulation
1. Remote/Local Output Voltage Sense Connections.
2. DIP Switches select 0-5V or 0-10V Programming and other functions.
3. DB25 (Female) connector allows (Non-isolated) Analog Program and Monitor and other functions.
4. RS-485 OUT to other Genesys™ Power Supplies.
5. RS-232/RS-485 IN Remote Serial Programming.
6. Output Connections: Rugged busbars for up to 60V Output; wire clamp connector for Outputs >60V.
7. Exit air assures reliable operation when zero stacked.
8. Wide-Range Input 85-265VAC continuous, 47/63Hz with Active Power Factor Correction (0.99 typical).
AC Input Connector: 750W (IEC320), 1500W (screw terminal-shown).
9. Optional Interface Position for IEEE 488.2 SCPI (shown) or Isolated Analog Interface or LAN Interface .
Front Panel Description
Rear Panel Description
1. AC ON/OFF Switch
2. Air Intake allows zero stacking for maximum system flexibility and power density.
3. Reliable encoder controls Output Voltage and sets Address.
4. Volt Display shows Output Voltage and directly displays OVP, UVL and Address settings.
5. Reliable encoder controls Output Current, sets baudrate and Advanced Parallel mode.
6. Current Display shows Output Current and displays baudrate.
7. Function/Status LEDs:
8. Pushbuttons allow flexible user configuration
Coarse and Fine adjustment of Output Voltage/Current and Advanced Parallel Master or Slave
Preview settings and set Voltage/Current with Output OFF, Front Panel Lockout
Set OVP and UVL Limits
Set Current Foldback
Local/Remote Mode and select Address and Baudrate
Output ON/OFF and Auto-Start/Safe-Start Mode
Alarm
Fine Control
Preview Settings
Foldback Mode
Remote Mode
Output On
|2
Genesys ™ 750W/1500W Specifications
Genesys™
750W/1500W-1U
3|
1.2 CONSTANT CURRENT MODE
1.Max.line regulation (0.01% of Io+ 2mA)(*4)
2.Max.load regulation (0.02% of Io+5mA)(*6)
3.Ripple r.m.s 5Hz~1MHz . (*7)
4.Max.line regulation (0.01% of Io+ 2mA)(*4)
5.Max.load regulation (0.02% of Io+5mA)(*6)
6.Ripple r.m.s 5Hz~1MHz .(*7)
7.Temp. coefficient
mA
mA
mA
mA
mA
mA
PPM/°C 100PPM/°C from rated output voltage,following 30 minutes warm up
X
X
X
X
X
X
X
X
1.3 PROTECTIVE FUNCTIONS
1. OCP
2. OCP Foldback
3. OVP type
4. OVP trip point
5. Over Temp. Protection
0~105% Constant Current
Output shut down when power supply change from CV to CC. User selectable.
Inverter shut-down, manual reset by AC input recycle or by OUT button or by communication port.
0.5~7.5V
User selectable , latched or non latched
0.5~10V
1~15V 1~24V
2~36V
2~44V 5~66V 5~88V
5~110V 5~165V 5~330V 5~660V
X
X
X
X
X
X
X
X
X
X
GEN
V
A
W
X
X
X
X
1.0 MODEL
1.Rated output voltage (*1)
2.Rated Output Current (*2)
3.Rated Output Power
750W 1500W
3.Rated Output Power
1.0 MODEL
1.Rated output voltage(*1)
2.Rated Output Current(*2)
4.Efficiency at 100/200Vac (*3)
GEN
V
A
W
%X
X
X
X
X
X
*1: Minimum voltage is guaranteed to maximum 0.2% of Vo Rated.
*2: Minimum current is guaranteed to maximum 0.4% of Io Rated
*7: For 6V models the ripple is measured at 2~6V output voltage and full output current. For other models, the ripple is measured at 10~100% output voltage and full output current.
*8: Time for the output voltage to recover within 0.5% of its rated for a load change 10~90% of rated output , Output set-point:10~100%.
*9
For 6V~300V models: measured with JEITA RC-9131A 1:1 probe. For 600V model: measured with 10:1 probe
*3: At maximum output power.
*4: 85~132V ac or 170~265Vac, constant load.
*5: From No-load to Full-load, constant input voltage.
*6: For load voltage change, equal to the unit voltage rating, constant input voltage.
Accuracy -Values have been calculated at Vo Rated & Io Rated
2.Display
3.Indications
Voltage
Current
Voltage, Current, Alarm, Fine, Preview, Foldback, Local, Output On, Front Panel Lock
4 digits , accuracy: 0.5%+/-1 count
4 digits, accuracy: 0.5%+/-1 count X
X
X
X
X
X
1.5 FRONT PANEL
1.Control functions Vout/ Iout manual adjust by separate encoders (coarse and fine adjustment selectable)
OVP/UVL manual adjust by Volt. Adjust encoder
Address selection by Voltage (or current) adjust encoder. Number of addresses:31
AC on/off, Output on/off, Re-start modes (auto, safe), Foldback control (CV to CC), Go to local control
RS232/485 and IEEE488.2 selection by IEEE enable switch and DIP switch
Baudrate selection: 1200,2400,4800,9600 and 19,200
X
X
X
X
X
X
X
X
X
X
X
X
5. OVP/UVL Programming
Resolution (0.1% of Vo Rated)
Accuracy (1% of Vo Rated)
mV
mV X
XX
X
4. Readback Current
Resolution (0.012% of Io Rated )
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output)
Resolution (0.012% of Io Rated )
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output)
mA
mA
mA
mA
X
XX
X
3. Readback Voltage
Resolution (0.012% of Vo Rated)
Accuracy (0.1%Vo Rated+0.1% of Vo Actual Output)
mV
mV X
XX
X
2. Remote Current Programming (16 bit)
Resolution (0.012% of Io Rated)
Accuracy (0.1% of Io Rated+0.1% of Io Actual Output)
Resolution (0.012% of Io Rated)
Accuracy (0.1% of Io Rated+0.1% of Io Actual Output)
mA
mA
mA
mA
X
XX
X
1.6 Interface RS-232&RS-485 or Optional GPIB / LAN Interface
Model
1. Remote Voltage Programming (16 bit)
Resolution (0.012% of Vo Rated)
Accuracy (0.05%Vo Rated+0.05% of Vo Actual Output)
V
mV
mV
750W
X
X
1500W
X
X
XX
1.4 ANALOG PROGRAMMING AND MONITORING
1.Vout Voltage Programming
2.Iout Voltage Programming
3.Vout Resistor Programming
4.Iout Resistor Programming
5.On/Off control (rear panel)
6.Output Current monitor
7.Output Voltage monitor
8.Power Supply OK signal
9. CV/CC indicator
10. Enable/Disable
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:+/-0.5% of rated Vout.
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:+/-1% of rated Iout.
0~100%, 0~5/10Kohm full scale,user select.,Accuracy and linearity:+/-1% of rated Vout.
0~100%, 0~5/10Kohm full scale,user select. Accuracy and linearity:+/-1.5% of rated Iout.
By electrical. Voltage: 0~0.6V/2~15V,or dry contact ,user selectable logic
0~5V or 0~10V , accuracy:1% , user selectable
0~5V or 0~10V ,accuracy:1% ,user selectable
TTL high (4~5V) -OK, 0V-Fail 500ohm series resistance
CV: TTL high (4~5V) source: 10mA, CC: TTL low (0~0.6V), sink current: 10mA
Dry contact. Open:off , Short: on. Max. voltage at Enable/Disable in: 6V
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
XX
X
11. Local/Remote analog control
12. Local/Remote analog control indicator By electrical signal or Open/Short: 0~0.6V or short: Remote, 4~5V or open: Local
Open collector, Local: Open, Remote: On. Maximum voltage: 30V, maximum sink current: 5mA.
1.1 CONSTANT VOLTAGE MODE
1.Max.line regulation ( 0.01% of Vo+ 2mV )(*4)
2.Max load regulation ( 0.01% of Vo+2mV )(*5)
3.Ripple and noise p-p 20MHz (*9)
4.Ripple r.m.s 5Hz~1MHz (*9)
5.Remote sense compensation/line
6.Temp. coefficient
7.Up-prog. response time, 0~Vo Rated
8.Down-prog response time full-load
9.Down-prog response time no-load
10.Transient response time (*8)
mV
mV
mV
mV
V
PPM/°C
mS
mS
mS
100PPM/°C of rated output voltage,following 30 minutes warm up
80mS , N.L/F.L , resistive load
Less than 1mSec for models up to and including 100V. 2msec for models above 100V
80 150mS , N.L/F.L , resistive load
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
12
25
200
22
45
400
6-100
6
100
600
6-200
6
200
1200
77/79
2.6
2.6
60
8
1
10
500
11
23
180
20
41
360
8-90
8
90
720
8-180
8
180
1440
78/81
2.8
2.8
60
8
1
600
8.0
17
120
14
29
240
12.5-60
12.5
60
750
12.5-120
12.5
120
1500
81/84
3.3
3.3
60
8
1
50
700
5.8
12.6
76
9.6
20.2
152
20-38
20
38
760
20-76
20
76
1520
83/86
4
4
60
8
1
800
4.5
10
63
7.0
15
125
30-25
30
25
750
30-50
30
50
1500
83/86
5
5
60
8
1.5
900
3.9
8.8
48
5.8
12.6
95
40-19
40
19
760
40-38
40
38
1520
84/88
6
6
60
8
2
1000
2.13
5.26
8
2.26
5.52
12
600-1.3
600
1.3
780
600-2.6
600
2.6
1560
83/87
62
62
300
60
5
4000
250
250
2.25
5.5
13
2.5
6.0
25
300-2.5
300
2.5
750
300-5
300
5
1500
83/87
32
32
150
25
5
2500
2.5
6.0
18
3.0
7.0
35
150-5
150
5
750
150-10
150
10
1500
84/88
17
17
100
10
5
150
2000
2.75
6.5
23
3.5
8.0
45
100-7.5
100
7.5
750
100-15
100
15
1500
84/88
12
12
80
8
5
1500
2.95
6.9
29
3.9
8.8
57
80-9.5
80
9.5
760
80-19
80
19
1520
84/88
10
10
80
8
4
1200
3.25
7.5
38
4.5
10
75
60-12.5
60
12.5
750
60-25
60
25
1500
84/88
8
8
60
8
3
1100
50-30
---
---
---
5
11
85
5~57V
---
---
---
---
50
30
1500
84/88
7
7
60
8
2
1100
6.0
---
120
100
---
---
3.60
60
50
6
50
50
500
---
3.60
0.72
400
800
12
12
200
24
400
6
0.72
6.0
6
60
12
24
0.96
360
720
16
10.8
180
21.6
360
8
0.96
8.0
8
80
10.8
21.6
2.40
152
304
40
4.56
76
9.12
152
20
2.40
20
20
200
4.56
9.12
1.50
240
480
25
7.2
120
14.4
240
12.5
1.50
12.5
12
125
7.2
14.4
3.60
100
200
60
3.0
50
6.0
100
30
3.60
30
30
300
3.0
6
4.80
76
152
80
2.28
38
4.56
76
40
4.80
40
40
400
2.28
4.56
7.2
50
100
120
1.50
25
3.0
50
60
7.2
60
60
600
1.50
3.0
9.6
38
76
160
1.14
19
2.28
38
80
9.6
80
80
800
1.14
2.28
12
30
60
200
0.90
15
1.80
30
100
12
100
100
1000
0.90
1.80
18
20
40
300
0.60
10
1.20
20
150
18
150
150
1500
0.60
1.20
36
10
20
600
0.30
5.0
0.60
10
300
36
300
300
3000
0.30
0.60
72
5.2
10.4
1200
0.16
2.6
0.32
5.2
600
72
600
600
6000
0.16
0.32
General Specifications Genesys™ 750W/1500W
Outline Drawing Genesys™ 750W/1500W Units
|4
NOTE
1. Bus bars for 6v to 60v models (shown)
Wire clamp connector for 80V to 600V models
2. Plug connectors included with the power supply
3. Chassis slides mounting holes #10-32 marked "A"
GENERAL DEVICES P/N: C-300-S-116 or equivalent
*1: For cases where conformance to various safety standards (UL, IEC etc.) is required, to be described as 100-240Vac (50/60Hz).
All specifications subject to change without notice.
2.1 INPUT CHARACTERISTICS
1. Input voltage/freq. (*1)
2. Power Factor 0.99 @100/200Vac, rated output power.
3. EN61000-3-2,3 compliance Complies with EN61000-3-2 class A and EN61000-3-3 at 20~100% output power.
4. Input current 100/200Vac
5. Inrush current 100/200Vac 750W :Less than 25A, 1500W :Less than 50A
6. Hold-up time More than 20mS , 100Vac , at 100% load.
750W :10.5A / 5A, 1500W :21A / 11A
85~265Vac continuous, 47~63Hz, single phase
2.6 MECHANICAL CONSTRUCTION
1. Cooling
2. Dimensions (WxHxD) W: 422.8mm, H: 43.6mm, D: 432.8mm (excluding connectors, encoders, handles, etc.)
3. Weight 750W: 7Kg (15 Lbs) 1500W: 8.5Kg (18 Lbs)
4. AC Input connector 750W: IEC320 AC Inlet.
1500W: Screw terminal block, Phoenix P/N: FRONT-4-H-7.62 , with strain relief
5.Output connectors 6V to 60V models: Bus-bars (hole Ø 8.5mm). 80V to 600V models:
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62
Forced air flow: from front to rear. No ventilation holes at the top or bottom of the chassis; Variable fan speed.
2.7 RELIABILITY SPECS
1. Warranty 5 years.
1. Parallel Operation
2. Series Operation Up to 4 units in master/slave mode with single wire current balance connection
Up to 2 units. with external diodes. 600V Max to Chassis ground
2.2 POWER SUPPLY CONFIGURATION
2.3 ENVIRONMENTAL CONDITIONS
1. Operating temp
2. Storage temp -20~70°C
3. Operating humidity 30~90% RH (non-condensing).
4. Storage humidity 10~95% RH (non-condensing).
5. Vibration MIL-810E, method 514.4 , test cond. I-3.3.1. The EUT is fixed to the vibrating surface.
6. Shock Less than 20G , half sine , 11mSec. Unit is unpacked.
7. Altitude
Operating: 10000ft (3000m), Derat output current by 2%/100m abouve 2000m, Non operating: 40000ft (12000m).
0~50°C, 100% load.
EN55022A, FCC part 15-A, VCCI-A.
2.4 EMC
1.Applicable Standards:
2.ESD IEC1000-4-2. Air-disch.-8KV, contact disch.-4KV
3.Fast transients IEC1000-4-4. 2KV
4.Surge immunity IEC1000-4-5. 1KV line to line, 2KV line to ground
5.Conducted immunity IEC1000-4-6, 3V
6.Radiated immunity IEC1000-4-3, 3V/m
7.Conducted emission EN55022B,FCC part 15J-B, VCCI-B.
8.Radiated emission EN55022A,FCC part 15-A, VCCI-A.
9.Voltage dips EN61000-4-11
10. Conducted emission
11. Radiated emission EN55022B, FCC part 15-B, VCCI-B.
More than 100Mohm at 25°C , 70% RH, 500Vdc
2.5 SAFETY
1.Applicable standards:
400<Vout<600V:Output is hazardous, IEEE/Isolated analog are not SELV.
2.Withstand voltage 60<Vout<600V models: Input-Haz. Output: 2.5KVrms 1min, Input-SELV: 3KVrms 1min.
Hazardous Output.-SELV: 1.9KVrms 1min, Hazardous Output-Ground:1.9KVrms 1min.
3.Insulation resistance
CE Mark, UL60950,EN60950 listed. Vout<60V:Output is SELV , IEEE/Isolated analog are SELV.
60<Vout<400V: Output is hazardous, IEEE/Isolated analog are SELV.
Vout<60V models :Input-Outputs (SELV): 3.0KVrms 1min, Input-Ground: 2.0KVrms 1min.
Input-Ground: 2KVrms 1min.
Genesys™ Power Parallel and Series Configurations
Remote Programming via RS-232 & RS-485 Interface
Standard Serial Interface allows chain control of up to 31 power
supplies on the same bus with built-in RS-232 & RS-485 Interface.
Genesys™
750W/1500W-1U
5|
Parallel operation - Master/Slave:
Active current sharing allows up to four identical units to be connected
in an auto-parallel configuration for four times the output power.
In Advanced Parallel Master/Slave Mode, total current is
programmed and reported by the Master, Up to four supplies
act as one.
Series operation
Up to two units may be connected in series to increase the output
voltage or to provide bipolar output. (Max 600V to Chassis Ground).
Programming Options (Factory installed)
Digital Programming via IEEE Interface P/N: IEEE
IEEE 488.2 SCPI Compliant
Program Voltage Program Current
Measure Voltage Measure Current
Over Voltage setting and shutdown Current Foldback shutdown
Error and Status Messages
New! Multi-Drop
Allows IEEE Master to control up to 31 slaves over RS-485 daisy-chain
Only the Master needs be equipped with IEEE Interface
Isolated Analog Programming
Four Channels to Program and Monitor Voltage and Current.
Isolation allows operation with floating references in harsh electrical environments.
Choose between programming with Voltage or Current.
Connection via removable terminal block: Phoenix MC1,5/8-ST-3.81.
Voltage Programming, user-selectable 0-5V or 0-10V signal. P/N: IS510
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
Current Programming with 4-20mA signal. P/N: IS420
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
LAN Interface Compliant to Class C P/N: LAN
• Meets all LXI-C Requirements
• Address Viewable on Front Panel
• Fixed and Dynamic Addressing
• Compatible with most standard Networks
VISA & SCPI Compatible
• LAN Fault Indicators
• Auto-detects LAN Cross-over Cable
• Fast Startup
Power Supply Identification / Accessories
Accessories
1. Communication cable
RS-232/RS-485 Cable is used to connect the power supply to the PC Controller.
AC Cords sets (750W only)
How to order
Models 750/1500W
2. Serial link cable*
Daisy-chain up to 31 Genesys™ power supplies.
|6
Mode
RS-485 Power Supply Connector Communication Cable
EIA/TIA-568A (RJ-45) Shield Ground L=50cm GEN/RJ45
P/N
* Included with power supply
Mode
P/N
RS-485 RS-232 RS-232
GEN/485-9 GEN/232-9 GEN/232-25
PC Connector
Communication Cable
Power Supply Connector
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-25F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
Output
Power
( W )
600
1200
720
1440
750
1500
760
1520
750
1500
760
1520
Output
Current
( A )
0~100
0~200
0~90
0~180
0~60
0~120
0~38
0~76
0~25
0~50
0~19
0~38
Output
Voltage
VDC
0~6V
0~8V
0~12.5V
0~20V
0~30V
0~40V
Model
GEN6-100
GEN6-200
GEN8-90
GEN8-180
GEN12.5-60
GEN12.5-120
GEN20-38
GEN20-76
GEN30-25
GEN30-50
GEN40-19
GEN40-38
Europe
750W
10A/250Vac L=2m
INT'L 7/VII
IEC320-C13
P/N: GEN/E
Japan
750W
13A/125Vac L=2m
IEC320-C13
P/N: GEN/J
Middle East
750W
10A/250Vac L=2m
SI-32
IEC320-C13
P/N: GEN/I
North America
750W
13A/125Vac L=2m
NEMA 5-15P
IEC320-C13
P/N : GEN/U
Region
Output Power
AC Cords
Wall Plug
Power Supply
Connector
Part Number
United Kingdom
750W
10A/250Vac L=2m
BS1363
IEC320-C13
P/N: GEN/GB
Model
GEN60-12.5
GEN60-25
GEN80-9.5
GEN80-19
GEN100-7.5
GEN100-15
GEN150-5
GEN150-10
GEN300-2.5
GEN300-5
GEN600-1.3
GEN600-2.6
Output
Power
( W )
750
1500
760
1520
750
1500
750
1500
750
1500
780
1560
Output
Voltage
VDC
0~60V
0~80V
0~100V
0~150V
0~300V
0~600V
Output
Current
( A )
0~12.5
0~25
0~9.5
0~19
0~7.5
0~15
0~5
0~10
0~2.5
0~5
0~1.3
0~2.6
GEN50-30 15000~50V 0~30
-
2.6
Output
Current
(0~2.6A)
-
IEEE
IS510
IS420
LAN
Option:
-
E - Europe
GB - United Kingdom
J - Japan
I - Middle East
U- North America
Region:
Factory Options AC Cable option is 750W only
600
Output
Voltage
(0~600V)
GEN
Series
Name
Factory option P/N
RS-232/RS-485 Interface built-in Standard -
GPIB Interface IEEE
Voltage Programming Isolated Analog Interface IS510
Current Programming Isolated Analog Interface IS420
LAN Interface (Complies with Class C) LAN
Programmable DC Power Supplies
2.4KW in 1U
Built in RS-232 & RS-485 Interface
Advanced Parallel Standard
Optional Interfaces:
Compliant LAN
IEEE488.2 SCPI (GPIB)
Isolated Analog Programming
The Genesys
TM family of programmable power supplies sets a new standard for flexible, reliable,
AC/DC power systems in OEM, Industrial and Laboratory applications.
Features include:
High Power Density 2.4kW in 1U
Wide Range of popular worldwide AC inputs, 1ø (230VAC) & 3ø (208VAC)
Active Power Factor Correction (Single-Phase & Three-Phase AC Input)
Output Voltage up to 600V, Current up to 300A
Built-in RS-232/RS-485 Interface Standard
Global Commands for Serial RS-232/RS-485 Interface
Auto-Re-Start / Safe-Start: user selectable
Last-Setting Memory
High Resolution 16 bit ADCs & DACs
Low Ripple & Noise
Front Panel Lock selectable from Front Panel or Software
Reliable Encoders for Voltage and Current Adjustment
Constant Voltage/Constant Current auto-crossover
Parallel Operation with Active Current Sharing; up to four identical units.
Advanced Parallel Master / Slave. Total Current is Programmed and Measured via the Master.
Independent Remote ON/OFF and Remote Enable/Disable
External Analog Programming and Monitoring (user selectable 0-5V & 0-10V)
Reliable Modular and SMT Design
19" Rack Mount capability for ATE and OEM applications
Optional Interfaces
Isolated Analog Programming and Monitoring Interface (0-5V/0-10V & 4-20mA)
IEEE 488.2 SCPI (GPIB) Multi-Drop
Compliant LAN
LabView® and LabWindows® drivers
Five Year Warranty
Worldwide Safety Agency Approvals; CE Mark for LVD and EMC Regulation
Genesys™
2.4kW-1U
1|
Applications
GenesysTM power supplies have been designed to meet the demands of a wide variety of applications.
System Designers will appreciate new, standard, remote programming features such as Global commands.
Also, new high-speed status monitoring is available for the RS-485 bus.
Test Systems using the IEEE-488 bus may achieve significant cost savings by incorporating the Optional IEEE
Multi-Drop Interface for a Master and up to 30 RS-485 Multi-Drop Slaves.
Higher power systems can be configured with up to four 2.4kW modules. Each module is 1U with zero space
between them (zero stack).
Flexible configuration is provided by the complete GenesysTM Family: 1U 750W Half-Rack, 1U 750W and 1500W
Full-Rack, 2U 3.3kW & 5kW. All are identical in Front Panel, Rear Panel Analog, and all Digital Interface Commands.
A wide variety of outputs allows testing of many different devices.
OEM Designers have a wide variety of Inputs and Outputs from which to select depending on application and
location.
1. Remote/Local Output Voltage Sense Connections.
2. DIP Switches select 0-5V or 0-10V Programming and other functions.
3. DB25 (Female) connector allows (Non-isolated) Analog Program and Monitor and other functions.
4. RS-485 OUT to other Genesys™ Power Supplies.
5. RS-232/RS-485 IN Remote Serial Programming.
6. Output Connections: Rugged busbars (shown) for up to 100V Output; wire clamp connector for Outputs >100V.
7. Exit air assures reliable operation when zero stacked.
8. Input: 230VAC Single Phase (shown), 208 VAC Three Phase, 50/60 Hz
AC Input Connector: Phoenix P/N: FRONT-4-H-7.62.
9. Optional Interface Position for IEEE 488.2 SCPI (shown) or Isolated Analog Interface or LAN Interface.
10. Auxiliary Output Voltage.
Front Panel Description
Rear Panel Description
1. ON/OFF Switch
2. Air Intake allows zero stacking for maximum system flexibility and power density.
3. Reliable encoder controls Output Voltage, Address, OVP and UVL settings.
4. Volt Display shows Output Voltage and directly displays OVP, UVL and Address settings.
5. Reliable encoder controls Output Current, sets baudrate and Advanced Parallel mode.
6. Current Display shows Output Current and displays Baud rate. Displays total current in Parallel Master/Slave Mode
7. Function/Status LEDs:
8. Pushbuttons allow flexible user configuration
Coarse and Fine adjustment of Output Voltage/Current and Advanced Parallel Master or Slave select.
Preview settings and set Voltage/Current with Output OFF, Front Panel Lock
Parallel Master/Slave
Set OVP and UVL Limits
Set Current Foldback Protection
Go to Local Mode and select Address and Baud rate
Output ON/OFF and Auto-Re-Start/Safe-Start Mode
Alarm
Foldback Mode
Fine Control
Remote Mode
|2
Preview Settings
Output On
1 2 3 4,5
9
6 7
8
10
1
2
3 4 6
7
8
5
Genesys™
2.4kW-1U
3|
Genesys ™ 2.4kW Specifications
*1: Minimum voltage is guaranteed to maximum 0.2% of rated output voltage.
*2: Minimum current is guaranteed to maximum 0.4% of rated output current.
*3: For cases where conformance to various safety standards (UL, IEC, etc) is required, to be
described as 190-240Vac (50/60Hz) for 3-Phase 208V models.
*4: 3-Phase 208V models: At 208Vac input voltage. With rated output power.
*5: Not including EMI filter inrush current, less than 0.2mSec.
*6: 3-Phase 208V models: 170~265Vac, constant load.
*7: From No-Load to Full-Load, constant input voltage. Maximum drop in Remote Sense.
*8: For 8V~300V models: Measured with JEITA RC-9131A (1:1) probe. For 600V model: Measured
with 10:1 probe.
*9: From 10% to 90% or 90% to 10% of Rated Output Voltage, with rated, resistive load.
*10:From 90% to 10% of Rated Output Voltage.
*11: For load voltage change, equal to the unit voltage rating, constant input voltage.
*12: For 8V~16V models the ripple is measured from 2V to rated output voltage and rated output
current. For other models, the ripple is measured at 10~100% of rated output voltage and rated
output current.
*13: The Constant Current programming readback and monitoring accuracy does not include the
warm-up and Load regulation thermal drift.
1.2 CONSTANT CURRENT MODE
1.Max.line regulation (0.01% of Io rated+2mA)(*6)
2.Max.load regulation (0.02% of Io rated+5mA)(*11)
3.Ripple r.m.s 5Hz~1MHz. (*12)
4.Temp. coefficient
8V
~
20V models: Less than ±0.5% of rated output current over 30 minutes following power On.
1.3 PROTECTIVE FUNCTIONS
1. OCP
2. OCP Foldback
3. OVP type
4. OVP trip point
6. Over Temp. Protection
0~105% Constant Current
Output shut down when power supply change from CV to CC. User selectable.
Inverter shut-down, manual reset by AC input recycle or by OUT button or by communication port command.
User selectable , latched or non-latched.
5~88V 5~110V
5~165V 5~330V 5~660V
3.Rated Output Power
1.0 MODEL
1.Rated output voltage(*1)
2.Rated Output Current(*2)
2.Display
3.Indications
Voltage: 4 digits , Accuracy: 0.5% of rated output Voltage ±1 count.
Current: 4 digits, Accuracy: 0.5% of rated output current ±1 count.
Voltage, Current, Alarm, Fine, Preview, Foldback, Local, Output On, Front Panel Lock, CV/CC.
1.5 FRONT PANEL
1.Control functions Vout/ Iout manual adjust by separate encoders (coarse and fine adjustment selectable).
OVP/UVL manual adjust by Volt. Adjust encoder.
Address selection by Voltage (or current) adjust encoder. Number of addresses:31.
On/Off, Output on/off, Re-start modes (auto, safe), Foldback control (CV to CC), Go to local control.
Re-start modes (automatic restart, safe mode).
Baud rate selection: 1200,2400,4800,9600 and 19,200.
1.4 ANALOG PROGRAMMING AND MONITORING
1.Vout Voltage Programming
2.Iout Voltage Programming (*13)
3.Vout Resistor Programming
4.Iout Resistor Programming (*13)
5.On/Off control (rear panel)
6.Output Current monitor (*13)
7.Output Voltage monitor
8.Power Supply OK signal
9. CV/CC Indicator
10. Enable/Disable
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±0.5% of rated Vout.
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±1% of rated Iout.
0~100%, 0~5/10Kohm full scale,user select.,Accuracy and linearity: ±1% of rated Vout.
0~100%, 0~5/10Kohm full scale,user select. Accuracy and linearity:±1.5% of rated Iout.
By electrical. Voltage: 0~0.6V/2~15V,or dry contact ,user selectable logic.
0~5V or 0~10V , Accuracy:±1% , user selectable.
0~5V or 0~10V ,Accuracy:±1% ,user selectable.
TTL high (4~5V) -OK, 0V-Fail 500ohm series resistance.
Dry contact. Open:off , Short: on. Max. voltage at Enable/Disable in: 6V.
11. Local/Remote analog control
12. Local/Remote analog control Indicator
Open collector, CC mode: On, CV mode: Off, Maximum voltage: 30V, maximum sink current: 10mA
Open collector, Local: Off, Remote: On. Maximum voltage: 30V, maximum sink current: 10mA.
1.Max.line regulation (0.01% of rated Vo+2mV)(*6)
2.Max load regulation (0.015% of rated Vo+5mV)(*7)
3.Ripple and noise p-p 20MHz (*8)
4.Ripple r.m.s 5Hz~1MHz
5.Remote sense compensation/wire
8.Warm-up drift
9.Up-prog. response time, 0~Vo Rated (*9)
10.Down-prog response time
No-load (*10)
11.Transient response time
100PPM/°C of rated output voltage,following 30 minutes warm-up
30mS
Time for output voltage to recover within 0.5% of its rated output for a load change 10-90% of rated output
current. Output set-point: 10-100%, local sense.
Less than 1mSec for models up to and including 100V. 2msec for models above 100V
100
100mS
80
7.Temp. stability
6.Temp. coefficient
Full-load (*9)
0.05% of rated Vout over 8hrs interval following 30 minutes warm-up. Constant line, load & temp.
Less than 0.05% of rated output voltage+2mV over 30 minutes following power On.
5.Temp. stability
6.Warm-up drift
100PPM/ºC from rated output current, following 30 minutes warm-up.
0.05% of rated Iout over 8hrs. interval following 30minutes warm-up. Constant line, load & temperature.
5. Output Under Voltage Limit Preset by front panel or communication port. Prevents from adjusting Vout below limit.
2.4
5.8
7
600
62
95
300
75
5
2.8
6.6
15
300
32
50
150
35
5
3000
3.6
8.2
30
150
17
27.5
100
25
5
2500
4.4
9.8
40
100
12
20
80
8
5
1500
5
11
50
80
10
17
80
8
5
1200
6
13
70
5~66V
60
8
14
60
8
5
1100
2~44V
8
17
120
40
6
11
60
8
5
1000
2~36V
10
21
220
30
5
9.5
60
8
5
900
1~24V
14
29
480
20
4
8
60
8
2
800
1~19V
17
35
600
16
3.6
7.4
60
8
2
30
700
0.5~12V
26
53
960
10
3
6.5
60
8
2
600
0.5~10V
32
65
1200
8
2.8
6.2
60
8
2
10
500
mA
mA
mA
V
A
W
mV
mV
mV
mV
V
PPM/°C
mS
mS
mS
mS
PPM/°C
60mS
1.1 CONSTANT VOLTAGE MODE
30V
~
600V models: Less than ±0.25% of rated output current over 30 minutes following power On.
4
2400
A
8
2400
B
16
2400
A
24
2400
C
30
2400
C
40
2400
A
60
2400
B
80
2400
B
120
2400
C
150
2400
B
240
2400
C
300
2400
A
4.Development Priority ---
By electrical signal or Open/Short: 0~0.6V or short: Remote, 2~15V or open: Local.
5. OVP/UVL Programming
Resolution (0.1% of Vo Rated)
Accuracy (1% of Vo Rated)
4. Readback Current
Resolution (0.012% of Io Rated )
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output) (*13)
3. Readback Voltage
Resolution (0.012% of Vo Rated)
Accuracy (0.1%Vo Rated+0.1% of Vo Actual Output)
2. Remote Current Programming (16 bit)
Resolution (0.012% of Io Rated)
Accuracy (0.2% of Io Rated+0.1% of Io Actual Output) (*13)
1.6 Interface RS-232&RS-485 or Optional GPIB / LAN Interface
Model
1. Remote Voltage Programming (16 bit)
Resolution (0.012% of Vo Rated)
Accuracy (0.05%Vo Rated+0.05% of Vo Actual Output)
72
16
1200
0.48
12
600
72
600
600
6000
0.48
36
32
600
0.96
24
300
36
300
300
3000
0.96
18
64
300
1.92
48
150
18
150
150
1500
1.92
12
96
200
2.88
72
100
12
100
100
1000
2.88
9.6
120
160
3.6
90
80
9.6
80
80
800
3.6
7.2
160
120
4.8
120
60
7.2
60
60
600
4.8
4.8
240
80
7.2
180
40
4.8
40
40
400
7.2
3.6
320
60
9.6
240
30
3.6
30
30
300
9.6
2.4
480
40
14.4
360
20
2.4
20
20
200
14.4
1.92
600
32
18
450
16
1.92
16
16
160
18
1.2
960
20
28.8
720
10
1.2
10
10
100
28.8
0.96
1200
16
36
900
8
0.96
8
8
80
36
mV
mV
mA
mA
mV
mV
mA
mA
V
mV
mV
GEN 8-300 10-240 16-150 20-120 30-80 40-60 600-4300-8150-16100-2480-3060-40
15mS
General Specifications Genesys™ 2.4kW
All specifications subject to change without notice.
2.1 INPUT CHARACTERISTICS
2.7 MECHANICAL CONSTRUCTION
1. Cooling
2. Dimensions (WxHxD) W: 423mm, H: 43.6mm, D: 432.8mm (excluding connectors, encoders, handles, etc.)
3. Weight 10 kg.
4. AC Input connector (with Protective Cover)
5.Output connectors 8V to 100V models: Bus-bars (hole Ø 8.5mm ). 150V to 600V models:
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62
Forced air flow: from front to rear. No ventilation holes at the top or bottom of the chassis; Variable fan speed.
2.8 RELIABILITY SPECS
1. Warranty 5 years.
1. Parallel Operation
2. Series Operation
Up to 4 identical units in master/slave mode
Up to 2 identical units. with external diodes. 600V Max to Chassis ground
2.3 POWER SUPPLY CONFIGURATION
2.4 ENVIRONMENTAL CONDITIONS
1. Operating temp
2. Storage temp -20~85°C
3. Operating humidity 20~90% RH (non-condensing).
4. Storage humidity 10~95% RH (non-condensing).
5. Vibration MIL-810F, method 514.5 , The EUT is fixed to the vibrating surface.
6. Shock Less than 20G , half sine , 11mSec. Unit is unpacked.
7. Altitude
Operating: 10000ft (3000m), Derate output current by 2%/100m above 2000m, Alternatively, derate maximum ambient temp.
by 1ºC/100m above 2000m. Non operating: 40000ft (12000m).
0~50°C, 100% load.
EN55022A, FCC part 15-A, VCCI-A.
2.5 EMC
1.Applicable Standards:
2.ESD IEC1000-4-2. Air-disch.-8KV, contact disch.-4KV
3.Fast transients IEC1000-4-4. 2KV
4.Surge immunity IEC1000-4-5. 1KV line to line, 2KV line to ground
5.Conducted immunity IEC1000-4-6, 3V
6.Radiated immunity IEC1000-4-3, 3V/m
7.Magnetic field immunity EN61000-4-8, 1A/m
8.Voltage dips EN61000-4-11
9.Conducted emission
10. Radiated emission
EN55022A, FCC part 15-A, VCCI-A.
More than 100Mohm at 25°C , 70% RH.
2.6 SAFETY
1.Applicable standards:
400<Vout600V:Output is hazardous, IEEE/Isolated analog are not SELV.
2.Withstand voltage
40<Vout100V models: Input-Haz. Output: 2600VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 1900VDC 1min, Hazardous Output-Ground:1200VDC 1min. Input-Ground: 2828VDC 1min.
3.Insulation resistance
CE Mark, UL60950,EN60950 listed. Vout40V:Output is SELV , IEEE/Isolated analog are SELV.
40<Vout400V: Output is hazardous, IEEE/Isolated analog are SELV.
Vout40V models :Input-Outputs (SELV): 4242VDC 1min, Input-Ground: 2828VDC 1min.
1. Input voltage/freq. (*3)
3. Power Factor (Typ)
4. Efficiency (*4)
10mSec for Single-Phase and 3-phase 208V models. Rated output power.
5. Inrush Current (*5)
6. Hold-up time (Typ)
Single Phase,230V models: 170~265Vac, 47~63Hz
3-Phase, 208V models: 170~265Vac, 47~63Hz
2. Maximum Input
current at 100% load
Single Phase,230V models:
3-Phase, 208V models: A
Single-Phase and 3-Phase 208V models: Less than 50A
mS
17
10.5
GEN 8-300 10-240 16-150 20-120 30-80 40-60 600-4300-8150-16100-2480-3060-40
17
10.5
17
10.5
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
16.3
9.8
Single Phase models: 0.99@230Vac, rated output power. 3-Phase models: 0.94@208 Vac, rated output power.
%84 84 84 86 86 88 88 88 88 88 88 88
8. RoHS Compliance
Complies with the requirements of RoHS directive.
100<Vout600V models: Input-Haz. Output: 4000VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 3550VDC 1min. Hazardous Output-Ground:2670VDC 1min. Input-Ground: 2828VDC 1min.
A
VAC
Single Phase,230V models,
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62, with Strain relief.
3-Phase, 208V models,
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62, with Strain relief.
Outline Drawing Genesys™ 2.4kW Units
|4
NOTE
Auxiliary autput Header: IMC 1.5/7-G-3.81, Plug: IMC 1.5/7-ST-3.81 (Phoenix Contact).
1. 15V output
2. 5V output
15V
±5%, 0.2A Max load, Ripple & Noise 50mVp-p. Referenced internally to the negative output potential.
5V
±5%, 0.2A Max load, Ripple & Noise 50mVp-p. Referenced internally to IF_com potential.
2.2 AUXILIARY OUTPUT
Genesys™ Power Parallel and Series Configurations
Parallel operation - Master/Slave:
Active current sharing allows up to four identical units to be connected
in an auto-parallel configuration for four times the output power.
In Advanced Parallel Master/Slave Mode, total current is
programmed and reported by the Master, Up to four supplies
act as one.
Series operation
Up to two units may be connected in series to increase the output
voltage or to provide bipolar output. (Max 600V to Chassis Ground).
Remote Programming via RS-232 & RS-485 Interface
Standard Serial Interface allows daisy-chain control of up to 31
power supplies on the same communication bus with built-in
RS-232 & RS-485 Interface.
Genesys™
2.4kW-1U
5|
Programming Options (Factory installed)
Digital Programming via IEEE Interface P/N: IEEE
IEEE 488.2 SCPI Compliant
Program Voltage Program Current
Measure Voltage Measure Current
Over Voltage setting and shutdown Current Foldback shutdown
Error and Status Messages
New! Multi-Drop
Allows IEEE Master to control up to 31 slaves over RS-485 daisy-chain
Only the Master needs be equipped with IEEE Interface
Isolated Analog Programming
Four Channels to Program and Monitor Voltage and Current.
Isolation allows operation with floating references in harsh electrical environments.
Choose between programming with Voltage or Current.
Connection via removable terminal block: Phoenix MC1,5/8-ST-3.81.
Voltage Programming, user-selectable 0-5V or 0-10V signal. P/N: IS510
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
Current Programming with 4-20mA signal. P/N: IS420
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
LAN Interface Compliant to Class C P/N: LAN
• Meets all LXI-C Requirements
• Address Viewable on Front Panel
• Fixed and Dynamic Addressing
• Compatible with most standard Networks
VISA & SCPI Compatible
• LAN Fault Indicators
• Auto-detects LAN Cross-over Cable
• Fast Startup
Power Supply Identification / Accessories
How to order
Models 2.4kW
|6
Accessories
1. Serial Communication cable
RS-232/RS-485 cable is used to connect the power supply to the Host PC.
2. Serial link cable*
Daisy-chain up to 31 Genesys™ power supplies.
Mode
RS-485
Power Supply Connector Communication Cable
EIA/TIA-568A (RJ-45) Shield Ground L=50cm GEN/RJ45
P/N
* Included with power supply
Mode
P/N
RS-485 RS-232 RS-232
GEN/485-9 GEN/232-9 GEN/232-25
PC Connector
Communication Cable
Power Supply Connector
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-25F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~8V
0~10V
0~15V
0~20V
0~30V
0~40V
Model
0~300
0~240
0~150
0~120
0~80
0~60
2400
2400
2400
2400
2400
2400
GEN 8-300
GEN 10-240
GEN 16-150
GEN 20-120
GEN 30-80
GEN 40-60
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~60V
0~80V
0~100V
0~150V
0~300V
0~600V
Model
0~40
0~30
0~24
0~16
0~8
0~4
2400
2400
2400
2400
2400
2400
GEN 60-40
GEN 80-30
GEN 100-24
GEN 150-16
GEN 300-8
GEN 600-4
-
300
Output
Current
(0~300A)
-
IEEE
IS510
IS420
LAN
-
1P230 (
Single Phase
170~265VAC)
3P208 (Three Phase 170~265VAC)
Factory Options: Factory AC Input Options:
8
Output
Voltage
(0~8V)
GEN
Series
Name
Option:
Factory option P/N
RS-232/RS-485 Interface built-in Standard -
GPIB Interface IEEE
Voltage Programming Isolated Analog Interface IS510
Current Programming Isolated Analog Interface IS420
LAN Interface (Complies with Class C) LAN
Also available, GenesysTM
1U Half Rack 750W
1U full Rack 750W/1500W/2400W
2U full Rack 3300W/5000W
Programmable DC Power Supplies
3.3KW in 2U
Built in RS-232 & RS-485 Interface
Advanced Parallel Standard
Optional Interfaces:
IEEE488.2 SCPI (GPIB)
Isolated Analog Programming
Compliant LAN
The Genesys
TM family of programmable power supplies sets a new standard for flexible, reliable,
AC/DC power systems in OEM, Industrial and Laboratory applications.
Features include:
High Power Density 3.3kW in 2U
Wide Range of popular worldwide AC inputs, 1ø (230VAC) & 3ø (208VAC, 400VAC)
Active Power Factor Correction (Single-Phase & Three-Phase AC Input)
Output Voltage up to 600V, Current up to 400A
Built-in RS-232/RS-485 Interface Standard
Global Commands for Serial RS-232/RS-485 Interface
Auto-Re-Start / Safe-Start: user selectable
Last-Setting Memory
High Resolution 16 bit ADCs & DACs
Low Ripple & Noise
Front Panel Lock selectable from Front Panel or Software
Reliable Encoders for Voltage and Current Adjustment
Constant Voltage/Constant Current auto-crossover
Parallel Operation with Active Current Sharing; up to four identical units.
Advanced Parallel Master / Slave. Total Current is Programmed and Measured via the Master.
Independent Remote ON/OFF and Remote Enable/Disable
External Analog Programming and Monitoring (user selectable 0-5V & 0-10V)
Reliable Modular and SMT Design
19" Rack Mount capability for ATE and OEM applications
Optional Interfaces
Isolated Analog Programming and Monitoring Interface (0-5V/0-10V & 4-20mA)
IEEE 488.2 SCPI (GPIB) Multi-Drop
Compliant LAN
LabView® and LabWindows® drivers
Five Year Warranty
Worldwide Safety Agency Approvals; CE Mark for LVD and EMC Regulation
Genesys™
3.3kW-2U
1|
Applications
GenesysTM power supplies have been designed to meet the demands of a wide variety of applications.
System Designers will appreciate new, standard, remote programming features such as Global commands.
Also, new high-speed status monitoring is available for the RS-485 bus.
Test Systems using the IEEE-488 bus may achieve significant cost savings by incorporating the Optional IEEE
Multi-Drop Interface for a Master and up to 30 RS-485 Multi-Drop Slaves.
Higher power systems can be configured with up to four 3.3kW modules. Each module is 2U with zero space
between them (zero stack).
Flexible configuration is provided by the complete GenesysTM Family: 1U 750W Half-Rack, 1U 750W and 1500W
Full-Rack. All are identical in Front Panel, Rear Panel Analog, and all Digital Interface Commands.
OEM Designers have a wide variety of Inputs and Outputs from which to select depending on application and
location.
1. Remote/Local Output Voltage Sense Connections.
2. DIP Switches select 0-5V or 0-10V Programming and other functions.
3. DB25 (Female) connector allows (Non-isolated) Analog Program and Monitor and other functions.
4. RS-485 OUT to other Genesys™ Power Supplies.
5. RS-232/RS-485 IN Remote Serial Programming.
6. Output Connections: Rugged busbars (shown) for up to 100V Output; wire clamp connector for Outputs >100V.
7. Exit air assures reliable operation when zero stacked.
8. Input: 230VAC Single Phase (shown), 208 & 400VAC Three Phase, 50/60 Hz
AC Input Connector: PHOENIX CONTACT Power Combicon PC 6/... Series with strain relief.
9. Optional Interface Position for IEEE 488.2 SCPI (shown) or Isolated Analog Interface or LAN Interface .
Front Panel Description
Rear Panel Description
1. ON/OFF Switch
2. Air Intake allows zero stacking for maximum system flexibility and power density.
3. Reliable encoder controls Output Voltage, Address, OVP and UVL settings.
4. Volt Display shows Output Voltage and directly displays OVP, UVL and Address settings.
5. Reliable encoder controls Output Current, sets baudrate and Advanced Parallel mode.
6. Current Display shows Output Current and displays Baud rate. Displays total current in Parallel Master/Slave Mode
7. Function/Status LEDs:
8. Pushbuttons allow flexible user configuration
Coarse and Fine adjustment of Output Voltage/Current and Advanced Parallel Master or Slave
Preview settings and set Voltage/Current with Output OFF, Front Panel Lock
Parallel Master/Slave
Set OVP and UVL Limits
Set Current Foldback Protection
Go to Local Mode and select Address and Baud rate
Output ON/OFF and Auto-Re-Start/Safe-Start Mode
Alarm
Foldback Mode
Fine Control
Remote Mode
|2
Preview Settings
Output On
1 2 3 4,5
9 6 7 8
1
2 3
4 6
7
8
5
Genesys™
3.3kW-2U
3|
*1: Minimum voltage is guaranteed to maximum 0.2% of rated output voltage.
*2: Minimum current is guaranteed to maximum 0.4% of rated output current.
*3: For cases where conformance to various safety standards (UL, IEC, etc) is required, to be
described as 190-240Vac (50/60Hz) for single phase and 3-Phase 208V models,
and 380~415Vac (50/60Hz) for 3-Phase 400V models.
*4: Single-Phase and 3-Phase 208V models: At 208Vac input voltage, 3-Phase 400V:
At 380Vac input voltage. With rated output power.
*5: Not including EMI filter inrush current, less than 0.2mSec.
*6: Single-Phase and 3-Phase 208V models: 170~265Vac, constant load. 3-Phase 400V
models: 342~460Vac, constant load.
*7: From No-Load to Full-Load, constant input voltage. Maximum drop in Remote Sense.
*8: For 8V~300V models: Measured with JEITA RC-9131A (1:1) probe. For 600V model: Measured
with 10:1 probe.
*9: From 10% to 90% or 90% to 10% of Rated Output Voltage, with rated, resistive load.
*10:From 90% to 10% of Rated Output Voltage.
*11:For load voltage change, equal to the unit voltage rating, constant input voltage.
*12:For 8V~15V models the ripple is measured from 2V to rated output voltage and rated output
current. For other models, the ripple is measured at 10~100% of rated output voltage and rated
output current.
*13: The Constant Current programming readback and monitoring accuracy does not include the
warm-up and Load regulation thermal drift.
Genesys ™ 3.3kW Specifications
1.2 CONSTANT CURRENT MODE
1.Max.line regulation (0.01% of rated Io+ 2mA)(*6)
2.Max.load regulation (0.02% of rated Io+5mA)(*11)
3.Ripple r.m.s 5Hz~1MHz . (*12)
5.Temp. coefficient
8V
~40V
models: Less than 0.5% of rated output current over 30 minutes following power On.
1.3 PROTECTIVE FUNCTIONS
1. OCP
2. OCP Foldback
3. OVP type
4. OVP trip point
6. Over Temp. Protection
0~105% Constant Current
Output shut down when power supply change from CV to CC. User selectable.
Inverter shut-down, manual reset by AC input recycle or by OUT button or by communication port command.
User selectable , latched or non-latched.
5~88V 5~110V
5~165V 5~330V 5~660V
3.Rated Output Power
1.0 MODEL
1.Rated output voltage(*1)
2.Rated Output Current(*2)
2.Display
3.Indications
Voltage: 4 digits , Accuracy: 0.5% of rated output Voltage ±1 count.
Current: 4 digits, Accuracy: 0.5% of rated output current ±1 count.
Voltage, Current, Alarm, Fine, Preview, Foldback, Local, Output On, Front Panel Lock, CVCC.
1.5 FRONT PANEL
1.Control functions Vout/ Iout manual adjust by separate encoders (coarse and fine adjustment selectable).
OVP/UVL manual adjust by Volt. Adjust encoder.
Address selection by Voltage (or current) adjust encoder. Number of addresses:31.
On/Off, Output on/off, Re-start modes (auto, safe), Foldback control (CV to CC), Go to local control.
Re-start modes (automatic restart, safe mode).
Baud rate selection: 1200,2400,4800,9600 and 19,200.
5. OVP/UVL Programming
Resolution (0.1% of Vo Rated)
Accuracy (1% of Vo Rated)
4. Readback Current
Resolution (0.012% of Io Rated )
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output) (*13)
3. Readback Voltage
Resolution (0.012% of Vo Rated)
Accuracy (0.1%Vo Rated+0.1% of Vo Actual Output)
2. Remote Current Programming (16 bit)
Resolution (0.012% of Io Rated)
Accuracy (0.2% of Io Rated+0.1% of Io Actual Output) (*13)
1.6 Interface RS-232&RS-485 or Optional GPIB / LAN Interface
Model
1. Remote Voltage Programming (16 bit)
Resolution (0.012% of Vo Rated)
Accuracy (0.05%Vo Rated+0.05% of Vo Actual Output)
1.4 ANALOG PROGRAMMING AND MONITORING
1.Vout Voltage Programming
2.Iout Voltage Programming (*13)
3.Vout Resistor Programming
4.Iout Resistor Programming (*13)
5.On/Off control (rear panel)
6.Output Current monitor (*13)
7.Output Voltage monitor
8.Power Supply OK signal
9. CV/CC Indicator
10. Enable/Disable
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±0.5% of rated Vout.
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±1% of rated Iout.
0~100%, 0~5/10Kohm full scale,user select.,Accuracy and linearity: ±1% of rated Vout.
0~100%, 0~5/10Kohm full scale,user select. Accuracy and linearity:±1.5% of rated Iout.
By electrical. Voltage: 0~0.6V/2~15V,or dry contact ,user selectable logic.
0~5V or 0~10V , Accuracy:±1% , user selectable.
0~5V or 0~10V ,Accuracy:±1% ,user selectable.
TTL high (4~5V) -OK, 0V-Fail 500ohm series resistance.
CV: TTL high (4~5V) source: 10mA, CC: TTL low (0~0.6V), sink current: 10mA.
Dry contact. Open:off , Short: on. Max. voltage at Enable/Disable in: 6V.
11. Local/Remote analog control
12. Local/Remote analog control Indicator By electrical signal or Open/Short: 0~0.6V or short: Remote, 4~5V or open: Local.
Open collector, Local: Off, Remote: On. Maximum voltage: 30V, maximum sink current: 10mA.
1.Max.line regulation ( 0.01% of rated Vo+ 2mV )(*6)
2.Max load regulation ( 0.015% of rated Vo+5mV )(*7)
3.Ripple and noise p-p 20MHz (*8)
4.Ripple r.m.s 5Hz~1MHz
5.Remote sense compensation/wire
8.Warm-up drift
9.Up-prog. response time, 0~Vo Rated (*9)
10.Down-prog response time No-load (*10)
11.Transient response time
100PPM/°C of rated output voltage,following 30 minutes warm-up
80
Time for output voltage to recover within 0.5% of its rated output for a load change 10-90% of rated output
current. Output set-point: 10-100%, local sense.
Less than 1mSec for models up to and including 100V. 2msec for models above 100V
160 500
250
300
7.Temp. stability
6.Temp. coefficient
Full-load (*9)
0.05% of rated Vout over 8hrs interval following 30 minutes warm-up. Constant line, load & temp.
Less than 0.05% of rated output voltage+2mV over 30 minutes following power On.
4.Load regulation thermal drift
6.Temp. stability
7.Warm-up drift
Less than 0.1% of rated output current over 30 minutes following load change.
200PPM/ºC from rated output current, following 30 minutes warm-up.
0.05% of rated Iout over 8hrs. interval following 30minutes warm-up. Constant line, load & temperature.
5. Output Under Voltage Limit Preset by front panel or communication port. Prevents from adjusting Vout below limit.
GEN 8-400 10-330 15-220 20-165 30-110 40-85 600-5.5300-11150-22100-3380-4260-55
2.6
6.1
10
600
5.5
3300
62
95
500
120
5
4000
72
22
1200
0.7
16.5
600
72
600
600
6000
0.7
3.1
7.2
50
300
11
3300
32
50
150
35
5
3500
36
44
600
1.3
33
300
36
300
300
3000
1.3
4.2
9.4
60
150
22
3300
17
27.5
100
25
5
2000
18
88
300
2.6
66
150
18
150
150
1500
2.6
5.3
11.6
90
100
33
3300
12
20
80
8
5
1500
12
132
200
4.0
99
100
12
100
100
1000
4.0
6.2
13.4
120
80
42
3360
10
17
80
8
5
1200
9.6
168
160
5.0
126
80
9.6
80
80
800
5.0
7.5
16
100
5~66V
60
55
3300
8
14
60
8
5
1100
7.2
220
120
6.6
165
60
7.2
60
60
600
6.6
2~44V
10.5
22
200
40
85
3400
6
11
60
8
5
1000
4.80
340
80
10.2
255
40
4.80
40
40
400
10.2
2~36V
13
27
250
30
110
3300
5
9.5
60
8
5
900
3.60
440
60
13.2
330
30
3.60
30
30
300
13.2
1~24V
18.5
38
300
20
165
3300
4
8
60
8
2
800
2.40
660
40
19.8
495
20
2.40
20
20
200
19.8
1~18V
24
49
440
15
220
3300
3.5
7.25
60
8
2
100
700
1.8
880
30
26.4
660
15
1.8
15
15
150
26.4
0.5~12V
35
71
660
10
330
3300
3
6.5
60
8
2
600
1.2
1320
20
39.6
990
10
1.2
10
10
100
39.6
0.5~10V
42
85
1300
8
400
3200
2.8
6.2
60
8
2
20
500
0.96
1600
16
48
1200
8
0.96
8
8
80
48
mA
mA
mA
V
A
W
mV
mV
mA
mA
mV
mV
mA
mA
V
mV
mV
mV
mV
mV
mV
V
PPM/°C
mS
mS
mS
mS
PPM/°C
150
1.1 CONSTANT VOLTAGE MODE
60V
~
600V models: Less than 0.25% of rated output current over 30 minutes following power On.
General Specifications Genesys™ 3.3kW
All specifications subject to change without notice.
2.1 INPUT CHARACTERISTICS
2.6 MECHANICAL CONSTRUCTION
1. Cooling
2. Dimensions (WxHxD) W: 423mm, H: 88mm, D: 442.5mm (excluding connectors, encoders, handles, etc.)
3. Weight 13 kg.
4. AC Input connector (with Protective Cover)
5.Output connectors 8V to 100V models: Bus-bars (hole Ø 10.5mm). 150V to 600V models:
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62
Forced air flow: from front to rear. No ventilation holes at the top or bottom of the chassis; Variable fan speed.
2.7 RELIABILITY SPECS
1. Warranty 5 years.
1. Parallel Operation
2. Series Operation Up to 4 identical units in master/slave mode
Up to 2 identical units. with external diodes. 600V Max to Chassis ground
2.2 POWER SUPPLY CONFIGURATION
2.3 ENVIRONMENTAL CONDITIONS
1. Operating temp
2. Storage temp -30~85°C
3. Operating humidity 20~90% RH (non-condensing).
4. Storage humidity 10~95% RH (non-condensing).
5. Vibration MIL-810F, method 514.5 , The EUT is fixed to the vibrating surface.
6. Shock Less than 20G , half sine , 11mSec. Unit is unpacked.
7. Altitude
Operating: 10000ft (3000m), Derate output current by 2%/100m above 2000m, Alternatively, derate maximum ambient temp.
by 1ºC/100m above 2000m. Non operating: 40000ft (12000m).
0~50°C, 100% load.
EN55022A, FCC part 15-A, VCCI-A.
2.4 EMC
1.Applicable Standards:
2.ESD IEC1000-4-2. Air-disch.-8KV, contact disch.-4KV
3.Fast transients IEC1000-4-4. 2KV
4.Surge immunity IEC1000-4-5. 1KV line to line, 2KV line to ground
5.Conducted immunity IEC1000-4-6, 3V
6.Radiated immunity IEC1000-4-3, 3V/m
7.Magnetic field immunity EN61000-4-8, 1A/m
8.Voltage dips EN61000-4-11
9.Conducted emission
10. Radiated emission EN55022A, FCC part 15-A, VCCI-A.
More than 100Mohm at 25°C , 70% RH.
2.5 SAFETY
1.Applicable standards:
400<Vout600V:Output is hazardous, IEEE/Isolated analog are not SELV.
2.Withstand voltage 40<Vout100V models: Input-Haz. Output: 2600VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 1900VDC 1min, Hazardous Output-Ground:1200VDC 1min. Input-Ground: 2828VDC 1min.
3.Insulation resistance
CE Mark, UL60950,EN60950 listed. Vout40V:Output is SELV , IEEE/Isolated analog are SELV.
40<Vout400V: Output is hazardous, IEEE/Isolated analog are SELV.
Vout40V models :Input-Outputs (SELV): 4242VDC 1min, Input-Ground: 2828VDC 1min.
1. Input voltage/freq. (*3)
3. Power Factor (Typ)
4. Efficiency (*4)
10mSec for Single-Phase and 3-phase 208V models, 6mSec for 3-Phase 400V models. Rated output power.
5. Inrush Current (*5)
6. Hold-up time (Typ)
Single Phase,230V models: 170~265Vac, 47~63Hz
3-Phase, 208V models: 170~265Vac, 47~63Hz
3-Phase, 400V models: 342~460Vac, 47~63Hz
2. Maximum
Input current
at 100% load
Single Phase,230V models:
3-Phase, 208V models:
3-Phase, 400V models: A
Single-Phase and 3-Phase 208V models: Less than 50A
3-Phase 400V models: Less than 20A
mS
24
15
7.5
GEN 8-400 10-330 15-220 20-165 30-110 40-85 600-5.5300-11150-22100-3380-4260-55
24
15
7.5
24
15
7.5
24
15
7.5
24
15
7.5
24
15
7.5
23
14.5
7
23
14.5
7
23
14.5
7
23
14.5
7
23
14.5
7
23
14.5
7
Single Phase models: 0.99@230Vac, rated output power. 3-Phase models: 0.94@208/380Vac, rated output power.
%82 84 84 86 86 88 88 88 88 88 88 87
8. RoHS Compliance
Complies with the requirements of RoHS directive.
100<Vout600V models: Input-Haz. Output: 4000VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 3550VDC 1min. Hazardous Output-Ground:2670VDC 1min. Input-Ground: 2828VDC 1min.
A
VAC
Single Phase,230V models, Power Combicon PC 6-16/3-GF-10,16 series, with Strain relief.
3-Phase, 208V & 400V models, Power Combicon PC 6-16/4-GF-10,16 series, with Strain relief.
Outline Drawing Genesys™ 3.3kW Units
|4
POWER
AAA
NOTE 2
NOTE 1
3 Phase Input Connector
NOTE 2
NOTE
1. Bus bars for 8V to 100V models (shown)
Wire clamp connector for 150V to 600V models
2. Plug connectors included with the power supply
3. Chassis slides mounting holes #10-32 marked "A"
GENERAL DEVICES P/N: C-300-S-116 or equivalent
Genesys™ Power Parallel and Series Configurations
Parallel operation - Master/Slave:
Active current sharing allows up to four identical units to be connected
in an auto-parallel configuration for four times the output power.
In Advanced Parallel Master/Slave Mode, total current is
programmed and reported by the Master, Up to four supplies
act as one.
Series operation
Up to two units may be connected in series to increase the output
voltage or to provide bipolar output. (Max 600V to Chassis Ground).
Remote Programming via RS-232 & RS-485 Interface
Standard Serial Interface allows daisy-chain control of up to 31
power supplies on the same communication bus with built-in
RS-232 & RS-485 Interface.
Genesys™
3.3kW-2U
5|
Programming Options (Factory installed)
Digital Programming via IEEE Interface P/N: IEEE
IEEE 488.2 SCPI Compliant
Program Voltage Program Current
Measure Voltage Measure Current
Over Voltage setting and shutdown Current Foldback shutdown
Error and Status Messages
New! Multi-Drop
Allows IEEE Master to control up to 31 slaves over RS-485 daisy-chain
Only the Master needs be equipped with IEEE Interface
Isolated Analog Programming
Four Channels to Program and Monitor Voltage and Current.
Isolation allows operation with floating references in harsh electrical environments.
Choose between programming with Voltage or Current.
Connection via removable terminal block: Phoenix MC1,5/8-ST-3.81.
Voltage Programming, user-selectable 0-5V or 0-10V signal. P/N: IS510
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
Current Programming with 4-20mA signal. P/N: IS420
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
LAN Interface Compliant to Class C P/N: LAN
• Meets all LXI-C Requirements
• Address Viewable on Front Panel
• Fixed and Dynamic Addressing
• Compatible with most standard Networks
VISA & SCPI Compatible
• LAN Fault Indicators
• Auto-detects LAN Cross-over Cable
• Fast Startup
Power Supply Identification / Accessories
How to order
Models 3.3kW
|6
Accessories
1. Serial Communication cable
RS-232/RS-485 cable is used to connect the power supply to the Host PC.
2. Serial link cable*
Daisy-chain up to 31 Genesys™ power supplies.
Mode
RS-485 Power Supply Connector Communication Cable
EIA/TIA-568A (RJ-45) Shield Ground L=50cm GEN/RJ45
P/N
* Included with power supply
Mode
P/N
RS-485 RS-232 RS-232
GEN/485-9 GEN/232-9 GEN/232-25
PC Connector
Communication Cable
Power Supply Connector
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-25F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~8V
0~10V
0~15V
0~20V
0~30V
0~40V
Model
0~400
0~330
0~220
0~165
0~110
0~85
3200
3300
3300
3300
3300
3400
GEN 8-400
GEN 10-330
GEN 15-220
GEN 20-165
GEN 30-110
GEN 40-85
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~60V
0~80V
0~100V
0~150V
0~300V
0~600V
Model
0~55
0~42
0~33
0~22
0~11
0~5.5
3300
3360
3300
3300
3300
3300
GEN 60-55
GEN 80-42
GEN 100-33
GEN 150-22
GEN 300-11
GEN 600-5.5
Also available, Genesys
TM
1U full Rack 750W/1500W
& Half Rack 750W
-
400
Output
Current
(0~400A)
-
IEEE
IS510
IS420
LAN
-
1P230 (
Single Phase
170~265VAC)
3P208 (Three Phase 170~265VAC)
3P400 (Three Phase 342~460VAC)
Factory Options: Factory AC Input Options:
8
Output
Voltage
(0~8V)
GEN
Series
Name Option:
Factory option P/N
RS-232/RS-485 Interface built-in Standard -
GPIB Interface IEEE
Voltage Programming Isolated Analog Interface IS510
Current Programming Isolated Analog Interface IS420
LAN Interface (Complies with Class C) LAN
Programmable DC Power Supplies
5KW in 2U
Built in RS-232 & RS-485 Interface
Advanced Parallel Standard
Optional Interfaces:
Compliant LAN
IEEE488.2 SCPI (GPIB)
Isolated Analog Programming
The GenesysTM family of programmable power supplies sets a new standard for flexible, reliable,
AC/DC power systems in OEM, Industrial and Laboratory applications.
Features include:
High Power Density 5kW in 2U
Wide Range of popular worldwide AC inputs, 3ø (208VAC, 400VAC)
Active Power Factor Correction (Three-Phase AC Input)
Output Voltage up to 600V, Current up to 600A
Built-in RS-232/RS-485 Interface Standard
Global Commands for Serial RS-232/RS-485 Interface
Auto-Re-Start / Safe-Start: user selectable
Last-Setting Memory
High Resolution 16 bit ADCs & DACs
Low Ripple & Noise
Front Panel Lock selectable from Front Panel or Software
Reliable Encoders for Voltage and Current Adjustment
Constant Voltage/Constant Current auto-crossover
Parallel Operation with Active Current Sharing; up to four identical units.
Advanced Parallel Master / Slave. Total Current is Programmed and Measured via the Master.
Independent Remote ON/OFF and Remote Enable/Disable
External Analog Programming and Monitoring (user selectable 0-5V & 0-10V)
Reliable Modular and SMT Design
19" Rack Mount capability for ATE and OEM applications
Optional Interfaces
Isolated Analog Programming and Monitoring Interface (0-5V/0-10V & 4-20mA)
IEEE 488.2 SCPI (GPIB) Multi-Drop
Compliant LAN
LabView® and LabWindows® drivers
Five Year Warranty
Worldwide Safety Agency Approvals; CE Mark for LVD and EMC Regulation
Genesys™ 5kW-2U
1|
Applications
GenesysTM power supplies have been designed to meet the demands of a wide variety of applications.
System Designers will appreciate new, standard, remote programming features such as Global commands.
Also, new high-speed status monitoring is available for the RS-485 bus.
Test Systems using the IEEE-488 bus may achieve significant cost savings by incorporating the Optional IEEE
Multi-Drop Interface for a Master and up to 30 RS-485 Multi-Drop Slaves.
Higher power systems can be configured with up to four 5kW modules. Each module is 2U with zero space
between them (zero stack).
Flexible configuration is provided by the complete GenesysTM Family: 1U 750W Half-Rack, 1U 750W/1500W
2U 3.3kW/5kW Full-Rack. All are identical in Front Panel, Rear Panel Analog, and all Digital Interface Commands.
OEM Designers have a wide variety of Inputs and Outputs from which to select depending on application and
location.
1. Remote/Local Output Voltage Sense Connections.
2. DIP Switches select 0-5V or 0-10V Programming and other functions.
3. DB25 (Female) connector allows (Non-isolated) Analog Program and Monitor and other functions.
4. RS-485 OUT to other Genesys™ Power Supplies.
5. RS-232/RS-485 IN Remote Serial Programming.
6. Output Connections: Rugged busbars (shown) for up to 100V Output; wire clamp connector for Outputs >100V.
7. Exit air assures reliable operation when zero stacked.
8. Input: 208 & 400VAC Three Phase, 50/60 Hz
AC Input Connector: PHOENIX CONTACT Power Combicon PC 6/... Series with strain relief.
9. Optional Interface Position for IEEE 488.2 SCPI (shown) or Isolated Analog Interface or LAN Interface .
Front Panel Description
Rear Panel Description
1. ON/OFF Switch
2. Air Intake allows zero stacking for maximum system flexibility and power density.
3. Reliable encoder controls Output Voltage, Address, OVP and UVL settings.
4. Volt Display shows Output Voltage and directly displays OVP, UVL and Address settings.
5. Reliable encoder controls Output Current, sets baudrate and Advanced Parallel mode.
6. Current Display shows Output Current and displays Baud rate. Displays total current in Parallel Master/Slave Mode
7. Function/Status LEDs:
8. Pushbuttons allow flexible user configuration
Coarse and Fine adjustment of Output Voltage/Current and Advanced Parallel Master or Slave
Preview settings and set Voltage/Current with Output OFF, Front Panel Lock
Parallel Master/Slave
Set OVP and UVL Limits
Set Current Foldback Protection
Go to Local Mode and select Address and Baud rate
Output ON/OFF and Auto-Re-Start/Safe-Start Mode
Alarm
Foldback Mode
Fine Control
Remote Mode
|2
Preview Settings
Output On
1 2 3 4,5
9 6 7 8
1
2 3
4 6
7
8
5
Genesys™ 5kW-2U
3|
*1: Minimum voltage is guaranteed to maximum 0.2% of rated output voltage.
*2: Minimum current is guaranteed to maximum 0.4% of rated output current.
*3: For cases where conformance to various safety standards (UL, IEC, etc) is required, to be
described as 190-240Vac (50/60Hz) for 3-Phase 208V models,
and 380~415Vac (50/60Hz) for 3-Phase 400V models.
*4: 3-Phase 208V models: At 208Vac input voltage, 3-Phase 400V:
At 380Vac input voltage. With rated output power.
*5: Not including EMI filter inrush current, less than 0.2mSec.
*6: 3-Phase 208V models: 170~265Vac, constant load. 3-Phase 400V
models: 342~460Vac, constant load.
*7: From No-Load to Full-Load, constant input voltage. Maximum drop in Remote Sense.
*8: For 8V~300V models: Measured with JEITA RC-9131A (1:1) probe. For 600V model: Measured
with 10:1 probe.
*9: From 10% to 90% or 90% to 10% of Rated Output Voltage, with rated, resistive load.
*10:From 90% to 10% of Rated Output Voltage.
*11:For load voltage change, equal to the unit voltage rating, constant input voltage.
*12:For 8V~16V models the ripple is measured from 2V to rated output voltage and rated output
current. For other models, the ripple is measured at 10~100% of rated output voltage and rated
output current.
*13: The Constant Current programming readback and monitoring accuracy does not include the
warm-up and Load regulation thermal drift.
Genesys ™ 5kW Specifications
1.2 CONSTANT CURRENT MODE
1.Max.line regulation (0.05% of Io rated)(*6)
2.Max.load regulation (0.1% of Io rated)(*11)
3.Ripple r.m.s 5Hz~1MHz. (*12)
4.Temp. coefficient
8V
~
16V models: Less than ±0.5% of rated output current over 30 minutes following power On.
1.3 PROTECTIVE FUNCTIONS
1. OCP
2. OCP Foldback
3. OVP type
4. OVP trip point
6. Over Temp. Protection
0~105% Constant Current
Output shut down when power supply change from CV to CC. User selectable.
Inverter shut-down, manual reset by AC input recycle or by OUT button or by communication port command.
User selectable , latched or non-latched.
5~88.2V 5~110.25V
5~165.3V 5~330.7V 5~661.5V
3.Rated Output Power
1.0 MODEL
1.Rated output voltage(*1)
2.Rated Output Current(*2)
2.Display
3.Indications
Voltage: 4 digits , Accuracy: 0.5% of rated output Voltage ±1 count.
Current: 4 digits, Accuracy: 0.5% of rated output current ±1 count.
Voltage, Current, Alarm, Fine, Preview, Foldback, Local, Output On, Front Panel Lock, CV/CC.
1.5 FRONT PANEL
1.Control functions Vout/ Iout manual adjust by separate encoders (coarse and fine adjustment selectable).
OVP/UVL manual adjust by Volt. Adjust encoder.
Address selection by Voltage (or current) adjust encoder. Number of addresses:31.
On/Off, Output on/off, Re-start modes (auto, safe), Foldback control (CV to CC), Go to local control.
Re-start modes (automatic restart, safe mode).
Baud rate selection: 1200,2400,4800,9600 and 19,200.
5. OVP/UVL Programming
Resolution (0.1% of Vo Rated)
Accuracy (1% of Vo Rated)
4. Readback Current
Resolution (0.012% of Io Rated )
Accuracy (0.4% of Io Rated)(*13)
3. Readback Voltage
Resolution (0.012% of Vo Rated)
Accuracy (0.15% of Vo Rated)
2. Remote Current Programming (16 bit)
Resolution (0.012% of Io Rated)
Accuracy (0.3% of Io Rated+0.1% of Io Actual Output)(*13)
1.6 Interface RS-232&RS-485 or Optional GPIB / LAN Interface
Model
1. Remote Voltage Programming (16 bit)
Resolution (0.012% of Vo Rated)
Accuracy (0.1% of Vo Rated)
1.4 ANALOG PROGRAMMING AND MONITORING
1.Vout Voltage Programming
2.Iout Voltage Programming (*13)
3.Vout Resistor Programming
4.Iout Resistor Programming (*13)
5.On/Off control (rear panel)
6.Output Current monitor (*13)
7.Output Voltage monitor
8.Power Supply OK signal
9. CV/CC Indicator
10. Enable/Disable
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±0.5% of rated Vout.
0~100%, 0~5V or 0~10V, user select. Accuracy and linearity:±1% of rated Iout.
0~100%, 0~5/10Kohm full scale,user select.,Accuracy and linearity: ±1% of rated Vout.
0~100%, 0~5/10Kohm full scale,user select. Accuracy and linearity:±1.5% of rated Iout.
By electrical. Voltage: 0~0.6V/2~15V,or dry contact ,user selectable logic.
0~5V or 0~10V , Accuracy:±1% , user selectable.
0~5V or 0~10V ,Accuracy:±1% ,user selectable.
TTL high (4~5V) -OK, 0V-Fail 500ohm series resistance.
Dry contact. Open:off , Short: on. Max. voltage at Enable/Disable in: 6V.
11. Local/Remote analog control
12. Local/Remote analog control Indicator
Open collector, CC mode: On, CV mode: Off, Maximum voltage: 30V, maximum sink current: 10mA
Open collector, Local: Off, Remote: On. Maximum voltage: 30V, maximum sink current: 10mA.
1.Max.line regulation (0.01% of rated Vo)(*6)
2.Max load regulation (0.015% of rated Vo+5mV)(*7)
3.Ripple and noise p-p 20MHz (*8)
4.Ripple r.m.s 5Hz~1MHz
5.Remote sense compensation/wire
8.Warm-up drift
9.Up-prog. response time, 0~Vo Rated (*9)
10.Down-prog response time No-load (*10)
11.Transient response time
100PPM/°C of rated output voltage,following 30 minutes warm-up
30mS
Time for output voltage to recover within 0.5% of its rated output for a load change 10-90% of rated output
current. Output set-point: 10-100%, local sense.
Less than 1mSec for models up to and including 100V. 2msec for models above 100V
80 200
100
100
7.Temp. stability
6.Temp. coefficient
Full-load (*9)
0.05% of rated Vout over 8hrs interval following 30 minutes warm-up. Constant line, load & temp.
Less than 0.05% of rated output voltage+2mV over 30 minutes following power On.
5.Temp. stability
6.Warm-up drift
100PPM/ºC from rated output current, following 30 minutes warm-up.
0.05% of rated Iout over 8hrs. interval following 30minutes warm-up. Constant line, load & temperature.
5. Output Under Voltage Limit Preset by front panel or communication port. Prevents from adjusting Vout below limit.
GEN 8-600 10-500 16-310 20-250 30-170 40-125 600-8.5300-17150-34100-5080-6560-85
4.25
8.5
15
600
60
95
500
120
5
3000
72
34
900
1.02
34
600
72
600
600
6000
1.02
8.5
17
30
300
30
50
300
35
5
2500
36
68
450
2.04
68
300
36
300
300
3000
2.04
17
34
90
150
15
27.5
120
25
5
2000
18
136
225
4.08
136
150
18
150
150
1500
4.08
25
50
100
100
10
20
100
15
5
1500
12
200
150
6.0
200
100
12
100
100
1000
6.0
32.5
65
120
80
8
17
85
12
5
1200
9.6
260
120
7.8
260
80
9.6
80
80
800
7.8
42.5
58
150
5~66.15V
60
6
14
75
10
5
1000
7.2
340
90
10.2
340
60
7.2
60
60
600
10.2
2~44.1V
62.5
125
300
40
4
11
75
10
5
900
4.80
500
60
15
500
40
4.8
40
40
400
15
2~36V
85
170
460
30
3
9.5
75
10
5
800
3.60
680
45
20.4
680
30
3.6
30
30
300
20.4
1~24V
125
250
1000
20
2
8
75
10
2
700
2.40
1000
30
30
1000
20
2.4
20
20
200
30
1~19V
155
310
1400
16
1.6
7.4
75
10
2
50
600
1.92
1240
24
37.2
1240
16
1.92
15
16
160
37.2
0.5~12V
250
500
1800
10
1.0
6.5
75
10
2
500
1.2
2000
15
60
2000
10
1.2
10
10
100
60
0.5~10V
300
600
1950
8
0.8
6.2
75
10
2
15
400
0.96
2400
12
72
2400
0.96
8
8
80
72
mA
mA
mA
V
A
W
mV
mV
mA
mA
mV
mV
mA
mA
V
mV
mV
mV
mV
mV
mV
V
PPM/°C
mS
mS
mS
mS
PPM/°C
50mS
1.1 CONSTANT VOLTAGE MODE
20V
~
600V models: Less than ±0.25% of rated output current over 30 minutes following power On.
8
8.5
5100
A
17
5100
B
34
5100
A
50
5000
C
65
5200
C
85
5100
A
125
5000
B
170
5100
B
250
5000
C
310
4960
B
500
5000
C
600
4800
A
4.Development Priority ---
By electrical signal or Open/Short: 0~0.6V or short: Remote, 2~15V or open: Local.
Outline Drawing Genesys™ 5kW Units
|4
POWER
AAA
NOTE
1. Bus bars for 8V to 100V models (shown)
Wire clamp connector for 150V to 600V models
2. Plug connectors included with the power supply
3. Chassis slides mounting holes #10-32 marked "A"
GENERAL DEVICES P/N: C-300-S-116 or equivalent
General Specifications Genesys™ 5kW
All specifications subject to change without notice.
2.1 INPUT CHARACTERISTICS
2.6 MECHANICAL CONSTRUCTION
1. Cooling
2. Dimensions (WxHxD) W: 423mm, H: 88mm, D: 442.5mm (excluding connectors, encoders, handles, etc.)
3. Weight 16 kg.
4. AC Input connector (with Protective Cover)
5.Output connectors 8V to 100V models: Bus-bars (hole Ø 10.5mm). 150V to 600V models:
wire clamp connector
, Phoenix P/N: FRONT-4-H-7.62
Forced air flow: from front to rear. No ventilation holes at the top or bottom of the chassis; Variable fan speed.
2.7 RELIABILITY SPECS
1. Warranty 5 years.
1. Parallel Operation
2. Series Operation Up to 4 identical units in master/slave mode
Up to 2 identical units. with external diodes. 600V Max to Chassis ground
2.2 POWER SUPPLY CONFIGURATION
2.3 ENVIRONMENTAL CONDITIONS
1. Operating temp
2. Storage temp -20~85°C
3. Operating humidity 20~90% RH (non-condensing).
4. Storage humidity 10~95% RH (non-condensing).
5. Vibration MIL-810F, method 514.5 , The EUT is fixed to the vibrating surface.
6. Shock Less than 20G , half sine , 11mSec. Unit is unpacked.
7. Altitude
Operating: 10000ft (3000m), Derate output current by 2%/100m above 2000m,
Non operating: 40000ft (12000m).
0~50°C, 100% load.
EN55022A, FCC part 15-A, VCCI-A.
2.4 EMC
1.Applicable Standards:
2.ESD IEC1000-4-2. Air-disch.-8KV, contact disch.-4KV
3.Fast transients IEC1000-4-4. 2KV
4.Surge immunity IEC1000-4-5. 1KV line to line, 2KV line to ground
5.Conducted immunity IEC1000-4-6, 3V
6.Radiated immunity IEC1000-4-3, 3V/m
7.Magnetic field immunity EN61000-4-8, 1A/m
8.Voltage dips EN61000-4-11
9.Conducted emission
10. Radiated emission EN55022A, FCC part 15-A, VCCI-A.
More than 100Mohm at 25°C , 70% RH.
2.5 SAFETY
1.Applicable standards:
400<Vout600V:Output is hazardous, IEEE/Isolated analog are not SELV.
2.Withstand voltage 40<Vout100V models: Input-Haz. Output: 2600VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 1900VDC 1min, Hazardous Output-Ground:1200VDC 1min. Input-Ground: 2828VDC 1min.
3.Insulation resistance
CE Mark, UL60950,EN60950 listed. Vout40V:Output is SELV , IEEE/Isolated analog are SELV.
40<Vout400V: Output is hazardous, IEEE/Isolated analog are SELV.
Vout40V models :Input-Outputs (SELV): 4242VDC 1min, Input-Ground: 2828VDC 1min.
1. Input voltage/freq. (*3)
3. Power Factor (Typ)
3-Phase 200Vac , 208Vac and 230Vac Models : 170~265Vrms , 47~63Hz
3-Phase, 400V models: 342~460Vac, 47~63Hz
2. MaximumInput
currentat 100% load 3-Phase, 170V models:
3-Phase, 342V models:
mS
21
10.5
GEN 8-600 10-500 16-310 20-250 30-170 40-125 600-8.5300-17150-34100-5080-6560-85
22
11 22
11 22
11 22
11 22
11 22
11 22
11 22
11 22
11 22
11 22
11
0.94 AT 100% LOAD AND 208V/380V/400V/415V
%
83
83 84
84 84
84 86
86 86
86 88
88 90
90 88
88 88
88 88
88 88
88 88
88
8. RoHS Compliance
Complies with the requirements of RoHS directive.
100<Vout600V models: Input-Haz. Output: 4000VDC 1min, Input-SELV: 4242VDC 1min.
Hazardous Output.-SELV: 3550VDC 1min. Hazardous Output-Ground:2670VDC 1min. Input-Ground: 2828VDC 1min.
A
VAC
3-Phase, 208V & 400V models, Power Combicon PC 6-16/4-GF-10,16 series, with Strain relief.
5mS T ypical
3-Phase 200V: 50A, 3-Phase 400V: 20A. Not including the EMI filter inrush current, less than 0.2mSec.
LESS THAN 3mA
5%
4. INRUSH PEAK CURRENT
5. EFFICIENCY AT 200V AND 380V
(A)
6. EFFICIENCY AT 170V AND 342V
(A)
7. HOLD UP TIME (CV MODE)
8. PHASE IMBALANCE
9. LEAKAGE CURRENT
NOTE 2
NOTE 1 NOTE 2
%
%
(A)
Genesys™ Power Parallel and Series Configurations
Parallel operation - Master/Slave:
Active current sharing allows up to four identical units to be connected
in an auto-parallel configuration for four times the output power.
In Advanced Parallel Master/Slave Mode, total current is
programmed and reported by the Master, Up to four supplies
act as one.
Series operation
Up to two units may be connected in series to increase the output
voltage or to provide bipolar output. (Max 600V to Chassis Ground).
Remote Programming via RS-232 & RS-485 Interface
Standard Serial Interface allows daisy-chain control of up to 31
power supplies on the same communication bus with built-in
RS-232 & RS-485 Interface.
Genesys™ 5kW-2U
5|
Programming Options (Factory installed)
Digital Programming via IEEE Interface P/N: IEEE
IEEE 488.2 SCPI Compliant
Program Voltage Program Current
Measure Voltage Measure Current
Over Voltage setting and shutdown Current Foldback shutdown
Error and Status Messages
New! Multi-Drop
Allows IEEE Master to control up to 31 slaves over RS-485 daisy-chain
Only the Master needs be equipped with IEEE Interface
Isolated Analog Programming
Four Channels to Program and Monitor Voltage and Current.
Isolation allows operation with floating references in harsh electrical environments.
Choose between programming with Voltage or Current.
Connection via removable terminal block: Phoenix MC1,5/8-ST-3.81.
Voltage Programming, user-selectable 0-5V or 0-10V signal. P/N: IS510
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
Current Programming with 4-20mA signal. P/N: IS420
Power supply Voltage and Current Programming Accuracy ±1%
Power supply Voltage and Current Monitoring Accuracy ±1.5%
LAN Interface Compliant to Class C P/N: LAN
• Meets all LXI-C Requirements
• Address Viewable on Front Panel
• Fixed and Dynamic Addressing
• Compatible with most standard Networks
VISA & SCPI Compatible
• LAN Fault Indicators
• Auto-detects LAN Cross-over Cable
• Fast Startup
Power Supply Identification / Accessories
How to order
Models 5kW
|6
Accessories
1. Serial Communication cable
RS-232/RS-485 cable is used to connect the power supply to the Host PC.
2. Serial link cable*
Daisy-chain up to 31 Genesys™ power supplies.
Mode
RS-485 Power Supply Connector Communication Cable
EIA/TIA-568A (RJ-45) Shield Ground L=50cm GEN/RJ45
P/N
* Included with power supply
Mode
P/N
RS-485 RS-232 RS-232
GEN/485-9 GEN/232-9 GEN/232-25
PC Connector
Communication Cable
Power Supply Connector
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-9F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
DB-25F
Shield Ground L=2m
EIA/TIA-568A (RJ-45)
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~8V
0~10V
0~16V
0~20V
0~30V
0~40V
Model
0~600
0~500
0~310
0~250
0~170
0~125
4800
5000
4960
5000
5100
5000
GEN 8-600
GEN 10-500
GEN 16-310
GEN 20-250
GEN 30-170
GEN 40-125
Output
Power
( W )
Output
Current
( A )
Output
Voltage
VDC
0~60V
0~80V
0~100V
0~150V
0~300V
0~600V
Model
0~85
0~65
0~50
0~34
0~17
0~8.5
5100
5200
5000
5100
5100
5100
GEN 60-85
GEN 80-65
GEN 100-50
GEN 150-34
GEN 300-17
GEN 600-8.5
Also available, GenesysTM
1U Half Rack 750W
1U full Rack 750W/1500W
2U full Rack 3300W
-
600
Output
Current
(0~600A)
-
IEEE
IS510
IS420
LAN
-
3P208 (Three Phase 170~265VAC)
3P400 (Three Phase 342~460VAC)
Factory Options: Factory AC Input Options:
8
Output
Voltage
(0~8V)
GEN
Series
Name Option:
Factory option P/N
RS-232/RS-485 Interface built-in Standard -
GPIB Interface IEEE
Voltage Programming Isolated Analog Interface IS510
Current Programming Isolated Analog Interface IS420
LAN Interface (Complies with Class C) LAN