IMA-x600 Power Supply Series
EMC
IMA-x600-xx
Applied standards Criteria
Radiated emissions 1) EN 55011, EN 55022 and FCC, Class B
Conducted emissions 1) EN 55011, EN 55022 and FCC, Class B
Power line harmonics EN 61000-3-2, Class A
Voltage icker EN 61000-3-3
ESD EN 61000-4-2, level 4, 8 kV contact, 15 kV air A
Radiated immunity EN 61000-4-3, level 3, 10 V/m A
Electrical fast transient EN 61000-4-4, level 4, ±4 kV A
Surge immunity EN 61000-4-5, level 4, 2 kV DM, 4 kV CM A
Conducted RF immunity EN 61000-4-6, level 3, 10 Vrms A
Power frequency magnetic eld EN 61000-4-8, level 3, 10 A/m A
Voltage dips and sags EN 61000-4-11, 30%, 500 ms
EN 61000-4-11, 60%, 100 ms
EN 61000-4-11,100%, 10 ms
EN 60601-1-2, 30%, 500 ms
EN 60601-1-2, 60%, 100 ms
EN 60601-1-2, 100%, 10 ms
EN 60601-1-2, 100%, 5000 ms
A
B
A
A
B
A
B
Ring wave EN 61000-4-12, level 3, 1 kV DM, 2 kV CM A
Voltage uctuations EN 61000-4-14, Class 3 A
1) Power Supply Unit inside a dummy system
Safety standards and directives 1)
IMA-x600-xx
IEC/EN 60950-1, Edition 2 and all national devia-
tions
UL 60950-1/CSA 22.2 No 60950-1, Edition 2;
5,000 m (16,400 ft) altitude, 120 V to 300 Vdc and 100 V to 240 ±10% Vac
(UL File E191395)
IEC/EN 60601-1, Edition 3 (tested against Edition
2, too) and all national deviations
IEC 60601-1(2005), EN60601-1(2006)
ANSI/AAMI ES 60601-1(2005)
CAN/CSA C22.2 No. 60601-1 (2008);
3,000 m (9,800 ft) altitude, 100 V to 240 Vac ±10% (UL File E325662)
Protection class I
1) Designed to support Type B Applied Part End Product Requirements