2 w ww .bk prec ision.c om
Test Bench, High Performance DMM
Models 388B, 389A, 390A & 391A
v040611
Specifications 391A 390A 389A 388B
Current True RMS reading
Ranges 200µA, 2mA, 20mA, 200mA, 20A* 400µA, 4mA, 40mA, 400mA, 20A* 400mA, 4mA, 40mA, 400mA, 2A, 20A*
Resolution 10nA, 100nA, 1µA, 10µA, 1mA 0.1µA, 1µA, 10µA, 100µA, 10µA 100nA, 1µA, 10µA,100µA,1mA,10mA
DC Accuracy 200µA to200mA:±(0.5% rdg + 5 dgt.)
20A: ±(2% rdg + 10 dgt.)
400µA - 400mA:±(1% rdg + 5dgt)
20A: ±(2% rdg + 3 dgt)
±(1% rdg + 1 dgt)
±(2% rdg + 3dgt)
400µA - 400mA: ±(1% rdg + 1 dgt)
2A: ±(1.5% rdg + 1 dgt)
20A: ±(3% rdg + 3 dgt)
AC Accuracy 200µA to200mA:±(1.2% rdg + 10 dgt.)
20A: ±(2.5% rdg + 20 dgt.)
400µA to 400mA:±(1.5% rdg + 4 dgt)
20A: ±(2.5% rdg + 4 dgt.)
400µA - 400mA:±(1.5% rdg + 1 dgt)
2A: ±(2% rdg + 4 dgt)
20A: ±(3.5% rdg + 4 dgt)
Input Protection µA/mA input:0.5A/500V fast blow ceramic fuse
20A input: 20A/600A fast blow ceramic fuse
µA/mA input: 0.5A/500V fast blow ceramic fuse
20A input: 20A/600A fast blow ceramic fuse
µA/mA input: 2A/600V fast blow ceramic fuse
20A input: 20A/600V fast blow ceramic fuse
Max. Burden Voltage 600mV (900mV on 20A range) 500mV (2V on 4mA, 400mA ranges) 600mV (900mV on 2A, 20A ranges)
Resistance
Ranges 200Ω; 2kΩ, 20kΩ, 200kΩ; 2MΩ, 20MΩ400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, 40MΩ
Resolution 10mΩ, 100mΩ, 1Ω, 10Ω, 100Ω, 1kΩ100mΩ, 1Ω, 10Ω, 100Ω, 1kΩ, 10kΩ
Accuracy
200Ω, 2MΩ: ±(0.25% rdg + 10 dgt.)
2kΩ to 200kΩ: ±(0.15% rdg + 3 dgt.)
20MΩ: ±(1.0% + 10 dgt.)
400Ω to 400kΩ: ±(0.5% rdg + 4 dgt.)
4MΩ: ±(1% rdg + 5 dgt.)
40MΩ: ±(2% rdg + 5 dgt.)
400Ω: ±(1% rdg + 4 dgt.)
4kΩ to 4MΩ: ±(0.8% rdg + 4 dgt.)
40MΩ: ±(2% rdg + 5 dgt.)
Open Circuit Voltage 3.2VDC typ. -0.45VDC typ. (-1.2VDC on 400Ω range) 0.6VDC typ. (3.2VDC on 400Ω range)
Overload Protection 500VDC or ACrms
Diode Test T
ested at 1mA, 3.2VDC max. type.
±(1% rdg + 10 dgt.) accuracy
T
ested at 1.2mA, 3.0VDC max. type.
±(1.5% rdg + 3 dgt.) accuracy
T
ested at 1mA, 3.2VDC max. type.
±(1.5% rdg + 3 dgt.) accuracy
Transistor Test (hFE) Does not apply Does not apply hFE range: 0 - 1000, hFE base current: 10µADC
Capacitance
Ranges Does not apply 4nF. 40nF, 400nF, 4µF, 40µF, 400µF, 4mF, 40mF 4nF. 40nF, 400nF, 4µF, 40µF
Resolution 1pF, 10pF, 100pF, 1nF, 10nF, 100nF, 1µF, 10µF 0.1pF, 1pF, 10pF, 100pF, 1nF
Accuracy 4nF: ±(3% rdg + 20 dgt.) 4nF to 40µF: ±(3% rdg + 5 dgt.)
400µFto 40mF: ±(5% rdg + 10 dgt.)
±(3% rdg + 4 dgt.)
Test Voltage <1V <3.5V
Overload Protection 500VDC or AC rms
Frequency
Ranges 2kHz, 20kHz, 200kHz 4kHz, 40kHz, 400kHz, 4MHz, 40MHz 4kHz, 40kHz, 400kHz, 4MHz
Resolution 0.1 Hz, 1Hz, 10Hz 1Hz, 10Hz, 100Hz, 10kHz, 100kHz 1Hz, 10Hz, 100Hz, 10kHz
Accuracy ±(0.1% rdg + 3 dgt.) ±(0.1% rdg + 3 dgt.) ±(0.25% + 4 Digits) ±(0.1% + 2 Digits)
Sensitivity 50mVrms min. (@>30 & <70% duty
cycle: 400mVrms min.)
1Hz - 4MHz: 1Vrms
4MHz - 40MHz: >2Vrms, <5Vrms
250mVrms min. on 10Hz to 1MHz
500mVrms min. on 1MHz to 4MHz
Minimum Pulse Width >25ns >25ns >2µs
Duty Cycle Limits >30% & <70% >30% & <70%
Minimum Input Range 2kHz:10Hz:20kHz:>60dgt; 200kHz>60dgt
Overload Protection 500VDC or AC rms 500VDC or AC rms
Logic Test
Logic Threshold Hi=2.8±0.8V, 0=0.8 ± 0.5V Does not apply Does not apply Hi=2.8±0.8V, 0=0.8 ± 0.5V
Frequency Response 20MHz 20MHz
Pulse Width 25ns 25ns
Pulse Limits >20% and <80% >20% and <80%
Indication 40m sec beep at logic 1 (Hi) 40m sec beep at logic 1 (Hi)
Overload Protection 500V DC or ACrms 500V DC or ACrms
Temperature
Range, Resolution Does not apply -58° to +2372°F, 1F°, (-50°to + 1300°C, 1C°) Does not apply Does not apply
Accuracy ±(0.8% rdg + 2°C) -50° - 400°C, ±(1% rdg + 2°C) 400° - 1300°C
Duty Cycle
Range, Resolution 0 to 90%, 0.1% Does not apply Does not apply Does not apply
Accuracy (5V logic) ±(2.0% rdg + 10 dgt.)
Minimum Pulse Width 10µs
Frequency Range 40Hz to 20kHz
Overload Protection 500VDC or AC rms
General
Display 20000 count, 4 1/2 digit LCD 4000 count, 3 3/4 digit LCD with 41 segment analog bar graph 4000 count, 3 3/4 digit LCD
Polarity Automatic, positive implied, negative polarity indication
Operating Temperature 32° to 122° (0˚ to 50˚C), 0 to 70% R.H.
Dimensions (HxWxD) 7.8 x 3.5 x 1.57" (198 x 90 x 40mm)
Weight 11.3 oz. (320g).
Three Year Warranty
SUPPLIED ACCESSORIES: 9 Volt Battery, Test Leads, Instruction Manual, Thermocouple Probe (Model 390A), Software and USB Interface Cable (Model 390A)
OPTIONAL ACCESSORIES: TL 2A Deluxe Test Leads, TL 3 Accessory Tip Kit (for TL 2A), PR 28A High Voltage Probe (40kVDC), TL 130A General Purpose DMM Kit, Carrying Case (not included): LC 29B
* 10 A continuous, 20 A for 30 seconds maximum