cui.com
date 07/10/2015 │ page 2 of 4CUI Inc │ SERIES: ETMA 30W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
INPUT
parameter conditions/description min typ max units
voltage 90 264 Vac
frequency 47 63 Hz
input current 0.6 A
inrush current at 240 Vac 50 A
no load power consumption 0.3 W
OUTPUT
parameter conditions/description min typ max units
line regulation1±1 %
load regulation2
5 Vdc output
9 Vdc output
12, 15 Vdc outputs
18, 24 Vdc outputs
±6
±4
±3
±2
%
%
%
%
voltage accuracy ±2 %
hold-up time at 115 Vac 16 ms
switching frequency 70 kHz
temperature coefcient ±0.05 %/°C
Note: 1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
PROTECTIONS
parameter conditions/description
over voltage protection TVS component to clamp
short circuit protection continuous, auto restart
SAFETY & COMPLIANCE
parameter conditions/description min typ max units
isolation voltage input to output 5,656 Vdc
safety approvals IEC 60601-1, EN 60601-1, UL 60601-1
EMI/EMC EN 55011, EN 60601-1-2, EN 61000-3-(2,3)
MTBF as per MIL-HDBK-217F, 115 Vac, 25 °C 300,000 hours
RoHS 2011/65/EU
ENVIRONMENTAL
parameter conditions/description min typ max units
operating temperature 0 40 °C
storage temperature -20 85 °C
humidity non-condensing 93 %
For more information, please visit the product page | For information on related products, view Dc Power Jacks